Deformation of transcatheter heart valves with mitral valve-in-valve

被引:1
|
作者
Fukui, Miho [1 ,2 ]
Okada, Atsushi [3 ]
Burns, Marcus R. [4 ]
Sato, Hirotomo [3 ]
Thao, Kiahltone R. [3 ]
Wang, Cheng [3 ]
Koike, Hideki [1 ,2 ]
Hamid, Nadira [3 ,4 ]
Enriquez-Sarano, Maurice [3 ]
Lesser, John R. [3 ,4 ]
Cavalcante, Joao L. [1 ,2 ,4 ]
Sorajja, Paul [3 ,4 ]
Bapat, Vinayak N. [3 ,4 ]
机构
[1] Minneapolis Heart Inst Fdn, Cardiovasc Imaging Res Ctr, Minneapolis, MN USA
[2] Minneapolis Heart Inst Fdn, Core Lab, Minneapolis, MN USA
[3] Minneapolis Heart Inst Fdn, Valve Sci Ctr, Minneapolis, MN USA
[4] Abbott NW Hosp, Allina Hlth Minneapolis Heart Inst Minneapolis, Minneapolis, MN USA
关键词
degenerative valve; imaging modalities; mitral valve disease; valve-in-valve; AORTIC-VALVE; REPLACEMENT; FRACTURE; BIOPROSTHESES; DURABILITY; PROSTHESES; THROMBOSIS; IMPACT;
D O I
10.4244/EIJ-D-23-00614
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The use of oversizing in mitral valve-in-valve (MViV) procedures can lead to non-uniform expansion of transcatheter heart valves (THV). This may have implications for THV durability. Aims: The objective of this study was to assess the extent and predictors of THV deformation in MViV procedures. Methods: We examined 33 patients who underwent MViV with SAPIEN prostheses. The extent of THV deformation (deformation index, eccentricity, neosinus volume, asymmetric leaflet expansion and vertical deformation) and hypoattenuating leaflet thickening (HALT) were assessed using cardiac computed tomography (CT), performed prospectively at 30 days post-procedure. For descriptive purposes, the THV deformation index was calculated, with values >1.00 representing a more hourglass shape. Results: Non-uniform underexpansion of THV was common after MViV implantation, with a median expansion area of 74.0% (interquartile range 68.1-84.1) at the narrowest level and a THV deformation index of 1.21 (1.13-1.29), but circularity was maintained with eccentricity ranging from 0.24 to 0.28. The degree of oversizing was a key factor associated with greater underexpansion and a higher deformation index (beta=-0.634; p<0.001; beta=0.594; p<0.001, respectively). Overall, the incidence of HALT on the 30-day postprocedural CT was 27.3% (9 of 33). Most patients (32 of 33) were on anticoagulation therapy, but the prothrombin time and international normalised ratio (PT-INR) at the time of the CT scan was <2.5 in 23 of 32 patients. Among patients with a PT-INR of <2.5, HALT was predominantly observed with a high THV deformation index of >= 1.18. Conclusions: THV deformation, i.e., underexpansion and an hourglass shape, commonly occurs after MViV implantation and is negatively affected by excessive oversizing. Optimising THV expansion during MViV could potentially prevent HALT.
引用
收藏
页码:E937 / +
页数:17
相关论文
共 50 条
  • [1] Transcatheter Mitral Valve Replacement for Mitral Valve-in-Valve, Valve-in-Ring, and Valve-in-MAC Using Balloon-Expandable Transcatheter Heart Valves
    Sticchi, Alessandro
    Reineke, David
    Praz, Fabien
    Windecker, Stephan
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (08) : 873 - 878
  • [2] Transcatheter valve-in-valve implantation for degenerated bioprosthetic aortic and mitral valves
    Gallo, Michele
    Dvir, Danny
    Demertzis, Stefanos
    Pedrazzini, Giovanni
    Berdajs, Denis
    Ferrari, Enrico
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (08) : 749 - 758
  • [3] Transcatheter Double Valve-in-Valve Replacement of Aortic and Mitral Bioprosthetic Valves
    Savoj, Javad
    Iftikhar, Syed
    Burstein, Steven
    Hu, Patrick
    [J]. CARDIOLOGY RESEARCH, 2019, 10 (03) : 193 - 198
  • [4] Transcatheter Valve-in-Valve Implantation for Failed Bioprosthetic Heart Valves
    Webb, John G.
    Wood, David A.
    Ye, Jian
    Gurvitch, Ronen
    Masson, Jean-Bernard
    Rodes-Cabau, Josep
    Osten, Mark
    Horlick, Eric
    Wendler, O.
    Dumont, Eric
    Carere, Ronald G.
    Wijesinghe, Namal
    Nietlispach, Fabian
    Johnson, Mark
    Thompson, Chrisopher R.
    Moss, Robert
    Leipsic, Jonathon
    Munt, Brad
    Lichtenstein, Samuel V.
    Cheung, Anson
    [J]. CIRCULATION, 2010, 121 (16) : 1848 - 1857
  • [5] Transcatheter Mitral Valve-in-Valve Implantation With a New Transcatheter Heart Valve for Bioprosthetic Degeneration
    Lu, Yuntao
    Yang, Ye
    Wang, Wenshuo
    Chen, Jinmiao
    Yin, Minyan
    Huang, Liqi
    Dong, Lili
    Wang, Chunsheng
    Wei, Lai
    [J]. FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 8
  • [6] Valve-in-Valve Transcatheter Aortic Valve Implantation for Degenerated Bioprosthetic Heart Valves
    Eggebrecht, Holger
    Schaefer, Ulrich
    Treede, Hendrik
    Boekstegers, Peter
    Babin-Ebell, Joerg
    Ferrari, Markus
    Moellmann, Helge
    Baumgartner, Helmut
    Carrel, Thierry
    Kahlert, Philipp
    Lange, Philipp
    Walther, Thomas
    Erbel, Raimund
    Mehta, Rajendra H.
    Thielmann, Matthias
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (11) : 1218 - 1227
  • [7] Deformation of Transcatheter Heart Valve Following Valve-in-Valve Transcatheter Aortic Valve Replacement Implications for Hemodynamics
    Fukui, Miho
    Sorajja, Paul
    Cavalcante, Joao L.
    Thao, Kiahltone R.
    Okada, Atsushi
    Sato, Hirotomo
    Wang, Cheng
    Koike, Hideki
    Hamid, Nadira
    Enriquez-Sarano, Maurice
    Lesser, John R.
    Bapat, Vinayak N.
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2023, 16 (05) : 515 - 526
  • [8] Deformation of Transcatheter Heart Valve Following Valve-in-Valve Transcatheter Aortic Valve Replacement: Implications for Hemodynamics
    Fukui, Miho
    Sorajja, Paul
    Enriquez-Sarano, Maurice
    Sato, Hirotomo
    Lesser, John Lesser
    Cavalcante, Joao
    Bapat, Vinayak
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2022, 80 (12) : B212 - B212
  • [9] Failed valve-in-valve transcatheter mitral valve implantation
    Bruschi, Giuseppe
    Botta, Luca
    Fratto, Pasquale
    Martinelli, Luigi
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2014, 45 (04) : E127 - E127
  • [10] Current Indications for Transcatheter Mitral Valve Replacement Using Transcatheter Aortic Valves Valve-in-Valve, Valve-in-Ring, and Valve-in-Mitral Annulus Calcification
    Urena, Marina
    Vahanian, Alec
    Brochet, Eric
    Ducrocq, Gregory
    Iung, Bernard
    Himbert, Dominique