Investigation of sound pressure leakage effect for primary calibration down to 10-2 Hz using a laser pistonphone system

被引:2
|
作者
Hirano, Koto [1 ]
Takahashi, Hironobu [1 ]
Yamada, Keisuke [1 ]
Nozato, Hideaki [1 ]
机构
[1] Natl Metrol Inst Japan, Natl Inst Adv Ind Sci & Technol, 1-1-1 Umezono, Tsukuba, Ibaraki 3058563, Japan
关键词
infrasound; laser pistonphone method; calibration; sound pressure leakage; pressure sensor;
D O I
10.1088/1361-6501/ad23c3
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Recently, the demand for sensor calibration in the infrasonic range has increased to obtain accurate information when monitoring infrasound generated by large-scale natural disasters. The laser pistonphone method is a primary calibration method to evaluate infrasound sensors in the low-frequency range. In this method, sound pressure is generated in a fixed sealed volume of air, by a volume change via an attached piston movement. By measuring the piston displacement using laser interferometer, the generated sound pressure can be calculated by multiplying the velocity of volume change by acoustic transfer impedance of a pistonphone. The main concern with this method is that sound pressure leakage from a gap adjacent to piston significantly changes the acoustic transfer impedance at lower frequencies. To apply the laser pistonphone method in the low-frequency range down to 10(-2) Hz, it is essential to accurately evaluate and compensate for the leakage effect, i.e. the gap's acoustic transfer impedance. In this study, we propose a technique for experimentally evaluating the gap impedance. The main idea is to determine the total acoustic transfer impedance by dividing the sound pressure by the volume velocity and then deducing the gap impedance by subtracting the chamber impedance from the obtained total acoustic transfer impedance. A digital pressure sensor was used to precisely measure sound pressure below 10(0) Hz because pressure sensors are suitable for accurate measurement of pressure fluctuations at low frequencies. We validated the proposed approach by calibrating an analog pressure sensor that outputs an analog voltage proportional to the absolute pressure. As a result, the sensitivity calibrated by the laser pistonphone method at 0.02 Hz agreed with the static sensitivity provided by the manufacturer within 0.02 dB.
引用
收藏
页数:12
相关论文
共 4 条
  • [1] INRIM continuous expansion system as high vacuum primary standard for gas pressure measurements below 9 x 10-2 Pa
    Bergoglio, M.
    Mari, D.
    VACUUM, 2009, 84 (01) : 270 - 273
  • [2] AC (50 HZ) AND DC ELECTRICAL BREAKDOWN OF VACUUM GAPS AND WITH VARIATION OF AIR-PRESSURE IN RANGE 10-9 -10-2 TORR USING OFHC COPPER, NICKEL, ALUMINUM, AND NIOBIUM PARALLEL PLANAR ELECTRODES
    HACKAM, R
    ALTCHEH, L
    JOURNAL OF APPLIED PHYSICS, 1975, 46 (02) : 627 - 636
  • [3] Final report of key comparison AFRIMETS.AUV.A-K5: primary pressure calibration of LS1P microphones according to IEC 61094-2, over the frequency range 2 Hz to 10 kHz.
    Nel, R.
    Avison, J.
    Harris, P.
    Blabla, M.
    Hamalainen, J.
    METROLOGIA, 2017, 54
  • [4] Generation of above 1010 Temporal Contrast, above 1020 W/cm2 Peak Intensity Pulses at a 10 Hz Repetition Rate using an OPCPA Preamplifier in a Double CPA, Ti:sapphire Laser System
    Kiriyama, Hiromitsu
    Mori, Michiaki
    Nakai, Yoshiki
    Shimomura, Takuya
    Tanoue, Manabu
    Okada, Hajime
    Kondo, Shuji
    Kanazawa, Shuhei
    Sagisaka, Akito
    Daito, Izuru
    Kotaki, Hideyuki
    Sasao, Hajime
    Wakai, Daisuke
    Tanaka, Momoko
    Ochi, Yoshihiro
    Sugiyama, Akira
    Daido, Hiroyuki
    Bulanov, Sergei
    Koike, Masato
    Bolton, Paul R.
    Kawanishi, Shunichi
    LASER-DRIVEN RELATIVISTIC PLASMAS APPLIED TO SCIENCE, INDUSTRY AND MEDICINE, 2009, 1153 : 3 - 6