Washable and Flexible Screen-Printed Ag/AgCl Electrode on Textiles for ECG Monitoring

被引:3
|
作者
Tu, Huating [1 ,2 ]
Li, Xiaoou [1 ]
Lin, Xiangde [1 ]
Lang, Chenhong [2 ]
Gao, Yang [3 ,4 ]
机构
[1] Shanghai Univ Med & Hlth Sci, Coll Med Instruments, Shanghai 201318, Peoples R China
[2] Zhejiang Sci Tech Univ, Key Lab Intelligent Text & Flexible Interconne, Hangzhou 310018, Peoples R China
[3] East China Univ Sci & Technol, Sch Mech & Power Engn, Shanghai 200237, Peoples R China
[4] Wuhan Textile Univ, State Key Lab New Text Mat & Adv Proc Technol, Wuhan 430200, Peoples R China
关键词
textiles electrode; screen printing; Ag/AgCl conductive ink; electrocardiography; DRY ELECTRODES;
D O I
10.3390/polym15183665
中图分类号
O63 [高分子化学(高聚物)];
学科分类号
070305 ; 080501 ; 081704 ;
摘要
Electrocardiogram (ECG) electrodes are important sensors for detecting heart disease whose performance determines the validity and accuracy of the collected original ECG signals. Due to the large drawbacks (e.g., allergy, shelf life) of traditional commercial gel electrodes, textile electrodes receive widespread attention for their excellent comfortability and breathability. This work demonstrated a dry electrode for ECG monitoring fabricated by screen printing silver/silver chloride (Ag/AgCl) conductive ink on ordinary polyester fabric. The results show that the screen-printed textile electrodes have good and stable electrical and electrochemical properties and excellent ECG signal acquisition performance. Furthermore, the resistance of the screen-printed textile electrode is maintained within 0.5 Omega/cm after 5000 bending cycles or 20 washing and drying cycles, exhibiting excellent flexibility and durability. This research provides favorable support for the design and preparation of flexible and wearable electrophysiological sensing platforms.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Washable and flexible screen printed graphene electrode on textiles for wearable healthcare monitoring
    Xu, Xiaowen
    Luo, Meng
    He, Pei
    Yang, Junliang
    [J]. JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2020, 53 (12)
  • [2] Screen-printed fabric electrode for electrocardiogram monitoring on commercial textiles
    Jin, Jiarui
    Jia, Wang
    Yang, Xuemei
    [J]. JOURNAL OF ENGINEERED FIBERS AND FABRICS, 2023, 18
  • [3] Fabrication and Characterization of Planar Screen-Printed Ag/AgCl Reference Electrode for Disposable Sensor Strip
    Idegami, Koutarou
    Chikae, Miyuki
    Nagatani, Naoki
    Tamiya, Eiichi
    Takamura, Yuzuru
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2010, 49 (09)
  • [4] A screen-printed Ag/AgCl reference electrode with long-term stability for electroanalytical applications
    Dawkins, Rebecca C.
    Wen, Dingchen
    Hart, Judy N.
    Vepsalainen, Mikko
    [J]. ELECTROCHIMICA ACTA, 2021, 393
  • [5] Fabrication of a Simple and Cheap Screen-Printed Silver/Silver Chloride (Ag/AgCl) Quasi-Reference Electrode
    Ferreira Oliveira, Ana Elisa
    Pereira, Arnaldo Cesar
    Campos de Resende, Mayra Asevedo
    Ferreira, Lucas Franco
    [J]. ELECTROANALYSIS, 2022, 34 (05) : 809 - 819
  • [6] Design and Characterization of Screen-Printed Textile Electrodes for ECG Monitoring
    Achilli, Andrea
    Bonfiglio, Annalisa
    Pani, Danilo
    [J]. IEEE SENSORS JOURNAL, 2018, 18 (10) : 4097 - 4107
  • [7] Stretchable and Washable Electroluminescent Display Screen-Printed on Textile
    Janczak, Daniel
    Zych, Marcin
    Raczynski, Tomasz
    Dybowska-Sarapuk, Lucja
    Peplowski, Andrzej
    Krzeminski, Jakub
    Sosna-Glebska, Aleksandra
    Znajdek, Katarzyna
    Sibinski, Maciej
    Jakubowska, Malgorzata
    [J]. NANOMATERIALS, 2019, 9 (09)
  • [8] Stability and Efficiency of Screen-Printed Wearable and Washable Antennas
    Scarpello, Maria Lucia
    Kazani, Ilda
    Hertleer, Carla
    Rogier, Hendrik
    Vande Ginste, Dries
    [J]. IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2012, 11 : 838 - 841
  • [9] Effect of Sweating on Electrode-Skin Contact Impedances and Artifacts in EEG Recordings With Various Screen-Printed Ag/Agcl Electrodes
    Kalevo, Laura
    Miettinen, Tomi
    Leino, Akseli
    Kainulainen, Samu
    Korkalainen, Henri
    Myllymaa, Katja
    Toyras, Juha
    Leppanen, Timo
    Laitinen, Tiina
    Myllymaa, Sami
    [J]. IEEE ACCESS, 2020, 8 : 50934 - 50943
  • [10] Fabrication of a planar-form screen-printed solid electrolyte modified Ag/AgCl reference electrode for application in a potentiometric biosensor
    Liao, Wei-Yin
    Chou, Tse-Chuan
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (12) : 4219 - 4223