A new digitised screen test for strabismus measurement

被引:2
|
作者
Oltrup, Theo [1 ,2 ]
Bende, Marvin [1 ]
Henseling, Celine [1 ]
Bende, Thomas [1 ]
Leitritz, Martin A. [1 ]
Bartz-Schmidt, Karl Ulrich [1 ]
机构
[1] Univ Tubingen, Dept Ophthalmol, Div Exp Ophthalm Surg & Refract Surg, Tubingen, Germany
[2] Dept Augenheilkunde Tubingen, Sekt Experimentelle Ophthalmochirurg & Refrakt Chi, Schleichstr 12-1, D-72076 Tubingen, Germany
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2023年 / 33卷 / 02期
关键词
Ocular motility analysis; Strabismus; Digital tangent screen test; Measurement uncertainty; DISTURBANCES;
D O I
10.1016/j.zemedi.2022.07.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Our study presents a digitised tangent screen test for ocular motility analysis according to the Harms and Hess tests (measurement of the squint angle in all fields of vision). This test uses an image beamer to display the tangent screen, a position sensor to measure the patient's head orientation, and a distance sensor to measure the fixation distance. Digital measurement of head orientation allows for a test procedure that eliminates the conventional requirement for a light pointer in the patient's hand. Thus, the digital screen test is presented, and the uncertainty of the measurement system is evaluated. Methods: A mathematical relationship was given between the measured squint angles, as well as the angle of diagnostic gaze direction, and the influence quantities on their measurement uncertainty. The individual uncertainties resulted from deviations in the measured values by the position and distance sensors, the calibration of the projection image of the beamer in length units, and the finite image resolution of the beamer. The individual standard uncertainties of the influence quantities were determined. The combined standard measurement uncertainties of the squint and gaze direction angles were given based on the model equation of the error propagation law at the tangent table according to Harms at a test distance of 2.5 m. The patient's uncertainty contribution to the mobility analysis was not considered. Results: The combined standard uncertainty of the measurement system (coverage factor k = 2 for 95% confidence level) for the squint angle is <= 0.43 degrees for the angle of diagnostic gaze direction <= 3.13 degrees at the test distance of 2.5 m. The individual standard uncertainties of the influence quantities on the angles are (k = 1): 1.55 degrees/1.01 degrees (horizontal/vertical angle of the position sensor), 0.19 degrees (distance sensor), 0.06 degrees (calibration of the projection image of the beamer), and 0.02 degrees (image resolution of the beamer). The maximum valid test distance of the digital screen test is 3.8 m. Conclusion: The digital screen test is compact and can be used at different locations. Compared to the traditional test, the time required for examination via the digitised test is less; additionally, its documentation is simplified. The measurement uncertainty of the diagnostic gaze direction angle is dominated by the sensor drift of the position sensor in the horizontal direction (yaw angle) and is due to the sensor technology. However, this drift error does not affect the squint angle measurement result nor its measurement uncertainty because the measurement principle used here is based on the congruence between the position cross and the fixation object and the confusion principle and compensates for the drift error. The measurement uncertainties of the determined measurement system are the lower limits of the uncertainties in the clinical use of the digital screen test if there are no effects due to significant patient deviations.
引用
收藏
页码:182 / 191
页数:10
相关论文
共 50 条
  • [1] Digital Hess Screen test : a novel method for quantitative measurement of strabismus
    Al-Mohamedi, Haroun
    Kelly-Perez, Ismael Kelly-Perez
    Prinz, Andreas
    Bende, Marvin
    Oltrup, Theo
    Leitritz, Martin
    Bende, Thomas
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [2] Virtual reality-based Harms tangent screen test for strabismus measurement
    Wenner, Yaroslava
    Schneider, Johann
    Drisser, Timo
    Yang, Yu Yi
    Demeter, Thilo
    Fronius, Maria
    Lorenz, Birgit
    Kohnen, Thomas
    Mueller, Michael
    Triesch, Jochen
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2025,
  • [3] HIRSCHBERG TEST AND MEASUREMENT OF STRABISMUS
    GRIFFIN, J
    BOYER, F
    AMERICAN JOURNAL OF OPTOMETRY AND PHYSIOLOGICAL OPTICS, 1973, 50 (09): : 742 - 742
  • [4] New database of digitised mammograms
    不详
    EUROPEAN JOURNAL OF CANCER, 2003, 39 (02) : 132 - 132
  • [5] Effectiveness and repeatability of eye-tracking-based test in strabismus measurement of children
    Zou, Leilei
    Tian, Tian
    Wygnanski-Jaffe, Tamara
    Yehezkel, Oren
    Wang, Shu
    Moshkovitz, Avital
    Sun, Xinghuai
    Liu, Hong
    Liu, Rui
    SEMINARS IN OPHTHALMOLOGY, 2022, 37 (04) : 502 - 508
  • [6] THE REAGIN SCREEN TEST - A NEW REAGIN CARD TEST FOR SYPHILIS
    MARCH, RW
    STILES, GE
    SEXUALLY TRANSMITTED DISEASES, 1980, 7 (02) : 66 - 70
  • [7] Digitised Justice: The New Two Tiers?
    Jane Loo
    Mark Findlay
    Criminal Law Forum, 2022, 33 : 1 - 38
  • [8] AN EVALUATION OF A NEW TEST KIT TO SCREEN FOR GALACTOSEMIA
    JINKS, DC
    VOLLMER, D
    ORFANOS, A
    GUTHRIE, R
    CLINICAL BIOCHEMISTRY, 1987, 20 (05) : 353 - 357
  • [9] New methods test shaker screen conductance
    Morgan, MC
    OIL & GAS JOURNAL, 2006, 104 (09) : 54 - +
  • [10] The Psychomotor Cognition Test for Measurement of Sleepiness/Fatigue on a Touch Screen
    Lee, Yongkyun
    Lee, Younggun
    Yoo, Seunghoon
    Shin, Seunghwan
    Park, Hyesu
    Kim, Dongsoo
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,