摘要
This review article describes some of the recent advances in ultra-trace determination of uranium in various natural water bodies such as drinking water, underground water, seawater, etc. using total reflection x-ray florescence (TXRF) spectrometry. Uranium is one of the most hazardous element having chemical as well as radio-toxicity. However, due to its low concentration in water samples, direct determination is seldom possible in samples with high salt content. Various sample preconcentration and selective extraction techniques for uranium such as solvent extraction, dispersive liquid-liquid micro extraction, solid phase extraction, and membrane based preconcentration methods are summarized in this article. TXRF is a very versatile micro-analytical technique with detection limits in picogram range and having multielemental analysis capability. TXRF determination of uranium using U L & alpha; as the analytical line has many analytical challenges and difficulties, mainly due to the severe interference with Rb and Br, both are commonly available in natural water samples. This article highlights some of these challenges, along with the advantages and disadvantages of the different sample preparation techniques used so far for uranium determination in natural water samples by TXRF.
引用
收藏
页码:326 / 339
页数:14
相关论文
共 50 条
  • [1] Determination of trace elements in uranium oxide by Total Reflection X-ray Fluorescence spectrometry
    Misra, NL
    Mudher, KDS
    Adya, VC
    Rajeswari, B
    Venugopal, V
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (06) : 834 - 840
  • [2] Trace determination of uranium preconcentrated using graphene oxide by total reflection X-ray fluorescence spectrometry
    Takahashi, Hiroaki
    Izumoto, Yukie
    Matsuyama, Tsugufumi
    Yoshii, Hiroshi
    X-RAY SPECTROMETRY, 2019, 48 (05) : 366 - 374
  • [3] Total reflection X-ray fluorescence spectrometric determination of ultra-trace uranium in natural water samples using a dispersive liquid-liquid micro-extraction method
    Shukla, Vikas Kumar
    Dhara, Sangita
    Mishra, N. L.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (08) : 1632 - 1640
  • [4] Determination of some trace elements in mineral spring waters by total reflection X-ray fluorescence spectrometry (TXRF)
    Dogan, M
    Soylak, M
    JOURNAL OF TRACE AND MICROPROBE TECHNIQUES, 2002, 20 (02): : 261 - 268
  • [5] Determination of Arsenic in Water by Total Reflection X-Ray Fluorescence Spectrometry (TXRF) by Capture on a Copper Film
    Bennun, Leonardo
    Parra, Lue Meru Marco
    Greaves, Eduardo D.
    ANALYTICAL LETTERS, 2025, 58 (04) : 680 - 687
  • [6] Uranium determination in seawater by total reflection X-ray fluorescence spectrometry
    Misra, N. L.
    Dhara, S.
    Mudher, K. D. Singh
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (10-11) : 1166 - 1169
  • [7] Determination of Halide Ions in Solution by Total Reflection X-ray Fluorescence (TXRF) Spectrometry
    Vander Hoogerstraete, Tom
    Jamar, Steven
    Wellens, Sil
    Binnemans, Koen
    ANALYTICAL CHEMISTRY, 2014, 86 (03) : 1391 - 1394
  • [8] Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis
    Tiwari, MK
    Gowrishankar, B
    Raghuvanshi, VK
    Nandedkar, RV
    Sawhney, KJS
    BULLETIN OF MATERIALS SCIENCE, 2002, 25 (05) : 435 - 441
  • [9] Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis
    M. K. Tiwari
    B. Gowrishankar
    V. K. Raghuvanshi
    R. V. Nandedkar
    K. J. S. Sawhney
    Bulletin of Materials Science, 2002, 25 : 435 - 441
  • [10] Analytical Possibilities of Total Reflection X-ray Spectrometry (TXRF) for Trace Selenium Determination in Soils
    Margui, E.
    Floor, G. H.
    Hidalgo, M.
    Kregsamer, P.
    Roman-Ross, G.
    Streli, C.
    Queralt, I.
    ANALYTICAL CHEMISTRY, 2010, 82 (18) : 7744 - 7751