Self-applied somnography: technical feasibility of electroencephalography and electro-oculography signal characteristics in sleep staging of suspected sleep-disordered adults

被引:5
|
作者
Rusanen, Matias [1 ,2 ]
Korkalainen, Henri [1 ,2 ]
Gretarsdottir, Heidur [3 ,4 ]
Siilak, Tiina [3 ]
Olafsdottir, Kristin Anna [3 ]
Toyras, Juha [1 ,5 ,6 ]
Myllymaa, Sami [1 ,2 ]
Arnardottir, Erna Sif [3 ,4 ]
Leppanen, Timo [1 ,2 ,6 ]
Kainulainen, Samu [1 ,2 ]
机构
[1] Univ Eastern Finland, Dept Tech Phys, Kuopio, Finland
[2] Kuopio Univ Hosp, Diagnost Imaging Ctr, Kuopio, Finland
[3] Reykjavik Univ, Sleep Inst, Sch Technol, Reykjavik, Iceland
[4] Landspitali Natl Univ Hosp Iceland, Reykjavik, Iceland
[5] Kuopio Univ Hosp, Sci Serv Ctr, Kuopio, Finland
[6] Univ Queensland, Sch Informat Technol & Elect Engn, Brisbane, Qld, Australia
关键词
home sleep apnea testing; portable diagnostics; sleep-disordered breathing; PORTABLE SLEEP; POLYSOMNOGRAPHY; APNEA; RECORDINGS; DIAGNOSIS;
D O I
10.1111/jsr.13977
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Sleep recordings are increasingly being conducted in patients' homes where patients apply the sensors themselves according to instructions. However, certain sensor types such as cup electrodes used in conventional polysomnography are unfeasible for self-application. To overcome this, self-applied forehead montages with electroencephalography and electro-oculography sensors have been developed. We evaluated the technical feasibility of a self-applied electrode set from Nox Medical (Reykjavik, Iceland) through home sleep recordings of healthy and suspected sleep-disordered adults (n = 174) in the context of sleep staging. Subjects slept with a double setup of conventional type II polysomnography sensors and self-applied forehead sensors. We found that the self-applied electroencephalography and electro-oculography electrodes had acceptable impedance levels but were more prone to losing proper skin-electrode contact than the conventional cup electrodes. Moreover, the forehead electroencephalography signals recorded using the self-applied electrodes expressed lower amplitudes (difference 25.3%-43.9%, p < 0.001) and less absolute power (at 1-40 Hz, p < 0.001) than the polysomnography electroencephalography signals in all sleep stages. However, the signals recorded with the self-applied electroencephalography electrodes expressed more relative power (p < 0.001) at very low frequencies (0.3-1.0 Hz) in all sleep stages. The electro-oculography signals recorded with the self-applied electrodes expressed comparable characteristics with standard electro-oculography. In conclusion, the results support the technical feasibility of the self-applied electroencephalography and electro-oculography for sleep staging in home sleep recordings, after adjustment for amplitude differences, especially for scoring Stage N3 sleep.