Accurate and reproducible enumeration of CD4 T cell counts and Hemoglobin levels using a point of care system: Comparison with conventional laboratory based testing systems in a clinical reference laboratory in Cameroon

被引:1
|
作者
Sagnia, Bertrand [1 ]
Mbakop Ghomsi, Fabrice [2 ]
Moudourou, Sylvie [3 ]
Gutierez, Ana [4 ]
Tchadji, Jules [1 ]
Sosso, Samuel Martin [3 ]
Ndjolo, Alexis [1 ]
Colizzi, Vittorio [5 ]
机构
[1] CIRCB, Microbiol & Immunol Lab, Yaounde, Cameroon
[2] Ctr Sante Catholique Nkolondom, Yaounde, Cameroon
[3] CIRCB, Med Anal Lab, Yaounde, Cameroon
[4] Catholic Hosp BIKOP, Mbalmayo, Cameroon
[5] Univ Roma Tor Vergata, Fac Sci, Lab Immunol, Rome, Italy
来源
PLOS ONE | 2024年 / 19卷 / 03期
关键词
HIV; THERAPY;
D O I
10.1371/journal.pone.0297790
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background Measurements of CD4 T cells and hemoglobin (Hb) are conventionally used to determine the immunological state and disease progression for HIV-infected patients. We obtained a small lightweight point-of-care device, the BD FACSPresto (TM) in order to demonstrate its ability to deliver CD4 and Hb analysis in comparison with two larger clinical machines the BDFACSCanto (TM) analyzer and Sysmex XN 1000 haematology analyzer. The advantages of using the POC device include access to HIV patient data in remote and in resource limited settings. Method The analytical performance of the BD FACSPresto (TM), compared with the FACSCanto (TM) II flow cytometer and the Sysmex XN 1000 haematology analyzer was evaluated by testing 241 routine clinical specimens collected in EDTA tubes from patients attending the Immunology and Microbiology laboratory of Chantal BIYA International Reference Centre (Yaounde, Cameroon) between January and May 2016. Results The mean in absolute counts and percentage of CD4 T cells was 606 cells/mL and 25% respectively via the FACSPresto (TM), and 574 cells/mL and 24% respectively via the BD FACSCanto (TM) II. The mean concentration of Hb levels was 11.90 on the Sysmex XN 1000 and 11.45 via the BD FACSPresto (TM), A high correlation (R-2 = 0.95, P < 0.001) of Hb level measurements was noted between the BD FACSPresto (TM) and Sysmex XN 1000 hematology analyzer. Overall, a Bland-Altman plot of the differences between the two methods showed an excellent agreement for absolute and percentage CD4 counts and hemoglobin measurements between POC and conventional methods evaluated here. Furthermore, the study demonstrated the ease of use of the BD FACSPresto (TM) POC technology in remote areas. Conclusion The BD FACPresto (TM) is a suitable tool for CD4 enumeration in resource-limited settings, specifically providing a deployable, reliable POC testing option. The BD FACSPresto (TM) performed appropriately in comparison to the conventional reference standard technologies. The BD FACSPresto (TM), system provides accurate, reliable, precise CD4/%CD4/Hb results on venous blood sampling. The data showed good agreement between the BD FACSPresto (TM), BD FACSCanto (TM) II and Sysmex XN 1000 XN 1000 systems.
引用
收藏
页数:11
相关论文
共 4 条
  • [1] Comparison of point-of-care versus laboratory-based CD4 cell enumeration in HIV-positive pregnant women
    Myer, Landon
    Daskilewicz, Kristen
    McIntyre, James
    Bekker, Linda-Gail
    JOURNAL OF THE INTERNATIONAL AIDS SOCIETY, 2013, 16
  • [2] Accurate CD4 T-cell enumeration and antiretroviral drug toxicity monitoring in primary healthcare clinics using point-of-care testing
    Jani, Ilesh V.
    Sitoe, Nadia E.
    Chongo, Patrina L.
    Alfai, Eunice R.
    Quevedo, Jorge I.
    Tobaiwa, Ocean
    Lehe, Jonathan D.
    Peter, Trevor F.
    AIDS, 2011, 25 (06) : 807 - 812
  • [3] Rapid enumeration of CD4 + T lymphocytes using an integrated microfluidic system based on Chemiluminescence image detection at point-of-care testing
    Xianbo Qiu
    Shuo Yang
    Di Wu
    Dong Wang
    Shan Qiao
    Shengxiang Ge
    Ningshao Xia
    Duli Yu
    Shizhi Qian
    Biomedical Microdevices, 2018, 20
  • [4] Rapid enumeration of CD4+T lymphocytes using an integrated microfluidic system based on Chemiluminescence image detection at point-of-care testing
    Qiu, Xianbo
    Yang, Shuo
    Wu, Di
    Wang, Dong
    Qiao, Shan
    Ge, Shengxiang
    Xia, Ningshao
    Yu, Duli
    Qian, Shizhi
    BIOMEDICAL MICRODEVICES, 2018, 20 (01)