Clinical performance of two new, fully integrated molecular platforms used for HIV-1, HBV and HCV viral load analysis, the NeuMoDx 288 and the Alinity m.

被引:5
|
作者
Mourik, K. [1 ]
Boers, S. A. [1 ]
van Rijn, A. L. [1 ]
Thijssen, J. C. P. [1 ]
Wessels, E. [1 ]
Claas, E. C. J. [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Med Microbiol, Leiden, Netherlands
关键词
Fully automated molecular platform; HIV; HBV; HCV; Viral load testing; COBAS AMPLIPREP/COBAS TAQMAN; REALTIME; ASSAY;
D O I
10.1016/j.jcv.2022.105376
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Viral load (VL) determination in patients with human immunodeficiency virus type 1 (HIV-1), hepatitis B virus (HBV) and hepatitis C virus (HCV) is essential for proper patient management and follow-up. New molecular platforms have been developed to fully automate these diagnostic assays.Objective: Evaluation of the clinical performance of HIV-1, HBV and HCV VL assays on the Alinity m (Abbott) and NeuMoDx (Qiagen) molecular platforms.Method: Test panels of the three viruses have been compiled of 100 plasma and/or serum samples per target containing non-detectable, non-quantifiable and quantifiable VLs. All samples were retrospectively tested on the Alinity m and NeuMoDx platforms according to manufacturers' instructions. Results: A total of 74, 86 and 66 samples with valid results for both platforms were included in the HIV-1, HBV and HCV analysis respectively. Overall qualitative agreement of the assays on both platforms was 78% for HIV-1, 93% for HBV and 100% for HCV. Quantitative agreement (less than 0.5 log difference) was shown to be 68% for HIV-1, 68% for HBV and 94% for HCV.Conclusion: The Alinity m and NeuMoDx HCV assay have a comparable performance. Quantification differences in the HIV-1 assay were mostly apparent in the lower VLs and under-quantification of the NeuMoDx HBV assay was observed.
引用
收藏
页数:6
相关论文
共 15 条
  • [1] Analytical performance of four molecular platforms used for HIV-1, HBV and HCV viral load determinations
    Aretzweiler, Gudrun
    Leuchter, Susanne
    Garcia-Alvarez, Monica
    Simon, Christian
    Marins, Ed
    Paxinos, Ellen
    Canchola, Jesse
    Delgado, Rafael
    Frontzek, Andre
    EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2019, 19 (10) : 941 - 949
  • [2] Clinical performance evaluation of HCV and HIV-1 assays on the fully automated molecular system Alinity m
    Phan, Tung
    Cravener, Zachary
    Olean, Janet
    Mccullough, Melissa
    Gribschaw, Jamie
    Wells, Alan
    JOURNAL OF CLINICAL VIROLOGY PLUS, 2023, 3 (04):
  • [3] Performance Evaluation of two new molecular platforms for HIV-1 RNA measurement: The COBAS HIV-1 5800 and the NeuMoDx 96
    Papachristou, E.
    Kitsou, K.
    Rokka, C.
    Iliopoulos, P.
    Maneka, L.
    Protopapas, K.
    Chini, M.
    Kosmopoulou, O.
    Psichogiou, M.
    Adamis, G.
    Astriti, M.
    Lagiou, P.
    Paraskevis, D.
    Magiorkinis, G.
    HIV MEDICINE, 2023, 24 : 347 - 348
  • [4] Comparison of the Abbott Alinity m and m2000 assays for the quantification of HIV-1, HCV and HBV in clinical samples
    Mouna, Lina
    Pallier, Coralie
    Proust, Stephanie
    Pregermain, Corinne
    Roque-Afonso, Anne-Marie
    JOURNAL OF CLINICAL VIROLOGY, 2020, 126
  • [5] Analytical and clinical performance of the procleix ultrio HIV-1/HCV/HBV assay in low viral load samples
    Katsoulidou, A.
    Moschidis, Z.
    Sypsa, V.
    Chini, M.
    Papatheodoridis, G.
    Tassopoulos, N.
    Mimidis, K.
    Karafoulidou, A.
    Hatzakis, A.
    VOX SANGUINIS, 2006, 91 : 84 - 84
  • [6] Analytical and clinical sensitivity of the Procleix Ultrio HIV-1/HCV/HBV assay in samples with a low viral load
    Katsoulidou, A.
    Moschidis, Z.
    Sypsa, V.
    Chini, M.
    Papatheodoridis, G. V.
    Tassopoulos, N. C.
    Mimidis, K.
    Karafoulidou, A.
    Hatzakis, A.
    VOX SANGUINIS, 2007, 92 (01) : 8 - 14
  • [7] Clinical performance evaluation of the Aptima viral assays for the quantitation of HIV-1, HCV, and HBV in plasma samples
    Lewis, Shawna
    Morehead, Elizabeth
    Cheung, Helen
    Mostafa, Heba H.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2023, 106 (03)
  • [8] Evaluation of the Alinity m HIV-1 assay for the quantification of HIV-1 RNA plasma viral load in a high-throughput molecular laboratory in South Africa
    Maree, Leana
    Krugel, Maria
    Reinhardt, Birgit
    Glass, Allison J.
    JOURNAL OF CLINICAL VIROLOGY, 2020, 132
  • [9] Performance Evaluation of Abbott Alinity m System to Detect HBV, HCV, and HIV-1 Infections: Comparison with Hologic Panther Aptima System
    Han, J.
    Woo, K.
    Jeong, I.
    Kwak, M.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2020, 22 (11): : S45 - S45
  • [10] Evaluation of the analytical and clinical performance of the Abbott Alinity m instrument for the quantification of HIV-1 RNA plasma viral load in a reference laboratory in Rio de Janeiro, Brazil
    Costa, Diego S.
    Quintana, Marcel S. B.
    Grinsztejn, Beatriz
    Veloso, Valdilea G.
    Silva, Simone C. C.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2024,