Acute human defibrillation performance of a subcutaneous implantable cardioverter-defibrillator with an additional coil electrode

被引:4
|
作者
Yap, Sing -Chien [7 ,8 ]
Oosterwerff, Erik F. J. [6 ]
Boersma, Lucas V. A. [1 ,2 ]
van der Stuijt, Willeke [2 ]
Lenssen, Anneke [4 ]
Hahn, Stephen J. [5 ]
Knops, Reinoud E. [3 ]
机构
[1] Erasmus MC, Cardiovasc Inst, Thorax Ctr, Dept Cardiol, Rotterdam, Netherlands
[2] Isala Klin, Dept Cardiol, Zwolle, Netherlands
[3] St Antonious Ziekenhuis, Dept Cardiol, Nieuwegein, Netherlands
[4] Univ Amsterdam, Med Ctr, Dept Cardiol, Amsterdam, Netherlands
[5] Boston Sci, Amsterdam, Netherlands
[6] Boston Sci CRM, St Paul, MN USA
[7] Flevo Hosp, Almere, Netherlands
[8] Erasmus MC, Dept Surg, Doctor Molenwaterpl 40, NL-3015 GD Rotterdam, Netherlands
关键词
Defibrillation; Subcutaneous; Human; Electrodes; Implantable cardioverter-defibrillator; OUTCOMES; TRIAL; ICD;
D O I
10.1016/j.hrthm.2023.08.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The subcutaneous implantable cardioverter-defibrillator (S-ICD) delivers 80 J shocks from an 8 cm left-parasternal coil to a 59 cm(3) left lateral pulse generator (PG). A system that defibrillates with lower energy could significantly reduce PG size. Computer modeling and animal studies suggested that a second shock coil either parallel to the left-parasternal coil or transverse from the xiphoid to the PG pocket would significantly reduce the defibrillation threshold.Objective The purpose of this study was to acutely assess the defibrillation efficacy of parallel and transverse configurations in patients receiving an S-ICD.Methods Testing was performed in patients receiving a conventional S-ICD system. Success at 65 J was required before investigational testing. A second electrode was temporarily inserted from the xiphoid incision connected to the PG with an investigational Y-adapter. Phase 1 (n = 11) tested the parallel configuration. Phase 2 (n = 21) tested both parallel and transverse configurations in random order.Results This study enrolled 35 patients (28 males (80%); mean age 51 +/- 17 years; left ventricular ejection fraction 40% +/- 15%; body mass index 26 +/- 4 kg/m(2); prior myocardial infarction 46%; congestive heart failure 49%; cardiomyopathy 63%). Compared to the conventional S-ICD system, mean shock impedance decreased for both parallel (69 +/- 15 Omega vs 86 +/- 20 Omega; n = 33; P < .001) and transverse (56 +/- 14 Omega vs 81 +/- 21 Omega; n = 20; P < .001) configurations. Shock success rates at 20, 30, and 40 J were 55%, 79%, 97%, and 25%, 70%, 90% for parallel and transverse configurations, respectively. Defibrillation threshold testing was well tolerated with no serious adverse events.Conclusion Adding a second shock coil, particularly in the parallel configuration, significantly reduced the impedance and had a high likelihood of defibrillation success at energies <= 40 J. This may enable the development of a smaller S-ICD.
引用
收藏
页码:1649 / 1656
页数:8
相关论文
共 50 条
  • [1] Subcutaneous Implantable Cardioverter-Defibrillator Implantation Without Defibrillation Testing
    Miller, Marc A.
    Palaniswamy, Chandrasekar
    Dukkipati, Srinivas R.
    Balulad, Sujata
    Smietana, Jeffrey
    Vigdor, Aaron
    Koruth, Jacob S.
    Choudry, Subbarao
    Whang, William
    Reddy, Vivek Y.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (25) : 3118 - 3119
  • [2] The subcutaneous implantable cardioverter-defibrillator
    Grace, Andrew
    CURRENT OPINION IN CARDIOLOGY, 2014, 29 (01) : 10 - 19
  • [3] Defibrillation testing of the subcutaneous implantable cardioverter-defibrillator at the time of generator replacement
    Schaller, Robert D.
    Hyman, Matthew
    Supple, Gregory E.
    Santangeli, Pasquale
    Riley, Michael P.
    Nazarian, Saman
    Arkles, Jeffrey
    Garcia, Fermin
    Lin, David
    Guandalini, Gustavo
    Kumareswaran, Ramanan
    Deo, Rajat
    Bode, Weeranun
    Markman, Timothy
    Epstein, Andrew
    Callans, David J.
    Dixit, Sanjay
    Brozoski, Joseph
    Marchlinski, Francis E.
    Frankel, David S.
    HEART RHYTHM, 2024, 21 (01) : 117 - 118
  • [4] Intermuscular Subcutaneous Implantable Cardioverter-Defibrillator Implantation Without Defibrillation Testing
    Mistry, Amar
    Pooranachandran, Vivetha
    Dhutia, Harshil
    Chelliah, Rajesh
    Pathmanathan, Ravi K.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2021, 7 (01) : 124 - 126
  • [5] Subcutaneous Implantable Cardioverter-Defibrillator Technology
    Gupta, Anurag
    Al-Ahmad, Amin
    Wang, Paul J.
    HEART FAILURE CLINICS, 2011, 7 (02) : 287 - +
  • [6] An Entirely Subcutaneous Implantable Cardioverter-Defibrillator
    Bardy, Gust H.
    Smith, Warren M.
    Hood, Margaret A.
    Crozier, Ian G.
    Melton, Iain C.
    Jordaens, Luc
    Theuns, Dominic
    Park, Robert E.
    Wright, David J.
    Connelly, Derek T.
    Fynn, Simon P.
    Murgatroyd, Francis D.
    Sperzel, Johannes
    Neuzner, Joerg
    Spitzer, Stefan G.
    Ardashev, Andrey V.
    Oduro, Amo
    Boersma, Lucas
    Maass, Alexander H.
    Van Gelder, Isabelle C.
    Wilde, Arthur A.
    van Dessel, Pascal F.
    Knops, Reinoud E.
    Barr, Craig S.
    Lupo, Pierpaolo
    Cappato, Riccardo
    Grace, Andrew A.
    NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (01): : 36 - 44
  • [7] The subcutaneous implantable cardioverter-defibrillator in review
    Kamp, Nicholas J.
    Al-Khatib, Sana M.
    AMERICAN HEART JOURNAL, 2019, 217 : 131 - 139
  • [8] Reduction in defibrillation threshold by modifications to the subcutaneous implantable-cardioverter defibrillator coil
    Heist, E. Kevin
    Knops, Reinoud E.
    Yap, Sing-Chien
    Boersma, Lucas V. A.
    Friedman, Daniel J.
    Poole, Jeanne E.
    Stahl, Wyatt
    Belalcazar, Andres
    HEART RHYTHM, 2023, 20 (05) : 783 - 785
  • [9] Subcutaneous transthoracic defibrillation in implantable cardioverter defibrillator patients using a pectoral can to posterior coil electrode configuration
    Lieberman, RA
    Shorofsky, SR
    Rashba, E
    Havel, WJ
    Christensen, JL
    De Groot, PJ
    CIRCULATION, 2005, 112 (17) : U429 - U429
  • [10] Dual coil system for the subcutaneous implantable cardioverter-defibrillator: Going back to the future
    Shim, David J.
    Henrikson, Charles A.
    HEART RHYTHM, 2023, 20 (12) : 1657 - 1658