Near infrared spectroscopy for determination of moisture content in lyophilized formulation

被引:1
|
作者
Khanolkar, Aruna [1 ,2 ]
Pawale, Pranita [1 ]
Thorat, Viraj [1 ]
Patil, Bhaskar [1 ]
Samanta, Gautam [1 ]
机构
[1] Cipla Ltd, QbD Dept, Integrated Prod Dev, Mumbai, India
[2] Cipla Ltd, QbD Dept, Integrated Prod Dev, LBS Rd,Vikhroli West, Mumbai 400083, India
关键词
Lyophilized; moisture content; near infrared; principal component analysis; partial least square; IN-LINE; RESIDUAL MOISTURE; DRUG SUBSTANCES; DRYING PROCESS; PROTEIN FORMULATIONS; GRANULE MOISTURE; BED GRANULATION; WATER-CONTENT; END-POINT; STABILITY;
D O I
10.1177/09670335241240309
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
A non-invasive near infrared (NIR) spectroscopic method was developed for the quantitative moisture determination in a lyophilized injection formulation. The calibration samples were prepared by exposing lyophilized samples at different temperatures and relative humidity. The samples from different scales and different process parameters were considered for adding robustness to the model. The NIR spectra were collected using a Fourier- transform (FT) NIR with a diffuse reflectance probe and the same samples were further analyzed by the Karl Fisher (KF) method for moisture content. The pre-treated NIR spectra were used for quantitative method development for moisture content. Partial least squares (PLS) regression was used to develop calibrations in the 5600-4950 cm-1 region with calibration coefficient of determination (R2) of 0.96 and root mean square error of calibration (RMSEC) of 0.149. The model was cross-validated internally using the Kernel algorithm with r2 = 0.96 and RMSECV = 0.15. The accuracy of the NIR method against the KF method, precision, and reproducibility were good and the model was robust in predicting different external validation samples. This work allowed NIR as an alternative measurement for moisture analysis as well as facilitate 100% monitoring before packaging and save the cost of sample and time of KF analysis.
引用
收藏
页码:18 / 28
页数:11
相关论文
共 50 条
  • [2] Determination of moisture content of lyophilized allergen vaccines by NIR spectroscopy
    Zheng, Yiwu
    Lai, Xuxin
    Bruun, Susanne Wrang
    Ipsen, Henrik
    Larsen, Jorgen Nedergaard
    Lowenstein, Henning
    Sondergaard, Ib
    Jacobsen, Susanne
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2008, 46 (03) : 592 - 596
  • [3] Determination of adequate moisture content for efficient dry-heat viral inactivation in lyophilized factor VIII by loss on drying and by near infrared spectroscopy
    Savage, M
    Torres, J
    Franks, L
    Masecar, B
    Hotta, J
    [J]. BIOLOGICALS, 1998, 26 (02) : 119 - 124
  • [4] Moisture content determination of pharmaceutical pellets by near infrared spectroscopy: Method development and validation
    Mantanus, J.
    Ziemons, E.
    Lebrun, P.
    Rozet, E.
    Klinkenberg, R.
    Streel, B.
    Evrard, B.
    Hubert, Ph.
    [J]. ANALYTICA CHIMICA ACTA, 2009, 642 (1-2) : 186 - 192
  • [5] Global regression model for moisture content determination using near-infrared spectroscopy
    Clavaud, Matthieu
    Roggo, Yves
    Degardin, Klara
    Sacre, Pierre-Yves
    Hubert, Philippe
    Ziemons, Eric
    [J]. EUROPEAN JOURNAL OF PHARMACEUTICS AND BIOPHARMACEUTICS, 2017, 119 : 343 - 352
  • [6] Determination of optical parameters and moisture content of wood with visible-near infrared spectroscopy
    Haddadi, Ataollah
    Hans, Guillaume
    Leblon, Brigitte
    Pirouz, Zarin
    Tsuchikawa, Satoru
    Nader, Joseph
    Groves, Kevin
    [J]. JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2016, 24 (06) : 571 - 585
  • [7] Determination of moisture in chestnuts using near infrared spectroscopy
    Liu, Jie
    Li, Xiaoyu
    Li, Peiwu
    Wang, Wei
    Zhou, Wei
    Zhang, Jun
    Zhang, Rongrong
    Xiao, Wu
    [J]. Nongye Gongcheng Xuebao/Transactions of the Chinese Society of Agricultural Engineering, 2010, 26 (02): : 338 - 341
  • [8] Determination of residual moisture in lyophilized protein pharmaceuticals using a rapid and non-invasive method: Near infrared spectroscopy
    Lin, TP
    Hsu, CC
    [J]. PDA JOURNAL OF PHARMACEUTICAL SCIENCE AND TECHNOLOGY, 2002, 56 (04) : 196 - 205
  • [9] Prediction of Veneer Moisture Content Based on Near Infrared Spectroscopy
    Zhou, Kang
    Chen, Yutang
    Sun, Chengshuo
    Na, Bin
    [J]. BIORESOURCES, 2022, 17 (04): : 5878 - 5889
  • [10] Determination of resin and moisture content in melamine-formaldehyde paper using near infrared spectroscopy
    Henriques, Ana
    Goncalves, Miguel
    Paiva, Nadia
    Ferra, Joao
    Martins, Jorge
    Magalhaes, FernaaoD
    Carvalho, Luisa H.
    [J]. JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2017, 25 (05) : 311 - 323