A passive dorsiflexing ankle prosthesis to increase minimum foot clearance during swing

被引:1
|
作者
Bartlett, Harrison L. [1 ]
Shepherd, Max K. [2 ,3 ]
Lawson, Brian E. [1 ]
机构
[1] Little Room Innovat LLC, Ann Arbor, MI USA
[2] Northeastern Univ, Dept Mech Engn, Boston, MA 02115 USA
[3] Northeastern Univ, Dept Phys Therapy, Boston, MA 02115 USA
来源
WEARABLE TECHNOLOGIES | 2023年 / 4卷
关键词
Amputation; biomechanics; minimum foot clearance; gait; foot; prosthesis; QUALITY-OF-LIFE; TOE CLEARANCE; BACK-PAIN; LEG-LENGTH; RESIDUAL LIMB; RISK-FACTORS; WALKING; PEOPLE; AMPUTATION; AMPUTEES;
D O I
10.1017/wtc.2023.10
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The biological ankle dorsiflexes several degrees during swing to provide adequate clearance between the foot and ground, but conventional energy storage and return (ESR) prosthetic feet remain in their neutral position, increasing the risk of toe scuffs and tripping. We present a new prosthetic ankle intended to reduce fall risk by dorsiflexing the ankle joint during swing, thereby increasing the minimum clearance between the foot and ground. Unlike previous approaches to providing swing dorsiflexion such as powered ankles or hydraulic systems with dissipative yielding in stance, our ankle device features a spring-loaded linkage that adopts a neutral angle during stance, allowing ESR, but adopts a dorsiflexed angle during swing. The ankle unit was designed, fabricated, and assessed in level ground walking trials on a unilateral transtibial prosthesis user to experimentally validate its stance and swing phase behaviors. The assessment consisted of three conditions: the ankle in an operational configuration, the ankle in a locked configuration (unable to dorsiflex), and the subject's daily use ESR prosthesis. When the ankle was operational, minimum foot clearance (MFC) increased by 13 mm relative to the locked configuration and 15 mm relative to his daily use prosthesis. Stance phase energy return was not significantly impacted in the operational configuration. The increase in MFC provided by the passive dorsiflexing ankle prosthesis may be sufficient to decrease the rate of falls experienced by prosthesis users in the real world.
引用
收藏
页数:19
相关论文
共 44 条
  • [1] Evaluation of an articulated passive ankle–foot prosthesis
    Elke Lathouwers
    Toon Ampe
    María Alejandra Díaz
    Romain Meeusen
    Kevin De Pauw
    BioMedical Engineering OnLine, 21
  • [2] Evaluation of an articulated passive ankle-foot prosthesis
    Lathouwers, Elke
    Ampe, Toon
    Diaz, Maria Alejandra
    Meeusen, Romain
    De Pauw, Kevin
    BIOMEDICAL ENGINEERING ONLINE, 2022, 21 (01)
  • [3] Does Gait with an Ankle Foot Orthosis Improve or Compromise Minimum Foot Clearance?
    Fonseca, Pedro
    Machado, Leandro
    Sousa, Manoela Vieira
    Sebastiao, Ricardo
    Sousa, Filipa
    Figueiredo, Joana
    Santos, Cristina P.
    Vilas-Boas, Joao Paulo
    SENSORS, 2021, 21 (23)
  • [4] Contribution of Prosthetic Knee and Ankle Mechanisms to Swing-Phase Foot Clearance
    Sensinger, Jonathon W.
    Intawachirarat, Nattaporn
    Gard, Steven A.
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2013, 21 (01) : 74 - 80
  • [5] A PASSIVE TWO-DEGREE-OF-FREEDOM ANKLE-FOOT PROSTHESIS
    Elley, Colin
    Nelson, Carl A.
    2018 DESIGN OF MEDICAL DEVICES CONFERENCE, 2018,
  • [6] The VSPA Foot: A Quasi-Passive Ankle-Foot Prosthesis With Continuously Variable Stiffness
    Shepherd, Max K.
    Rouse, Elliott J.
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2017, 25 (12) : 2375 - 2386
  • [7] Passive Ankle-Foot Prosthesis Prototype with Extended Push-Off
    Brackx, Branko
    Van Damme, Michael
    Matthys, Arnout
    Vanderborght, Bram
    Lefeber, Dirk
    INTERNATIONAL JOURNAL OF ADVANCED ROBOTIC SYSTEMS, 2013, 10
  • [8] Design and Characterization of a Quasi-Passive Pneumatic Foot-Ankle Prosthesis
    Lee, Jeffrey D.
    Mooney, Luke M.
    Rouse, Elliott J.
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2017, 25 (07) : 823 - 831
  • [9] The ankle dorsiflexion kinetics demand to increase swing phase foot-ground clearance: implications for assistive device design and energy demands
    Bajelan, Soheil
    Sparrow, W. A.
    Begg, Rezaul
    JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2024, 21 (01)
  • [10] Prosthesis foot that has an action of ankle joint during walking
    Takahama, Itsurou
    Seimitsu Kogaku Kaishi/Journal of the Japan Society for Precision Engineering, 1994, 60 (05): : 647 - 651