Effects of spatial and temporal resolution on cardiovascular magnetic resonance feature tracking measurements using a simple realistic numerical phantom

被引:1
|
作者
Adams, David M. [1 ]
Boubertakh, Redha [2 ]
Miquel, Marc Eric [1 ,3 ,4 ]
机构
[1] Barts Hlth NHS Trust, Clin Phys, London, England
[2] Natl Heart Res Inst Singapore NHRIS, 5 Hosp Dr, Singapore, Singapore
[3] Queen Mary Univ London, William Res Inst, Ctr Adv Cardiovasc Imaging, NIHR Barts Biomed Res Ctr BRC, Charterhouse Sq, London, England
[4] Queen Mary Univ London, Digital Environm Res Inst DERI, London, England
来源
BRITISH JOURNAL OF RADIOLOGY | 2023年 / 96卷 / 1143期
关键词
SPECKLE-TRACKING; CARDIAC IMAGES; GENERATION; FRAMEWORK; ULTRASOUND; SEQUENCES; MODEL;
D O I
10.1259/bjr.20220233
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives To develop a single -slice numerical phantom with known myocardial motion, at several temporal and in -plane spatial resolutions, for testing and comparison of Cardiovascular Magnetic Resonance (CMR) feature tracking (FT) software.Methods The phantom was developed based on CMR acquisitions of one volunteer (acquired cine, tagging cine, T1 map, T2 map, proton density weighted image). The numerical MRI simulator JEMRIS was used, and the phantom was generated at several in -plane spatial reso-lutions (1.4 x 1.4 mm(2) to 3.0 x 3.0 mm(2)) and temporal resolutions (20 to 40 cardiac phases). Two feature tracking software packages were tested: Medical Image Tracking Toolbox (MITT) and two versions of cvi42 (v5.3.8 and v5.13.7). The effect of resolution on strain results was investigated with reference to ground -truth radial and circumferential strain. Results Peak radial strain was consistently undermeas-ured more for cvi42 v5.13.7 than for v5.3.8. Increased pixel size produced a trend of increased difference from ground -truth peak strain, with the largest changes for cvi42 obtained using v5.13.7 between 1.4 x 1.4 mm(2) and 3.0 x 3.0 mm(2), at 9.17 percentage points (radial) and 8.42 percentage points (circumferential).Conclusions The results corroborate the presence of intervendor differences in feature tracking results and show the magnitude of strain differences between soft-ware versions.Advances in knowledge This study shows how temporal and in -plane spatial resolution can affect feature tracking with reference to the ground -truth strain of a numer-ical phantom. Results reaffirm the need for numerical phantom development for the validation and testing of FT software.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Impact of temporal and spatial resolution on atrial feature tracking cardiovascular magnetic resonance imaging
    Schmidt-Rimpler, Jonas
    Backhaus, Soeren J.
    Hartmann, Finn P.
    Schaten, Philip
    Lange, Torben
    Evertz, Ruben
    Schulz, Alexander
    Kowallick, Johannes T.
    Lapinskas, Tomas
    Hasenfuss, Gerd
    Kelle, Sebastian
    Schuster, Andreas
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2024, 396
  • [2] Defining the optimal temporal and spatial resolution for cardiovascular magnetic resonance imaging feature tracking
    Backhaus, Soren J.
    Metschies, Georg
    Billing, Marcus
    Schmidt-Rimpler, Jonas
    Kowallick, Johannes T.
    Gertz, Roman J.
    Lapinskas, Tomas
    Pieske-Kraigher, Elisabeth
    Pieske, Burkert
    Lotz, Joachim
    Bigalke, Boris
    Kutty, Shelby
    Hasenfuss, Gerd
    Kelle, Sebastian
    Schuster, Andreas
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2021, 23 (01)
  • [3] Defining the optimal temporal and spatial resolution for cardiovascular magnetic resonance imaging feature tracking
    Sören J. Backhaus
    Georg Metschies
    Marcus Billing
    Jonas Schmidt-Rimpler
    Johannes T. Kowallick
    Roman J. Gertz
    Tomas Lapinskas
    Elisabeth Pieske-Kraigher
    Burkert Pieske
    Joachim Lotz
    Boris Bigalke
    Shelby Kutty
    Gerd Hasenfuß
    Sebastian Kelle
    Andreas Schuster
    Journal of Cardiovascular Magnetic Resonance, 23
  • [4] The cure: demonstrating temporal resolution of mechanical dyssynchrony in patients with takotsubo syndrome using cardiovascular magnetic resonance myocardial feature-tracking
    Lange, T.
    Stiermaier, T.
    Lamata, P.
    Raaz, U.
    Kowallick, J. T.
    Lotz, J.
    Hasenfuss, G.
    Thiele, H.
    Eitel, I.
    Schuster, A.
    EUROPEAN HEART JOURNAL, 2017, 38 : 1088 - 1089
  • [5] Feature tracking compared with tissue tagging measurements of segmental strain by cardiovascular magnetic resonance
    Wu, LiNa
    Germans, Tjeerd
    Guclu, Ahmet
    Heymans, Martijn W.
    Allaart, Cornelis P.
    van Rossum, Albert C.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2014, 16
  • [6] Feature tracking compared with tissue tagging measurements of segmental strain by cardiovascular magnetic resonance
    LiNa Wu
    Tjeerd Germans
    Ahmet Güçlü
    Martijn W Heymans
    Cornelis P Allaart
    Albert C van Rossum
    Journal of Cardiovascular Magnetic Resonance, 16
  • [7] Realistic numerical simulations of diffusion tensor cardiovascular magnetic resonance: The effects of perfusion and membrane permeability
    Alemany, Ignasi
    Rose, Jan N.
    Ferreira, Pedro F.
    Pennell, Dudley J.
    Nielles-Vallespin, Sonia
    Scott, Andrew D.
    Doorly, Denis J.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (04) : 1641 - 1656
  • [8] Myocardial deformation assessment using cardiovascular magnetic resonance-feature tracking technique
    Almutairi, Haifa M.
    Boubertakh, Redha
    Miquel, Marc E.
    Petersen, Steffen E.
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1080):
  • [9] The effects of extracellular contrast agent (Gadobutrol) on the precision and reproducibility of cardiovascular magnetic resonance feature tracking
    Kuetting, Daniel L. R.
    Dabir, Darius
    Homsi, Rami
    Sprinkart, Alois M.
    Luetkens, Julian
    Schild, Hans H.
    Thomas, Daniel K.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2016, 18
  • [10] The effects of extracellular contrast agent (Gadobutrol) on the precision and reproducibility of cardiovascular magnetic resonance feature tracking
    Daniel L. R. Kuetting
    Darius Dabir
    Rami Homsi
    Alois M. Sprinkart
    Julian Luetkens
    Hans H. Schild
    Daniel K. Thomas
    Journal of Cardiovascular Magnetic Resonance, 18