Radiotherapy Infrastructure Shielding Calculations: software development for shielding calculations in a linear accelerator radiotherapy room

被引:0
|
作者
Ganou, Charalampia [1 ]
Ganos, Aristotelis [2 ]
Papatheodorou, Spiridon [3 ]
Metaxas, Vasileios [1 ]
Dimitroukas, Christos [1 ,3 ]
Panayiotakis, George [1 ,3 ]
机构
[1] Univ Patras, Sch Med, Dept Med Phys, Patras 26504, Greece
[2] Univ Patras, Dept Civil Engn, Patras 26504, Greece
[3] Univ Hosp Patras, Dept Med Phys, Patras 26504, Greece
关键词
REQUIREMENTS; MODEL;
D O I
10.1093/rpd/ncad189
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This work aimed to develop and validate software that calculates the shielding thickness required for a radiotherapy room with a linear accelerator utilising geometric and dosimetric data. The software "Radiotherapy Infrastructure Shielding Calculations" (RISC) was developed using MATLAB programming. It does not require the installation of the MATLAB platform, and the user only needs to download and install the application, which displays a graphical user interface (GUI). The GUI includes empty cells to insert numerical values for several parameters to calculate the proper shielding thickness. The GUI comprises two main interfaces, one for the primary and one for the secondary barrier calculation. The interface of the primary barrier is divided into four tabs: (a) primary radiation, (b) patient scattered and leakage radiation, (c) IMRT techniques and (d) the shielding cost calculations. The interface of the secondary barrier includes three tabs: (a) patient scattered and leakage radiation, (b) IMRT techniques and (c) the shielding cost calculations. Each tab consists of two sections: one for input and one for output of the necessary data. The RISC is based on the methods and formulae of the NCRP 151 and calculates the primary and secondary barrier thickness for ordinary concrete with a density of 2.35 g/cm(3) and the cost for a radiotherapy room with a linear accelerator that performs conventional or IMRT techniques. Calculations can be performed for photon energies of 4, 6, 10, 15, 18, 20, 25 and 30 MV of a dual-energy linear accelerator, while instantaneous dose rate (IDR) calculations are also performed. The RISC has been validated using all comparative examples of NCRP 151 and the calculations from shielding reports of the Varian IX linear accelerator at Methodist Hospital of Willowbrook and Elekta Infinity at the University Hospital of Patras. The RISC is accompanied by two text files: (a) "Terminology," extensively describing all parameters, and (b) "User's Manual," providing useful instructions to the user. The RISC is user-friendly, simple, fast and precise, providing accurate shielding calculations and quickly and easily reproducing different shielding scenarios for a radiotherapy room with a linear accelerator. Additionally, it could be used during the educational process of shielding calculations by graduate students or trainee medical physicists. As a future work, the RISC will be updated with new features such as skyshine radiation, door shielding, and other types of machines and shielding materials.
引用
收藏
页码:1357 / 1366
页数:10
相关论文
共 50 条
  • [1] Linear accelerator bunker shielding for stereotactic radiotherapy
    Rijken, James
    Bhat, Madhava
    Crowe, Scott
    Kairn, Tanya
    Trapp, Jamie
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (21):
  • [2] SHIELDING CALCULATIONS FOR A 250 MEV PROTON RADIOTHERAPY FACILITY
    ALLEN, M
    AWSCHALOM, M
    MILLER, DW
    MEDICAL PHYSICS, 1987, 14 (03) : 468 - 468
  • [3] COMPUTER-ASSISTED LINEAR-ACCELERATOR SHIELDING CALCULATIONS
    TRAPPIER, AS
    JOHNSON, LP
    JOURNAL OF THE NATIONAL MEDICAL ASSOCIATION, 1990, 82 (03) : 201 - &
  • [4] SHIELDING CALCULATIONS FOR A 400-MEV ELECTRON ACCELERATOR
    ALSMILLER, RG
    BARISH, J
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1973, 16 (JUN): : 334 - 335
  • [5] A versatile program for the calculation of linear accelerator room shielding
    Hassan, Zeinab El-Taher
    Farag, Nehad M.
    Elshemey, Wael M.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 666 - 677
  • [6] Shielding Tensor Calculations
    Facelli J.C.
    eMagRes, 2011, 2011
  • [7] COMPARISON OF SHIELDING CALCULATIONS
    BRIDGES, JE
    MILLER, DA
    SCHULZ, RB
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 1968, EM10 (01) : 175 - &
  • [8] Installation Shielding in Radiotherapy
    Calandrino, Riccardo
    RADIATION PROTECTION IN MEDICAL PHYSICS, 2009, : 133 - 139
  • [9] Optimized Shielding Calculations for Medical Linear Accelerators (LINACs)
    Muhammad, W.
    Lee, S.
    Hussain, A.
    MEDICAL PHYSICS, 2012, 39 (06) : 3765 - 3765
  • [10] Photoneutron dose levels in a linear accelerator room with lead shielding
    Li, Z
    Klein, E
    Esthappan, J
    Drzymala, R
    Purdy, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1325 - 1325