Impact of setup errors on the robustness of linac-based single-isocenter coplanar and non-coplanar VMAT plans for multiple brain metastases

被引:0
|
作者
Sun, Xiaohuan [1 ]
Guan, Fada [2 ]
Yun, Qinghui [3 ]
Jennings, Matthew [4 ]
Biggs, Simon [5 ]
Wang, Zhongfei [1 ]
Wang, Wei [1 ]
Zhang, Te [1 ]
Shi, Mei [1 ]
Zhao, Lina [1 ]
机构
[1] Air Force Med Univ, Xijing Hosp, Dept Radiat Oncol, Xian, Peoples R China
[2] Yale Univ, Sch Med, Dept Therapeut Radiol, New Haven, CT USA
[3] Air Force Med Univ, Xijing Hosp, Dept Equipment, Xian, Peoples R China
[4] Townsville Univ Hosp, Dept Med Phys, Douglas, Qld, Australia
[5] Radiotherapy AI Pty Ltd, Wagga Wagga, Australia
来源
基金
中国国家自然科学基金;
关键词
coplanar and noncoplanar VMAT plans; multiple brain metastases; patient setup errors; single isocenter; RADIOSURGERY; CONFORMITY; THERAPY; SRS;
D O I
10.1002/acm2.14317
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Patient setup errors have been a primary concern impacting the dose delivery accuracy in radiation therapy. A robust treatment plan might mitigate the effects of patient setup errors. In this reported study, we aimed to evaluate the impact of translational and rotational errors on the robustness of linac-based, single-isocenter, coplanar, and non-coplanar volumetric modulated arc therapy treatment plans for multiple brain metastases. Methods: Fifteen patients were retrospectively selected for this study with a combined total of 49 gross tumor volumes (GTVs). Single-isocenter coplanar and non-coplanar plans were generated first with a prescribed dose of 40 Gy in 5 fractions or 42 Gy in 7 fractions to cover 95% of planning target volume (PTV). Next, four setup errors (+1 and +2 mm translation, and +1 degrees and +2 degrees rotation) were applied individually to generate modified plans. Different plan quality evaluation metrics were compared between coplanar and non-coplanar plans. 3D gamma analysis (3%/2 mm) was performed to compare the modified plans (+2 mm and +2 degrees only) and the original plans. Paired t-test was conducted for statistical analysis. Results: After applying setup errors, variations of all plan evaluation metrics were similar (p > 0.05). The worst case for V100% to GTV was 92.07% +/- 6.13% in the case of +2 mm translational error. 3D gamma pass rates were > 90% for both coplanar (+2 mm and +2 degrees) and the +2 mm non-coplanar groups but was 87.40% +/- 6.89% for the +2 degrees non-coplanar group. Conclusion: Translational errors have a greater impact on PTV and GTV dose coverage for both planning methods. Rotational errors have a greater negative impact on gamma pass rates of non-coplanar plans. Plan evaluation metrics after applying setup errors showed that both coplanar and non-coplanar plans were robust and clinically acceptable.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Cost-effectiveness of Linac-based single-isocenter non-coplanar technique (HyperArcTM) for brain metastases radiosurgery
    Filippo Alongi
    Alba Fiorentino
    Ruggero Ruggieri
    Francesco Ricchetti
    Patrick Kupelian
    [J]. Clinical & Experimental Metastasis, 2018, 35 : 601 - 603
  • [2] Cost-effectiveness of Linac-based single-isocenter non-coplanar technique (HyperArcTM) for brain metastases radiosurgery
    Alongi, Filippo
    Fiorentino, Alba
    Ruggieri, Ruggero
    Ricchetti, Francesco
    Kupelian, Patrick
    [J]. CLINICAL & EXPERIMENTAL METASTASIS, 2018, 35 (07) : 601 - 603
  • [3] Single-Isocenter Linac-Based Radiosurgery for Brain Metastases with Coplanar Arcs: A Dosimetric and Clinical Analysis
    Faccenda, Valeria
    Panizza, Denis
    Pisoni, Valerio
    Trivellato, Sara
    Daniotti, Martina Camilla
    Bianchi, Sofia Paola
    De Ponti, Elena
    Arcangeli, Stefano
    [J]. CANCERS, 2023, 15 (18)
  • [4] Dosimetric Impact of Setup Errors in Single-Isocenter VMAT Radiosurgery for Brain Metastases
    Faccenda, V.
    Panizza, D.
    Trivellato, S.
    Daniotti, M. C.
    Pisoni, V.
    Lucchini, R.
    Arcangeli, S.
    De Ponti, E.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E102 - E103
  • [5] Dosimetric impact of setup errors in single-isocenter VMAT radiosurgery for brain metastases
    Panizza, D.
    Faccenda, V.
    Trivellato, S.
    Caricato, P.
    Pisoni, V.
    Lucchini, R.
    De Ponti, E.
    Arcangeli, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S253 - S253
  • [6] Single-isocenter stereotactic radiosurgery for multiple brain metastases: Impact of patient misalignments on target coverage in non-coplanar treatments
    Eder, Michael Martin
    Reiner, Michael
    Heinz, Christian
    Garny, Sylvia
    Freislederer, Philipp
    Landry, Guillaume
    Niyazi, Maximilian
    Belka, Claus
    Riboldi, Marco
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2022, 32 (03): : 296 - 311
  • [7] Linac-based radiosurgery for multiple brain metastases: Comparison between two mono-isocenter techniques with multiple non-coplanar arcs
    Ruggieri, Ruggero
    Naccarato, Stefania
    Mazzola, Rosario
    Ricchetti, Francesco
    Corradini, Stefanie
    Fiorentino, Alba
    Alongi, Filippo
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 132 : 70 - 78
  • [8] First experience and clinical results using a new non-coplanar mono-isocenter technique (HyperArc) for Linac-based VMAT radiosurgery in brain metastases
    Alongi, Filippo
    Fiorentino, Alba
    Gregucci, Fabiana
    Corradini, Stefanie
    Giaj-Levra, Niccolo
    Romano, Luigi
    Rigo, Michele
    Ricchetti, Francesco
    Beltramello, Alberto
    Lunardi, Gianluigi
    Mazzola, Rosario
    Ruggieri, Ruggero
    [J]. JOURNAL OF CANCER RESEARCH AND CLINICAL ONCOLOGY, 2019, 145 (01) : 193 - 200
  • [9] First experience and clinical results using a new non-coplanar mono-isocenter technique (HyperArc™) for Linac-based VMAT radiosurgery in brain metastases
    Filippo Alongi
    Alba Fiorentino
    Fabiana Gregucci
    Stefanie Corradini
    Niccolo Giaj-Levra
    Luigi Romano
    Michele Rigo
    Francesco Ricchetti
    Alberto Beltramello
    Gianluigi Lunardi
    Rosario Mazzola
    Ruggero Ruggieri
    [J]. Journal of Cancer Research and Clinical Oncology, 2019, 145 : 193 - 200
  • [10] Single-isocenter stereotactic non-coplanar arc treatment of 200 patients with brain metastases: multileaf collimator size and setup uncertainties
    Calmels, Lucie
    Biancardo, Susan Blak Nyrup
    Sibolt, Patrik
    Bekke, Susanne Norring
    Bjelkengren, Ulf
    Wilken, Eva
    Geertsen, Poul
    Sjostrom, David
    Behrens, Claus F.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2022, 198 (05) : 436 - 447