A miniature filamentless X-ray tube for the on-board calibration of space-borne X-ray detectors

被引:0
|
作者
Wang, Huan [1 ,2 ]
Cao, Yang [1 ]
Sang, Peng [1 ]
Li, Fan [1 ,2 ]
Yu, Daochun [1 ,2 ]
Li, Baoquan [1 ,2 ]
机构
[1] Chinese Acad Sci, Natl Space Sci Ctr, Beijing, Peoples R China
[2] Univ Chinese Acad Sci, Beijing, Peoples R China
关键词
On-orbit calibration of X-ray detectors; Miniature filamentless X-ray tube; Linearity calibration method; L-series of characteristic X-ray of tungsten; target;
D O I
10.1016/j.nima.2023.168133
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The on-orbit calibration of the space X-ray detectors is important to obtain the accurate science data. Radioactive isotopes such as 241Am and 55Fe are generally used to implement calibration for space X-ray mission. However, the radioactive source is potential risk for the health safety and also causes a persistent radiation background during X-ray detector operation in space. On-orbit calibration of X-ray detectors has always been a difficult problem for space missions. This paper presents a miniature filamentless X-ray tube (MFXT) with tungsten target which can be used to calibrate X-ray detector on-orbit. At -12 kV high-voltage applied on MFXT, the strong L-series of characteristic X-ray of tungsten target is produced, so the gain and offset parameters of detector can be easily determined by the energy peaks of characteristic X-rays of tungsten. And after about 24 h of cumulative experiment, at least two characteristic ray peaks that can be used for calibration are formed with 5 minutes cumulative time after turning on it again with 5.2E+3 counts/s/sr radiation intensity by MFXT. As of the miniature, low power consumption and high reliability, this MFXT is an alternative to replace the radioactive source for on-orbit detector calibration in space missions.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] An alpha tagged X-ray source for the calibration of space borne X-ray detectors
    Rao, A. R.
    Naik, Sachindra
    Patil, Milind
    Malkar, J. P.
    Kumar, R. P. S. Kalyan
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 616 (01): : 55 - 58
  • [2] Thick CZT detectors for space-borne x-ray astronomy
    Krawczynski, H
    Jung, I
    Perkins, J
    Burger, A
    Groza, M
    HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS VI, 2004, 5540 : 1 - 13
  • [3] A space-borne X-ray imaging polarimeter
    Weisskopf, Martin C. C.
    Soffitta, Paolo
    Ramsey, Brian D. D.
    Baldini, Luca
    O'Dell, Steve L. L.
    Costa, Enrico
    Kaaret, Philip
    NATURE ASTRONOMY, 2023, 7 (05) : 635 - 635
  • [4] A space-borne X-ray imaging polarimeter
    Martin C. Weisskopf
    Paolo Soffitta
    Brian D. Ramsey
    Luca Baldini
    Steve L. O’Dell
    Enrico Costa
    Philip Kaaret
    Nature Astronomy, 2023, 7 : 635 - 635
  • [5] A miniature x-ray tube
    Haga, A
    Senda, S
    Sakai, Y
    Mizuta, Y
    Kita, S
    Okuyama, F
    APPLIED PHYSICS LETTERS, 2004, 84 (12) : 2208 - 2210
  • [6] MINIATURE LOW ENERGY X-RAY CALIBRATION TUBE
    GRADER, RJ
    SCUDDER, JK
    TINDALL, WE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1969, 40 (09): : 1160 - &
  • [7] MINIATURE X-RAY POINT SOURCE FOR ALIGNMENT AND CALIBRATION OF X-RAY OPTICS
    PRICE, RH
    BOYLE, MJ
    GLAROS, SS
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1977, 22 (09): : 1196 - 1196
  • [8] X-ray polarimetry on-board HXMT
    Soffitta, Paolo
    Bellazzini, Ronaldo
    Tagliaferri, Gianpiero
    Costa, Enrico
    Pareschi, Giovanni
    Basso, Stefano
    Cotroneo, Vincenzo
    Frutti, Massimo
    Lazzarotto, Francesco
    Muleri, Fabio
    Rubini, Alda
    Spaiidre, Gloria
    Bez, Alessandro
    Baldini, Luca
    Bregeon, Jean
    Minuti, Massimo
    Matt, Giorgio
    Frontera, Fillipo
    SPACE TELESCOPES AND INSTRUMENTATION 2008: ULTRAVIOLET TO GAMMA RAY, PTS 1 AND 2, 2008, 7011
  • [9] INVESTIGATION OF PULSE SHAPE ANALYZERS FOR PHOSWICH DETECTORS IN SPACE-BORNE HARD X-RAY EXPERIMENTS
    BLEEKER, JAM
    OVERTOOM, JM
    NUCLEAR INSTRUMENTS & METHODS, 1979, 167 (03): : 505 - 513
  • [10] On-board event processing algorithms for a CCD-based space borne X-ray spectrometer
    Chun, HJ
    Bowles, JA
    BranduardiRaymont, G
    Gowen, RA
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 376 (02): : 254 - 262