Specific absorption rate (SAR) simulations for low-field (< 0.1 T) MRI systems

被引:6
|
作者
Parsa, Javad [1 ,2 ]
Webb, Andrew [1 ]
机构
[1] Leiden Univ, Dept Radiol, CJ Gorter MRI Ctr, Med Ctr, Leiden, Netherlands
[2] Percuros BV, Leiden, Netherlands
基金
欧盟地平线“2020”;
关键词
Low field MRI; Specific absorption rate; Transmit efficiency; Electromagnetic simulations; Point-of-care MRI;
D O I
10.1007/s10334-023-01073-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To simulate the magnetic and electric fields produced by RF coil geometries commonly used at low field. Based on these simulations, the specific absorption rate (SAR) efficiency can be derived to ensure safe operation even when using short RF pulses and high duty cycles. Methods Electromagnetic simulations were performed at four different field strengths between 0.05 and 0.1 T, corresponding to the lower and upper limits of current point-of-care (POC) neuroimaging systems. Transmit magnetic and electric fields, as well as transmit efficiency and SAR efficiency were simulated. The effects of a close-fitting shield on the EM fields were also assessed. SAR calculations were performed as a function of RF pulse length in turbo-spin echo (TSE) sequences. Results Simulations of RF coil characteristics and B-1(+) transmit efficiencies agreed well with corresponding experimentally determined parameters. Overall, the SAR efficiency was, as expected, higher at the lower frequencies studied, and many orders of magnitude greater than at conventional clinical field strengths. The tight-fitting transmit coil results in the highest SAR in the nose and skull, which are not thermally sensitive tissues. The calculated SAR efficiencies showed that only when 180 degrees refocusing pulses of duration similar to 10 ms are used for TSE sequences does SAR need to be carefully considered. Conclusion This work presents a comprehensive overview of the transmit and SAR efficiencies for RF coils used for POC MRI neuroimaging. While SAR is not a problem for conventional sequences, the values derived here should be useful for RF intensive sequences such as T-1 rho, and also demonstrate that if very short RF pulses are required then SAR calculations should be performed.
引用
收藏
页码:429 / 438
页数:10
相关论文
共 50 条
  • [1] Specific absorption rate (SAR) simulations for low-field (< 0.1 T) MRI systems
    Javad Parsa
    Andrew Webb
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2023, 36 : 429 - 438
  • [2] Low-field 0.55 T MRI evaluation of the fetus
    Ponrartana, Skorn
    Nguyen, HaiThuy N.
    Cui, Sophia X.
    Tian, Ye
    Kumar, Prakash
    Wood, John C.
    Nayak, Krishna S.
    PEDIATRIC RADIOLOGY, 2023, 53 (07) : 1469 - 1475
  • [3] Low-field 0.55 T MRI evaluation of the fetus
    Skorn Ponrartana
    HaiThuy N. Nguyen
    Sophia X. Cui
    Ye Tian
    Prakash Kumar
    John C. Wood
    Krishna S. Nayak
    Pediatric Radiology, 2023, 53 : 1469 - 1475
  • [4] Calorimetric Calibration of Specific Absorption Rate (SAR) On a Compact 3T MRI Scanner
    Shu, Y.
    Tao, S.
    Yeo, D.
    Foo, T.
    Weavers, P.
    Trzasko, J.
    Huston, J.
    Bernstein, M.
    MEDICAL PHYSICS, 2017, 44 (06) : 3048 - 3048
  • [5] Modern low-field MRI
    Pogarell, Tobias
    Heiss, Rafael
    Janka, Rolf
    Nagel, Armin M.
    Uder, Michael
    Roemer, Frank W.
    SKELETAL RADIOLOGY, 2024, 53 (09) : 1751 - 1760
  • [6] Low-field MRI pelvimetry
    T. A. Tukeva
    H. J. Aronen
    P. T. Karjalainen
    P. J. Mäkelä
    European Radiology, 1997, 7 : 230 - 234
  • [7] Low-field musculoskeletal MRI
    Ghazinoor, Shaya
    Crues, John V., III
    Crowley, Chris
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (02) : 234 - 244
  • [8] Low-field MRI pelvimetry
    Tukeva, TA
    Aronen, HJ
    Karjalainen, PT
    Makela, PJ
    EUROPEAN RADIOLOGY, 1997, 7 (02) : 230 - 234
  • [9] Fistula in ano: Evaluation with low-field magnetic resonance imaging (0.1 T)
    Madsen, SM
    Myschetzky, PS
    Heldmann, U
    Rasmussen, OO
    Thomsen, HS
    SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1999, 34 (12) : 1253 - 1256
  • [10] LOW-FIELD MRI AND SCAPHOID FRACTURE
    LEPISTO, J
    MATTILA, K
    NIEMINEN, S
    SATTLER, B
    KORMANO, M
    JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 1995, 20B (04): : 539 - 542