Isokinetic Dynamometer Leg Extensor Peak Torque Measurement: A Time-Delayed Reliability and Score Selection Analysis Study

被引:2
|
作者
Thompson, Brennan J. [1 ,2 ]
Xu, Jennifer [3 ]
机构
[1] Utah State Univ, Dept Kinesiol & Hlth Sci, Logan, UT 84322 USA
[2] Utah State Univ, Sorenson Legacy Fdn Ctr Clin Excellence, Movement Res Clin, Logan, UT 84322 USA
[3] Univ Virginia, Dept Kinesiol, Charlottesville, VA 22904 USA
关键词
Biodex; muscle strength; muscle function; strength testing; peak force; quadriceps; ACCELERATION ENERGY; KNEE; STRENGTH; HEALTHY; BIODEX; POWER; YOUNG; WORK;
D O I
10.3390/jfmk8020062
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
The reliability of isokinetic peak torque (PT) has been reported mostly using a short-term (<similar to 10 day) inter-trial testing time frame. However, many studies and programs utilize a long-term (several weeks to months) inter-trial testing period. Additionally, the methods by which the PT value is selected and reported from a multiple rep testing scheme have not been well investigated for both reliability and PT absolute performance comparisons. The purpose of this study was to investigate the long-term reliability of isokinetic and isometric PT of the leg extensors with an emphasis on the differences among several PT score selection methods. Thirteen men and women (age = 19.5 years) underwent two testing trials separated by 28.8 (+/- 1.8) days. Testing included maximal voluntary contractions of three sets of three reps for two isokinetic contraction conditions of 60 (Isok60) and 240 (Isok240) deg/s velocities, and three sets of one rep of isometric contractions for the leg extensors. The PT score was derived from seven different methods (see text for descriptions). Reliability as assessed from intraclass correlation coefficients (ICCs) varied widely across contraction conditions and PT score selection parameters. The Isok60 velocity overall had lower reliability (ICCs = 0.48-0.81) than Isok240 (0.77-0.87) across the conditions whereas the isometric PT variables showed moderate reliability (0.71-0.73). Overall the set 1 PT score selection parameters were generally lower (p <= 0.05) than those that involved sets two and three. Systematic error (p <= 0.05) was shown for 6 out of the 17 PT selection variables. On a subjective interpretation basis, when taking everything into account the best overall combination of time/trial efficiency, reliability, best/highest PT score parameter, and reduced risk of systematic bias appears to be the PT variable that uses the average of the highest two reps of the first two sets of three reps-i.e., averaging the highest two values of the six total reps from the first two sets.
引用
收藏
页数:12
相关论文
共 2 条
  • [1] Reliability of measurement of knee extensor and flexor muscle peak torque on the KinCom isokinetic dynamometer in individuals with rheumatoid arthritis
    McMeeken, JM
    Story, IH
    Smith, J
    Kent, P
    ISOKINETICS AND EXERCISE SCIENCE, 1998, 7 (02) : 67 - 73
  • [2] Real-time measurement of isometric peak torque and rate of torque development using a novel strength testing device: a validity and reliability study
    Palmer, Ty B.
    Blinch, Jarrod
    Farrow, Ahalee C.
    Agu-Udemba, Chinonye C.
    Mitchell, Ethan A.
    PHYSIOLOGICAL MEASUREMENT, 2020, 41 (11)