Accuracy of the estimations of respiratory mechanics using an expiratory time constant in passive and active breathing conditions: a bench study

被引:0
|
作者
Chen, Yuqing [1 ]
Yuan, Yueyang [2 ]
Zhang, Hai [1 ]
Li, Feng [1 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Chest Hosp, Dept Resp Med, Shanghai 200030, Peoples R China
[2] Hu Nan City Univ, Sch Mech & Elect Engn, Yi Yang 413099, Peoples R China
关键词
Mechanical ventilation; Respiratory mechanics; Lung compliance; Airway resistance; PRESSURE-SUPPORT VENTILATION; PROPORTIONAL-ASSIST VENTILATION; PULMONARY MECHANICS; DISTRESS-SYNDROME; PARAMETERS; SYSTEM; ASYNCHRONY; FAILURE;
D O I
10.1186/s40001-023-01146-y
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
BackgroundRespiratory mechanics monitoring provides useful information for guiding mechanical ventilation, but many measuring methods are inappropriate for awake patients. This study aimed to evaluate the accuracy of dynamic mechanics estimation using expiratory time constant (RCexp) calculation during noninvasive pressure support ventilation (PSV) with air leak in different lung models.MethodsA Respironics V60 ventilator was connected to an active breathing simulator for modeling five profiles: normal adult, restrictive, mildly and severely obstructive, and mixed obstructive/restrictive. Inspiratory pressure support was adjusted to maintain tidal volumes (V-T), achieving 5.0, 7.0, and 10.0 ml/kg body weight. PEEP was set at 5 cmH(2)O, and the back-up rate was 10 bpm. Measurements were conducted at system leaks of 25-28 L/min. RCexp was estimated from the ratio at 75% exhaled V-T and flow rate, which was then used to determine respiratory system compliance (C-rs) and airway resistance (R-aw).ResultsIn non-obstructive conditions (R-aw & LE; 10 cmH(2)O/L/s), the C-rs was overestimated in the PSV mode. Peak inspiratory and expiratory flow and V-T increased with PS levels, as calculated C-rs decreased. In passive breathing, the difference of C-rs between different V-T was no significant. Underestimations of inspiratory resistance and expiratory resistance were observed at V-T of 5.0 ml/kg. The difference was minimal at V-T of 7.0 ml/kg. During non-invasive PSV, the estimation of airway resistance with the RCexp method was accurately at V-T of 7.0 ml/kg.ConclusionsThe difference between the calculated C-rs and the preset value was influenced by the volume, status and inspiratory effort in spontaneously breathing.
引用
收藏
页数:11
相关论文
共 20 条
  • [1] Accuracy of the estimations of respiratory mechanics using an expiratory time constant in passive and active breathing conditions: a bench study
    Yuqing Chen
    Yueyang Yuan
    Hai Zhang
    Feng Li
    European Journal of Medical Research, 28
  • [2] Accuracy of the Ventilator Automated Displayed Respiratory Mechanics in Passive and Active Breathing Conditions: A Bench Study
    Daoud, Ehab G.
    Katigbak, Reynaldo
    Ottochian, Marcus
    RESPIRATORY CARE, 2019, 64 (12) : 1555 - 1560
  • [3] ACTIVE AND PASSIVE RESPIRATORY MECHANICS AND CONTROL OF BREATHING IN KITTENS
    ZIN, WA
    MARLOT, D
    BONORA, M
    SIAFAKAS, NM
    GAUTIER, H
    MILICEMILI, J
    JOURNAL OF APPLIED PHYSIOLOGY, 1983, 55 (01) : 183 - 190
  • [4] Determining respiratory rate using measured expiratory time constant: A prospective observational study
    Depta, Filip
    Gentile, Michael A.
    Kallet, Richard H.
    Firment, Peter
    Leskanic, Jozef
    Rybar, Dusan
    Torok, Pavol
    Zdravkovic, Marko
    JOURNAL OF CRITICAL CARE, 2023, 73
  • [5] Non-invasive determination of respiratory system mechanics in pressure support ventilation using the expiratory time constant? Response
    Becher, Tobias
    Schaedler, Dirk
    Frerichs, Inez
    Weiler, Norbert
    CRITICAL CARE, 2013, 17 (02):
  • [6] Inspiratory respiratory mechanics estimation by using expiratory data for reverse-triggered breathing cycles
    Howe, S. L.
    Chase, J. G.
    Redmond, D. P.
    Morton, S. E.
    Kim, K. T.
    Pretty, C.
    Shaw, G. M.
    Tawhai, M. H.
    Desaive, T.
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 186
  • [7] Passive expiratory flow-volume curve is not an accurate method to measure the respiratory time constant in rabbits
    Vialet, R
    Monnier, A
    Lagier, P
    Jammes, Y
    Toméi, C
    Martin, C
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1999, 43 (10) : 1017 - 1020
  • [8] Performance of low- flow oxygen delivery systems with mouth breathing and altered respiratory mechanics: a bench study.
    Poncin, William
    Arrouch, Shani
    Martinez-Alejos, Roberto
    Fresnel, Emeline
    Fresnel, Emeline
    EUROPEAN RESPIRATORY JOURNAL, 2024, 64
  • [9] Accuracy of the dynamic signal analysis approach in respiratory mechanics during noninvasive pressure support ventilation: a bench study
    Chen, Yuqing
    Yuan, Yueyang
    Zhang, Hai
    Li, Feng
    Zhou, Xin
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2021, 49 (02)
  • [10] PASSIVE EXPIRATORY FLOW-VOLUME RECORDINGS IN IMMATURE NEWBORN RABBITS - EFFECT OF SURFACTANT REPLACEMENT ON THE TIME CONSTANT OF THE RESPIRATORY SYSTEM
    NOACK, G
    CURSTEDT, T
    GROSSMANN, G
    NILSSON, R
    ROBERTSON, B
    RESPIRATION, 1990, 57 (01) : 1 - 5