6-month migration sufficient for evaluation of total knee replacements: a systematic review and meta-analysis

被引:5
|
作者
Puijk, Raymond [1 ]
Puijk, Rowan H. [1 ]
Laende, Elise K. [2 ,3 ,4 ]
Dunbar, Michael J. [3 ,4 ,5 ]
Plevier, Jose W. M. [6 ]
Nolte, Peter A. [1 ,7 ,8 ]
Pijls, Bart G. C. W. [9 ]
机构
[1] Spaarne Gasthuis, Dept Orthopaed, Hoofddorp, Netherlands
[2] Queens Univ, Mech & Mat Engn, Kingston, ON, Canada
[3] Dalhousie Univ, Dept Surg, Div Orthopaed, Halifax, NS, Canada
[4] Nova Scotia Hlth Author, QEII Hlth Sci Ctr, Halifax, NS, Canada
[5] Dalhousie Univ, Sch Biomed Engn, Halifax, NS, Canada
[6] Leiden Univ, Med Ctr, Walaeus Lib, Leiden, Netherlands
[7] Univ Amsterdam, Acad Ctr Dent ACTA, Dept Oral Cell Biol, Amsterdam, Netherlands
[8] Vrije Univ Amsterdam, Amsterdam, Netherlands
[9] Leiden Univ, Med Ctr, Dept Orthopaed, Leiden, Netherlands
关键词
RANDOMIZED CONTROLLED-TRIAL; TIBIAL COMPONENT MIGRATION; PATIENT-SPECIFIC INSTRUMENTATION; 5-YEAR FOLLOW-UP; RADIOSTEREOMETRIC ANALYSIS; ALL-POLYETHYLENE; FIXED-BEARING; CONTACT KINEMATICS; IMPLANT MIGRATION; HIP-ARTHROPLASTY;
D O I
10.2340/17453674.2023.24579
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose - This updated meta-analysis evaluates the migration pattern of the tibial component of primary total knee replacements measured with radiostereometric analysis (RSA). We aimed to evaluate whether 6-month maximum total point motion (MTPM) values could be used instead of 1-year MTPM for RSA threshold testing and to present the pooled migration patterns for different implant designs that can be used as a benchmark.Patients and methods - The search included all published RSA studies on migration patterns of tibial components until 2023. Study groups were classified according to their prosthesis brand, fixation, and insert (PFI). Sub -analyses were performed to compare the mean tibial component migration patterns of different implant variables, stratified according to fixation.Results - 96 studies (43 new studies), including 197 study groups and 4,706 knees, were included. Most migration occurred within the first 6 postoperative months (126 study groups: mean 0.58 mm, 95% confidence interval [CI] 0.50-0.65), followed by minimal migration between 6 and 12 months (197 study groups: mean 0.04 mm, CI 0.03-0.06), irrespective of the fixation method used. Distinct migration patterns were observed among the different fixation methods. No differences were found in migration patterns among cemented components in any of the sub-group analyses conducted. For uncemented implants, trabecular metal surfaced components seemed to migrate less than porous-coated or uncoated componentsConclusion - Based on the small difference between MTPM values at 6 months and 1 year, MTPM at 6 months could be used instead of MTPM at 1 year for RSA threshold testing. The pooled migration patterns can be used as benchmark for evaluation of new implants by defining fixation -specific RSA thresholds when combined with implant survival.
引用
收藏
页码:577 / 587
页数:11
相关论文
共 50 条
  • [1] RSA migration of total knee replacements: A systematic review and meta-analysis
    Pijls, Bart G.
    Plevier, Jose W. M.
    Nelissen, Rob G. H. H.
    ACTA ORTHOPAEDICA, 2018, 89 (03) : 320 - 328
  • [2] Detection of aseptic loosening in total knee replacements: a systematic review and meta-analysis
    Lara Barnsley
    Les Barnsley
    Skeletal Radiology, 2019, 48 : 1565 - 1572
  • [3] Detection of aseptic loosening in total knee replacements: a systematic review and meta-analysis
    Barnsley, Lara
    Barnsley, Les
    SKELETAL RADIOLOGY, 2019, 48 (10) : 1565 - 1572
  • [4] EFFICACY OF SYMPTOMATIC TREATMENTS FOR KNEE OA: A SYSTEMATIC REVIEW AND NETWORK META-ANALYSIS WITH A 6-MONTH TIME-HORIZON
    Beaudart, C.
    Lengele, L.
    Geerinck, A.
    Leclercq, V.
    Rodriguez, D. Sanchez
    Bruyere, O.
    Reginster, J. -Y.
    OSTEOPOROSIS INTERNATIONAL, 2020, 31 (SUPPL 1) : S56 - S56
  • [5] Symptomatic Efficacy of Pharmacological Treatments for Knee Osteoarthritis: A Systematic Review and a Network Meta-Analysis with a 6-Month Time Horizon
    C. Beaudart
    L. Lengelé
    V. Leclercq
    A. Geerinck
    D. Sanchez-Rodriguez
    O. Bruyère
    J. Y. Reginster
    Drugs, 2020, 80 : 1947 - 1959
  • [6] Symptomatic Efficacy of Pharmacological Treatments for Knee Osteoarthritis: A Systematic Review and a Network Meta-Analysis with a 6-Month Time Horizon
    Beaudart, C.
    Lengele, L.
    Leclercq, V.
    Geerinck, A.
    Sanchez-Rodriguez, D.
    Bruyere, O.
    Reginster, J. Y.
    DRUGS, 2020, 80 (18) : 1947 - 1959
  • [7] Osteotomies and Total Knee Arthroplasty: Systematic Review and Meta-Analysis
    Krzysztof, Kulinski
    Trams, Ewa
    Pomianowski, Stanislaw
    Kaminski, Rafal
    LIFE-BASEL, 2022, 12 (08):
  • [8] Impact of Use of Intramedullary and Extramedullary Guides on Tibial Component Geometry in Total Knee Replacements: A Systematic Review and Meta-Analysis
    Feeley, Iain
    Hegarty, Aidan
    Hickey, Anne
    Glynn, Aaron
    JOURNAL OF KNEE SURGERY, 2016, 29 (06) : 487 - 496
  • [9] Vascular Injuries in Total Knee Arthroplasty A Systematic Review and Meta-Analysis
    Sundaram, Kavin
    Udo-Inyang, Inyang
    Mont, Michael A.
    Molloy, Robert
    Higuera-Rueda, Carlos
    Piuzzi, Nicolas S.
    JBJS REVIEWS, 2020, 8 (01)
  • [10] Patellar Resurfacing in Total Knee Arthroplasty: Systematic Review and Meta-Analysis
    Longo, Umile G.
    Ciuffreda, Mauro
    Mannering, Nicholas
    D'Andrea, Valerio
    Cimmino, Matteo
    Denaro, Vincenzo
    JOURNAL OF ARTHROPLASTY, 2018, 33 (02): : 620 - 632