Adaptive Calibration in Electrothermal Atomic Absorption Spectrometry

被引:0
|
作者
Sadagov, Yu. M. [1 ]
Sadagov, A. Yu. [1 ]
机构
[1] Cortec Ltd, Moscow 119602, Russia
关键词
ETAAS; transformation function; calibration dependence; adaptive calibration; matrix effects;
D O I
10.1134/S1061934823080142
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method is developed for determining concentrations in Zeeman electrothermal atomic absorption spectrometry based on the conversion of the transformation function of an atomic absorption spectrometer into a calibration dependence adapted to the matrix of the analyzed sample. The transformation function is expressed as a one-parameter saturation function, which is an integral optical characteristic of the spectrometer, independent of the composition of the analyzed sample. The algorithm of the method includes measurement of the analytical signals of an analyzed sample and a same sample with a known additive of the analyte and calculation of the matrix coefficient and the desired concentration of the analyte in the analyzed sample from these signals. The adaptive calibration method is tested for As, Au, Cd, Cu, Mo, Pd, Pb, Mo, Pd, Pb, Ti, V. The relative systematic error in determining the concentration of an element in the range of measuring the absorbance of atomic vapor does not exceed 10%. The analytical ranges for the elements exceed three orders of magnitude.
引用
收藏
页码:988 / 994
页数:7
相关论文
共 50 条
  • [1] Adaptive Calibration in Electrothermal Atomic Absorption Spectrometry
    Yu. M. Sadagov
    A. Yu. Sadagov
    Journal of Analytical Chemistry, 2023, 78 : 988 - 994
  • [2] CORRECTION FOR CURVATURE IN CALIBRATION CURVES IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY
    BAYUNOV, PA
    LVOV, BV
    JOURNAL OF ANALYTICAL CHEMISTRY OF THE USSR, 1987, 42 (04): : 495 - 500
  • [3] METHODS OF CALIBRATION IN THE DIRECT ANALYSIS OF SOLID SAMPLES BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY
    BERGLUND, M
    BAXTER, DC
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (14) : E1567 - E1586
  • [4] Enhancement of the atomic absorption signal by silicate at electrothermal atomic absorption spectrometry
    Nukatsuka, I
    Anezaki, K
    Kubota, S
    Ohzeki, K
    ANALYTICA CHIMICA ACTA, 2000, 415 (1-2) : 221 - 227
  • [5] Electrothermal atomic absorption spectrometry and related techniques
    Jackson, KW
    ANALYTICAL CHEMISTRY, 2000, 72 (12) : 159R - 167R
  • [6] Electrothermal metallic furnace atomic absorption spectrometry
    Morzan, E.
    Carrone, G.
    Tudino, M.
    Etchenique, R.
    ANALYTICAL METHODS, 2017, 9 (05) : 756 - 760
  • [7] Palladium release in electrothermal atomic absorption spectrometry
    Slaveykova, VI
    Manev, SG
    Lazarov, DL
    SPECTROSCOPY LETTERS, 1997, 30 (02) : 297 - 307
  • [8] Stabilizing Agents for Calibration in the Determination of Mercury Using Solid Sampling Electrothermal Atomic Absorption Spectrometry
    Zelinkova, Hana
    Cervenka, Rostislav
    Komarek, Josef
    SCIENTIFIC WORLD JOURNAL, 2012,
  • [9] Lead-depleted blood for matrix-matched calibration with electrothermal atomic absorption spectrometry
    Ilkov, A
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (04) : 313 - 315
  • [10] CALIBRATION IN ATOMIC-ABSORPTION SPECTROMETRY
    KABAILA, P
    LWIN, T
    STRODE, P
    ANALYST, 1987, 112 (11) : 1541 - 1546