The application of inverted lateral clavicle locking compression plate for medial clavicle fractures: biomechanical study and clinical implementation

被引:2
|
作者
Kritsaneephaiboon, Apipop [1 ]
Jitprapaikulsarn, Surasak [2 ]
Chantarapanich, Nattapon [3 ]
Klabklay, Prapakorn [1 ]
Gromprasit, Arthit [2 ]
Patamamongkonchai, Chawanan [2 ]
Dissaneewate, Pornpanit [1 ]
机构
[1] Prince Songkla Univ, Dept Orthoped Surg & Phys Med, Hat Yai, Thailand
[2] Buddhachinaraj Hosp, Dept Orthoped, Phitsanulok, Thailand
[3] Kasetsart Univ, Fac Engn Sriracha, Dept Mech Engn, Sriracha, Chonburi, Thailand
关键词
Medial clavicle fracture; Inverted lateral clavicle locking compression plate; Finite element study; END; EPIDEMIOLOGY; FIXATION;
D O I
10.1007/s00590-023-03826-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundThere has been no absolute consensus on the optimal implant for fixation of medial clavicle fracture. The purpose of the present study was to test the biomechanical efficacy of inverted lateral clavicle locking compression plate (LCP-LC) in fixation of this specific fracture location by finite element analysis.MethodsTransverse and comminuted medial clavicle fracture simulated models stabilized by three different devices including inverted LCP-LC, superior clavicle LCP (LCP-SC), and dual reconstruction (LCP-RP) with LCP-SC were investigated biomechanical performance under three loading conditions, i.e., axial compression, inferior bending, and axial torsion.ResultsEQV stress exhibited on implant and elastic strain at fracture site under inferior bending was greater than other loading cases. LCP-SC construct represented the greatest EQV stress and elastic strain. The inverted LCP-LC construct demonstrated lower EQV stress than the LCP-SC construct and was comparable to dual plating. Under axial compression and axial torsion, elastic strain produced from the inverted LCP-LC construct was comparable to dual plating, but greater than dual plating when subjected to inferior bending.ConclusionBy the results of FE analysis, inverted LCP-LC could be a potential application for fixation of clavicle fracture with short medial fragment.
引用
收藏
页码:1465 / 1478
页数:14
相关论文
共 50 条
  • [1] The application of inverted lateral clavicle locking compression plate for medial clavicle fractures: biomechanical study and clinical implementation
    Apipop Kritsaneephaiboon
    Surasak Jitprapaikulsarn
    Nattapon Chantarapanich
    Prapakorn Klabklay
    Arthit Gromprasit
    Chawanan Patamamongkonchai
    Pornpanit Dissaneewate
    European Journal of Orthopaedic Surgery & Traumatology, 2024, 34 : 1465 - 1478
  • [2] Inverted distal clavicle anatomic locking plate for displaced medial clavicle fracture
    Wang, Yong
    Jiang, Jiannong
    Dou, Bin
    Zhang, Panjun
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2015, 135 (09) : 1241 - 1245
  • [3] Inverted distal clavicle anatomic locking plate for displaced medial clavicle fracture
    Yong Wang
    Jiannong Jiang
    Bin Dou
    Panjun Zhang
    Archives of Orthopaedic and Trauma Surgery, 2015, 135 : 1241 - 1245
  • [4] Medial clavicle pseudarthrosis successfully treated with an inverted distal clavicle locking plate
    Sasaki, Yu
    Lee, Sang Yang
    Iwakura, Takashi
    Fukui, Tomoaki
    Oe, Keisuke
    Matsumoto, Tomoyuki
    Matsushita, Takehiko
    Kawamoto, Teruya
    Mifune, Yutaka
    Kuroda, Ryosuke
    Niikura, Takahiro
    ANNALS OF MEDICINE AND SURGERY, 2019, 44 : 1 - 4
  • [5] Displaced medial clavicle fractures; operative treatment with locking compression plate fixation
    Frima, H.
    Houwert, R. M.
    Sommer, C.
    BRITISH JOURNAL OF SURGERY, 2018, 105 : 10 - 10
  • [6] Displaced medial clavicle fractures: operative treatment with locking compression plate fixation
    Herman Frima
    Roderick M. Houwert
    Christoph Sommer
    European Journal of Trauma and Emergency Surgery, 2020, 46 : 207 - 213
  • [7] Displaced medial clavicle fractures: operative treatment with locking compression plate fixation
    Frima, Herman
    Houwert, Roderick M.
    Sommer, Christoph
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2020, 46 (01) : 207 - 213
  • [8] Excellent clinical and radiological outcome following locking compression plate fixation of displaced medial clavicle fractures
    Markus Wurm
    Sebastian Siebenlist
    Michael Zyskowski
    Patrick Pflüger
    Peter Biberthaler
    Marc Beirer
    Chlodwig Kirchhoff
    BMC Musculoskeletal Disorders, 22
  • [9] Excellent clinical and radiological outcome following locking compression plate fixation of displaced medial clavicle fractures
    Wurm, Markus
    Siebenlist, Sebastian
    Zyskowski, Michael
    Pflueger, Patrick
    Biberthaler, Peter
    Beirer, Marc
    Kirchhoff, Chlodwig
    BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [10] A CLINICAL STUDY OF DISPLACED CLAVICLE FRACTURES TREATED WITH ANATOMICALLY PRECONTOURED LOCKING COMPRESSION PLATE
    Mulimani, Venkatesh M.
    Ramesh, Manikya
    Babu, Chavan Pramod
    Hosthota, Akash
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2016, 5 (83): : 6218 - 6222