Blood Glucose Regulation Models in Artificial Pancreas for Type-1 Diabetic Patients

被引:1
|
作者
Chandrasekhar, Abishek [1 ]
Padhi, Radhakant [1 ]
机构
[1] Indian Inst Sci, Ctr Cyber Phys Syst, Bengaluru 560012, India
关键词
Artificial pancreas; Minimal models; Physiological modeling; Type-1 diabetes mellitus; Biomedical models; INSULIN; EXERCISE; MELLITUS; GLUCAGON; ADULTS;
D O I
10.1007/s41745-023-00362-z
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Development, validation, and testing of algorithms for artificial pancreas (AP) systems require mathematical models for the glucose-insulin dynamics inside the body. These physiological models have been extensively studied over the past decades. Two broad types of models are available in diabetic research, each with its own unique purpose: (i) minimal models, which are relatively simple but still manages to capture the macroscopic behavior of the glucose-insulin dynamics of the body, and (ii) high-fidelity models, which are complex and precisely describe the internal dynamics of the glucose-insulin interaction in the body. The minimal models are primarily utilized for control algorithm synthesis, whereas the high-fidelity models are used as platforms for testing and validating AP systems. The most well-known variants of these physiological models are discussed in detail. In addition to these systems, data-driven models such as the auto-regressive moving average with exogenous inputs (ARMAX) models are also used widely in control algorithm synthesis for AP systems. High-fidelity models are utilized for simulating virtual diabetic patients for in silico testing and validation of artificial pancreas systems. Two currently available high-fidelity models are reviewed in this paper for completeness, including the Type-1 diabetes mellitus (T1DM) simulator approved by the food and drug administration of USA. Models accounting for exercise and also glucagon infusion (for dual-hormone AP systems) are also included, which are essential in developing control algorithms with better autonomy and minimal risk.
引用
收藏
页码:353 / 364
页数:12
相关论文
共 50 条
  • [1] Blood Glucose Regulation Models in Artificial Pancreas for Type-1 Diabetic Patients
    Abishek Chandrasekhar
    Radhakant Padhi
    Journal of the Indian Institute of Science, 2023, 103 : 353 - 364
  • [2] Type-1 Fuzzy Controller for Blood Glucose Regulation in Type-1 Diabetic Patients
    Thakur, Priyanka
    Pillay, Yrithu T.
    Watkins, John
    Sawan, Edwin
    SOUTHEASTCON 2024, 2024, : 1205 - 1209
  • [3] Performance Analysis of Fuzzy-PID Controller for Blood Glucose Regulation in Type-1 Diabetic Patients
    Jyoti Yadav
    Asha Rani
    Vijander Singh
    Journal of Medical Systems, 2016, 40
  • [4] Performance Analysis of Fuzzy-PID Controller for Blood Glucose Regulation in Type-1 Diabetic Patients
    Yadav, Jyoti
    Rani, Asha
    Singh, Vijander
    JOURNAL OF MEDICAL SYSTEMS, 2016, 40 (12)
  • [5] An artificial pancreas for automated blood glucose control in patients with Type 1 diabetes
    Schmidt, Signe
    Boiroux, Dimitri
    Ranjan, Ajenthen
    Jorgensen, John Bagterp
    Madsen, Henrik
    Norgaard, Kirsten
    THERAPEUTIC DELIVERY, 2015, 6 (05) : 609 - 619
  • [6] EFFECTS OF METICULOUS BLOOD-GLUCOSE REGULATION ON TYPE-1 DIABETIC GLOMERULOSCLEROSIS (DGS)
    ULDALL, PR
    GORMAN, CK
    FROM, G
    JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 1984, 20 (6B): : 1496 - 1496
  • [7] Blood Glucose Regulation in Type-1 Diabetic Patients using Sliding Mode Control Based on Nonlinear Transformation
    Xavier, Joanofarc
    Divya, Noonela
    Meena, Krithiga
    Patnaile, S. K.
    Panda, R. C.
    IFAC PAPERSONLINE, 2022, 55 (01): : 902 - 907
  • [8] ARTIFICIAL ENDOCRINE PANCREAS AND OPTIMAL BLOOD-GLUCOSE REGULATION IN DIABETIC-PATIENTS - FROM BEDSIDE-TYPE TO WEARABLE-TYPE
    SHICHIRI, M
    KISHIKAWA, H
    SAKAKIDA, M
    KAJIWARA, K
    HASHIGUCHI, Y
    NISHIDA, K
    UEMURA, T
    KONNO, Y
    ICHINOSE, K
    DIABETES RESEARCH AND CLINICAL PRACTICE, 1994, 24 : S251 - S259
  • [9] Robust controller for artificial pancreas for patients with type-1 diabetes
    Mandal S.
    Sutradhar A.
    Research on Biomedical Engineering, 2023, 39 (02) : 437 - 450
  • [10] INSULIN INSENSITIVITY AND GLUCOSE EFFECTIVENESS IN TYPE-1 DIABETIC ALLOGRAFT PANCREAS RECIPIENTS
    OSEI, K
    COTTRELL, DA
    HENRY, ML
    FERGUSON, RM
    TESI, RJ
    ODORISIO, TM
    CLINICAL RESEARCH, 1991, 39 (02): : A246 - A246