Pseudoaneurysm of aorta after transcatheter closure of patent ductus arteriosus

被引:0
|
作者
Sharma, Anuj [1 ]
Arya, Adhi [1 ]
Wadile, Santosh [2 ]
Borgohain, Shyamadeep [3 ]
Kulkarni, Snehal [2 ]
机构
[1] Sri Sathya Sai Sanjeevani Int Ctr Child Heart Care, Dept Pediat Cardiol, NH-2,Delhi Mathura Rd, Baghola 121102, Haryana, India
[2] Sri Sathya Sai Sanjeevani Ctr Child Heart Care & T, Dept Pediat Cardiol, Navi Mumbai 410210, Maharashtra, India
[3] Sri Sathya Sai Sanjeevani Ctr Child Heart Care & T, Dept Pediat Cardiac Surg, Navi Mumbai 410210, Maharashtra, India
关键词
Patent ductus arteriosus; Transcatheter device closure; Aortic aneurysm; Device retrieval;
D O I
10.1007/s12055-023-01619-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aortic pseudoaneurysm is a rare but life-threatening complication after transcatheter closure of patent ductus arteriosus (PDA) with only a few reported cases in literature. We report a case of an 18-month-old child who was diagnosed to have pseudoaneurysm of the aorta after 6 months of transcatheter closure of PDA. It was diagnosed on routine follow-up on echocardiography evaluation. Computed tomography chest is the imaging modality of choice for evaluation of these patients which showed large saccular aneurysm arising from distal aortic arch beyond origin of left subclavian artery. Any aneurysm more than 30 mm should be treated irrespective of symptoms. Treatment modalities are surgical resection of the aneurysm (as in our patient) or thoracic endovascular aortic repair. In our case, surgical resection of aneurysm was performed.
引用
收藏
页码:103 / 106
页数:4
相关论文
共 50 条
  • [1] Pseudoaneurysm of aorta after transcatheter closure of patent ductus arteriosus
    Anuj Sharma
    Adhi Arya
    Santosh Wadile
    Shyamadeep Borgohain
    Snehal Kulkarni
    Indian Journal of Thoracic and Cardiovascular Surgery, 2024, 40 : 103 - 106
  • [2] Complications after transcatheter closure of patent ductus arteriosus
    Jang, Gi Young
    Son, Chang Sung
    Lee, Joo Won
    Lee, Jae Young
    Kim, Soo Jin
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2007, 22 (03) : 484 - 490
  • [3] Transcatheter closure of patent ductus arteriosus
    Alvarez J, Alejandro
    Carrillo R, Vanessa
    Sanhueza H, Eugenio
    Nazal L, Rodrigo
    REVISTA CHILENA DE PEDIATRIA-CHILE, 2012, 83 (06): : 577 - 581
  • [4] Transcatheter closure of patent ductus arteriosus
    Hess, J
    EUROPEAN HEART JOURNAL, 1997, 18 (03) : 359 - 360
  • [5] Transcatheter closure of the patent ductus arteriosus
    Grifka, RG
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 61 (04) : 554 - 570
  • [6] Transcatheter closure of patent ductus arteriosus and ductus in children
    Irdem, Ahmet
    Kervancioglu, Mehmet
    Baspinar, Osman
    Kilinc, Metin
    EUROPEAN JOURNAL OF THERAPEUTICS, 2012, 18 (03): : 132 - 138
  • [7] Transcatheter closure of residual patent ductus arteriosus
    El-Saiedi, Sonia Ali
    Elshedoudy, Sahar Abdulla
    El-Sisi, Ammal Mahmoud
    Hanna, Baher Matta
    Fattouh, Aya Mohammed
    Hijazi, Ziyad
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 95 (01) : 78 - 82
  • [8] Successful stent for a pseudoaneurysm after the patent ductus arteriosus closure procedure
    Lu, Hua
    Yan, Tao
    Wang, Jun
    Zhang, Weida
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2013, 17 (06) : 1038 - 1039
  • [9] Transcatheter Device Closure of Patent Ductus Arteriosus
    Sultan, Mehboob
    Ullah, Maad
    Sadiq, Nadeem
    Akhtar, Khurram
    Akbar, Hajira
    JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2014, 24 (10): : 710 - 713
  • [10] Hybrid retrieval of embolized device in abdominal aorta after transcatheter closure of large patent ductus arteriosus
    Soliman, Mosaad
    Mowaphy, Khaled
    Elsaadany, Nshaat A.
    Soliman, Reem
    JOURNAL OF VASCULAR SURGERY CASES INNOVATIONS AND TECHNIQUES, 2021, 7 (01): : 56 - 60