Development of a multi-channel gamma-blind fast neutron detector based on wavelength shifting fibers embedded in a ZnS:Ag epoxy mixture

被引:0
|
作者
Wolfertz, Alexander [1 ]
Adams, Robert [2 ]
Perret, Gregory [1 ]
Lamirand, Vincent [3 ]
机构
[1] Paul Scherrer Inst PSI, Nucl Energy & Safety Res Div, Forschungsstr 111, CH-5232 Villigen, Switzerland
[2] Swiss Fed Inst Technol Zurich ETHZ, Lab Nucl Energy Syst, Sonneggstr 3, CH-8092 Zurich, Switzerland
[3] Ecole Polytech Fed Lausanne EPFL, Lab Reactor Phys & Syst Behav, EPFL SB IPHYS LRS, PH D2 435 Batiment PH,Stn 3, CH-1015 Lausanne, Switzerland
关键词
Fast neutron detector; Spent fuel characterization; Wavelength shifting fibers; Post-irradiation experiments; ZnS(Ag) scintillator;
D O I
10.1016/j.nima.2023.168385
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper presents a fast neutron detector based on elastic scattering with hydrogen, a silver-activated zinc sulfide scintillator to convert the recoil proton energy to light, and wavelength-shifting fibers (WLSFs) to collect the scintillation light. The detector uses silicon photomultipliers (SiPMs) to recognize individual scintillation photons and a digital filter algorithm based on single photon counting to find clusters of photons belonging to neutron events. The detector presented in this paper features four detection channels, arranged in a 2 x 2 square. The sensitive volume of each detection channel covers a similar to 5mm by similar to 5mm area from the frontal direction, is 3 cm long, and contains 49 WLSFs. The detector is versatile and performs well under different conditions. Its performance can be tuned to match different applications by simply changing some parameters of the digital filter algorithm. This is illustrated in this paper by extensive measurements in different environments. Using one set of parameters, the detector achieved a gamma-blindness of 10-8 with an intrinsic neutron detection efficiency of similar to 1%. With another set of parameters and with lower requirements for gamma blindness, the intrinsic neutron detection efficiency was increased to similar to 11%. Yet another set of parameters allows the detector to time incoming fast neutrons with an accuracy of similar to 60ns. Additionally, the decay time of the scintillation light created by neutron events was measured, falling to 10% of its peak value in similar to 10 mu s. Finally, the detector was exposed to strong gamma radiation for a prolonged time to test its radiation resistance. The detection efficiency dropped about linearly with the accumulated gamma fluence, reaching a drop of 40% compared to the initial efficiency at a total gamma fluence of similar to 2 & sdot;1013 cm(-2).
引用
收藏
页数:12
相关论文
共 11 条
  • [1] First tests of a gamma-blind fast neutron detector using a ZnS:Ag-epoxy mixture cast around wavelength-shifting fibers
    Wolfertz, Alexander
    Adams, Robert
    Perret, Gregory
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 971
  • [2] Gamma Sensitivity of a ZnS:Ag(6-LiF) Wavelength Shifting Fiber Neutron Detector in Mixed Neutron-gamma Fields
    Sykora, Garrett Jeff
    Schooneveld, Erik M.
    Rhodes, Nigel J.
    Van Eijck, Lambert
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 1567 - 1571
  • [3] Cold neutron radiation dose effects on a 6LiF:ZnS(Ag) neutron detector with wavelength shifting fibers and SiPM photodetector
    Pritchard, K.
    Hussey, D.
    Osovizky, A.
    Ziegler, J.
    Binkley, E.
    Tsai, P.
    Hadad, N.
    Jackson, M.
    Hurlbut, C.
    Baltic, G. M.
    Majkrzak, C. F.
    Maliszewskyj, N. C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 986
  • [4] Development of a two-dimensional imaging detector based on a neutron scintillator with wavelength-shifting fibers
    Sakai, K
    Adachi, T
    Oku, T
    Gorin, A
    Ino, T
    Kuroda, K
    Manuilov, I
    Morimoto, K
    Ryazantsev, A
    Shimizu, HM
    Suzuki, J
    Tokanai, F
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2002, 74 (Suppl 1): : S243 - S245
  • [5] Development of a two-dimensional imaging detector based on a neutron scintillator with wavelength-shifting fibers
    K. Sakai
    T. Adachi
    T. Oku
    A. Gorin
    T. Ino
    K. Kuroda
    I. Manuilov
    K. Morimoto
    A. Ryazantsev
    H.M. Shimizu
    J. Suzuki
    F. Tokanai
    Applied Physics A, 2002, 74 : s243 - s245
  • [6] Method with Digital Signal Processing for Increasing the Efficiency of Thermal Neutron Detection by a Scintillation Detector Based on a 6LiF ZnS(Ag) Screen and Wavelength-Shifting Fibers
    Podlesny, M. M.
    Kruglov, V. V.
    Bodnarchuk, V. I.
    Milkov, V. M.
    Bogdzel, A. A.
    PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2024, 21 (02) : 146 - 153
  • [7] Development of a multi-channel readout ASIC for a fast neutron spectrometer based on GEM-TPC
    何力
    邓智
    刘以农
    李玉兰
    Chinese Physics C, 2014, (10) : 41 - 45
  • [8] Development of a multi-channel readout ASIC for a fast neutron spectrometer based on GEM-TPC
    He Li
    Deng Zhi
    Liu Yi-Nong
    Li Yu-Lan
    CHINESE PHYSICS C, 2014, 38 (10)
  • [9] Development of a multi-channel readout ASIC for a fast neutron spectrometer based on GEM-TPC
    何力
    邓智
    刘以农
    李玉兰
    Chinese Physics C, 2014, 38 (10) : 41 - 45
  • [10] A Time-of-Flight Neutron Imaging Detector Using ZnS/6LiF Scintillator and Wavelength Shifting Fibers with High Spatial Resolution and with Low Gamma-ray Sensitivity
    Nakamura, T.
    Toh, K.
    Kawasaki, T.
    Honda, K.
    Birumachi, A.
    Ebine, M.
    Sakasai, K.
    Soyama, K.
    Katagiri, M.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,