A suggested protocol for value assignment using isotope-dilution-LC-MS/MS in reference measurement procedures - Exemplary application for ciprofloxacin in serum

被引:1
|
作者
Schaffler, Judith [1 ]
Habler, Katharina [1 ]
Vogeser, Michael [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, Univ Hosp, Inst Lab Med, Munich, Germany
关键词
Isotope -dilution liquid chromatography; tandem mass spectrometry (ID-LC-MS/MS); Reference measurement procedures; Metrological traceability; ISO; 15; 193; Standardization;
D O I
10.1016/j.cca.2023.117556
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Reference measurement procedures for assigning values to calibration materials play a crucial role in the concept of metrological traceability in laboratory medicine. ISO standard 15,193 specifies the requirements for such measurement procedures, albeit in a very general terms. Materials and methods: A standard structure of analysis series for value assignment by LC-MS/MS was developed and tested,this structure was complemented by a spreadsheet file for result calculation, metadata evaluation, and finally validation and confirmation of individual analysis runs and individual results, and measurement uncertainty evaluation. This framework was applied to a procedure for the quantification of ciprofloxacin in serum as an example. Results: The approach of a detailed description of the analytical procedures of isotope dilution LC-MS/MS reference measurement methods together with a highly standardized spreadsheet-based method for data processing was found to be practical and efficient. The described measurement procedure for the quantification of ciprofloxacin in serum was found to be fit for purpose. Conclusion: A standardized, detailed procedure for the application of isotope dilution LC-MS/MS in reference measurement procedures can complement the ISO 15193 standard with respect to this particular analytical technique, which is now widely used in the context of metrological traceability in laboratory medicine.
引用
收藏
页数:6
相关论文
共 43 条
  • [1] Isotope-dilution-LC-MS/MS candidate reference measurement procedure for cefepime in human serum
    Schaeffler, Judith
    Vogeser, Michael
    Habler, Katharina
    JOURNAL OF MASS SPECTROMETRY AND ADVANCES IN THE CLINICAL LAB, 2024, 33 : 43 - 48
  • [2] Measurement Procedure for Glucose in Serum Using Liquid Chromatography Isotope Dilution - Tandem Mass Spectrometry (LC-ID-MS/MS)
    Kress, M.
    Meissner, D.
    Kramer, U.
    Spannagl, M.
    Kaiser, P.
    CLINICAL LABORATORY, 2014, 60 (09) : 1509 - 1515
  • [3] An isotope dilution LC-MS/MS-based candidate reference method for the quantification of androstenedione in human serum and plasma
    Gradl, Katrin
    Taibon, Judith
    Singh, Neeraj
    Albrecht, Eva
    Geistanger, Andrea
    Pongratz, Stephan
    Hutzler, Stefan
    Mayer, Magdalena
    Kleinschmidt, Christine
    Geletneky, Christian
    Hofmann, Verena
    Koeppl, Daniel
    Rauh, Manfred
    Kobold, Uwe
    CLINICAL MASS SPECTROMETRY, 2020, 16 : 1 - 10
  • [4] Quantitation of plasma metanephrines using isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS): a candidate reference measurement procedure and its application to evaluating routine ID-LC/MS/MS methods
    Liu, Zhenni
    Liu, Qingxiang
    Deng, Yuhang
    Zhao, Haijian
    Zeng, Jie
    Zhang, Tianjiao
    Zhang, Jiangtao
    Wang, Jing
    Zhou, Weiyan
    Zhang, Chuanbao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2021, 413 (30) : 7509 - 7520
  • [5] Quantitation of plasma metanephrines using isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS): a candidate reference measurement procedure and its application to evaluating routine ID-LC/MS/MS methods
    Zhenni Liu
    Qingxiang Liu
    Yuhang Deng
    Haijian Zhao
    Jie Zeng
    Tianjiao Zhang
    Jiangtao Zhang
    Jing Wang
    Weiyan Zhou
    Chuanbao Zhang
    Analytical and Bioanalytical Chemistry, 2021, 413 : 7509 - 7520
  • [6] Isotope Dilution-LC-MS/MS Reference Method Assessment of Serum Folate Assay Accuracy and Proficiency Testing Consensus Mean
    Blackmore, Sheena
    Pfeiffer, Christine M.
    Lee, Anne
    Fazili, Zia
    Hamilton, Malcolm S.
    CLINICAL CHEMISTRY, 2011, 57 (07) : 986 - 994
  • [7] Improved Stable Isotope Dilution Assay for Dietary Folates Using LC-MS/MS and Its Application to Strawberries
    Striegel, Lisa
    Chebib, Soraya
    Netzel, Michael E.
    Rychlik, Michael
    FRONTIERS IN CHEMISTRY, 2018, 6
  • [8] Clinical application of an optimized reference measurement procedure for serum digoxin using bracketing calibration method by ID-LC-MS/MS
    Yan, Qiaofang
    Huang, Fei
    Wang, Chuyang
    Zhan, Min
    Zhang, Qiaoxuan
    Huang, Di
    Yan, Jun
    Lin, Haibiao
    Huang, Xianzhang
    Han, Liqiao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (29) : 6909 - 6918
  • [9] Target analyte quantification by isotope dilution LC-MS/MS directly referring to internal standard concentrations - validation for serum cortisol measurement
    Maier, Barbara
    Vogeser, Michael
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (04) : 833 - 837
  • [10] Quantitative measurement of dihydrouridine in RNA using isotope dilution liquid chromatography mass spectrometry (LC/MS)
    Dalluge, JJ
    Hashizume, T
    McCloskey, JA
    NUCLEIC ACIDS RESEARCH, 1996, 24 (16) : 3242 - 3245