Stability Analysis of Plate-Screw Fixation for Femoral Midshaft Fractures

被引:2
|
作者
Basirom, Izzawati [1 ]
Daud, Ruslizam [1 ]
Ijaz, Muhammad Farzik [2 ]
Rojan, Mohd Afendi [1 ]
Basaruddin, Khairul Salleh [1 ]
机构
[1] Univ Malaysia Perlis, Fac Mech Engn Technol, Fracture & Damage Mech FDM, Arau 02600, Perlis, Malaysia
[2] King Saud Univ, Coll Engn, Mech Engn Dept, POB 800, Riyadh 11421, Saudi Arabia
关键词
Implant stability; interfragmentary strain; transverse fracture; screw configuration; COMPRESSION PLATE; DENTAL IMPLANT; BONE; TITANIUM; GAP; OPTIMIZATION; BIOMECHANICS; INTERFACE; EVOLUTION; ALLOYS;
D O I
10.3390/ma16175958
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
An understanding of the biomechanical characteristics and configuration of flexible and locked plating in order to provide balance stability and flexibility of implant fixation will help to construct and promote fast bone healing. The relationship between applied loading and implantation configuration for best bone healing is still under debate. This study aims to investigate the relationship between implant strength, working length, and interfragmentary strain (eIFM) on implant stability for femoral midshaft transverse fractures. The transverse fracture was fixed with a fragment locking compression plate (LCP) system. Finite element analysis was performed and subsequently characterised based on compression loading (600 N up to 900 N) and screw designs (conventional and locking) with different penetration depths (unicortical and bicortical). Strain theory was used to evaluate the stability of the model. The correlation of screw configuration with screw type shows a unicortical depth for both types (p < 0.01) for 700 N and 800 N loads and (p < 0.05) for configurations 134 and 124. Interfragmentary strain affected only the 600 N load (p < 0.01) for the bicortical conventional type (group BC), and the screw configurations that were influenced were 1234 and 123 (p < 0.05). The low steepness of the slope indicates the least eIFM for the corresponding biomechanical characteristic in good-quality stability. A strain value of =2% promotes callus formation and is classified as absolute stability, which is the minimum required value for the induction of callus and the maximum value that allows bony bridging. The outcomes have provided the correlation of screw configuration in femoral midshaft transverse fracture implantation which is important to promote essential primary stability.
引用
收藏
页数:21
相关论文
共 50 条
  • [1] SLIDING SCREW PLATE FIXATION OF INTERTROCHANTERIC FEMORAL FRACTURES
    MULHOLLAND, RC
    GUNN, DR
    ORTH, MC
    JOURNAL OF TRAUMA, 1972, 12 (07): : 581 - +
  • [2] Finite element analysis of spiral plate and Herbert screw fixation for treatment of midshaft clavicle fractures
    Zhang, Xiaojuan
    Cheng, Xiaodong
    Yin, Bing
    Wang, Jianzhao
    Li, Sheng
    Liu, Guobin
    Hu, Zusheng
    Wu, Weiwei
    Zhang, Yingze
    MEDICINE, 2019, 98 (34)
  • [3] Intramedullary screw fixation for midshaft fractures of the clavicle
    Khalil, Ayman
    INTERNATIONAL ORTHOPAEDICS, 2009, 33 (05) : 1421 - 1424
  • [4] Intramedullary Screw Fixation for Midshaft Clavicle Fractures
    Bents, Robert Thurston
    Bents, Easton James
    ARTHROSCOPY TECHNIQUES, 2024, 13 (03):
  • [5] Intramedullary screw fixation for midshaft fractures of the clavicle
    Ayman Khalil
    International Orthopaedics, 2009, 33 : 1421 - 1424
  • [6] Biomechanical comparison of plate-screw and screw fixation in medial tibial plateau fractures (Schatzker 4). A model study
    Cift, H.
    Cetik, O.
    Kalaycioglu, B.
    Dirikoglu, M. H.
    Ozkan, K.
    Eksioglu, F.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2010, 96 (03) : 263 - 267
  • [7] A comparison of intramedullary nailing and plate-screw fixation in the treatment for ipsilateral fracture of the hip and femoral shaft
    Kesemenli C.C.
    Tosun B.
    Kim N.S.Y.
    MUSCULOSKELETAL SURGERY, 2012, 96 (2) : 117 - 124
  • [8] ANATOMIC CONSIDERATIONS FOR DORSAL SACRAL PLATE-SCREW FIXATION
    XU, RM
    EBRAHEIM, NA
    MOHAMED, A
    ELGAMAL, H
    YEASTING, RA
    JOURNAL OF SPINAL DISORDERS, 1995, 8 (05): : 352 - 356
  • [9] A mechanical performance study of the plate-screw system applied to fractures in the femoral shaft - An in vitro study
    Dalmina, V
    Borges, PC
    Cziulik, C
    Petschow, R
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 1835 - 1838
  • [10] Treatment of pediatric femur supracondylar fractures: comparison of K-wire fixation versus plate-screw fixation
    Tekin, Sezgin Bahadir
    Karabulut, Cagri
    Bozgeyik, Bahri
    Gokcen, H. Bahadir
    JOURNAL OF PEDIATRIC ORTHOPAEDICS-PART B, 2024, 33 (01): : 70 - 75