Comparisons of biometric parameters measurements by OPD-SCAN III and Pentacam in cataract

被引:0
|
作者
Dong, Shiqi [1 ]
Xu, Rong [1 ]
Wang, Yue [1 ]
Xiao, Wei [1 ]
Jiao, Ya [1 ]
机构
[1] Hankou Aier Eye Hosp, Dept Cataract, Wuhan, Hubei, Peoples R China
关键词
Chord kappa; Chord alpha; OPD-Scan III; Pentacam; Cataract; CORNEAL SPHERICAL ABERRATIONS; INTRAOCULAR LENSES; ANGLE ALPHA; CENTRATION; SURGERY;
D O I
10.1016/j.pdpdt.2023.103911
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Comparison of OPD-Scan III and Pentacam measurements of relevant parameters guiding the selection of multifocal intraocular lenses in cataract patients. Methods: A retrospective study included 120 patients (120 eyes) with cataract. Their chord kappa, chord alpha, corneal spherical aberration, and pupil size were measured by OPD-Scan III and Pentacam. Pairwise t-tests, Pearson tests, and Bland-Altman analyses were used, respectively, to assess the difference, correlation, and agreement between the devices. Results: There was no significant difference in photopic chord kappa and corneal spherical aberration between the two instruments (P = 0.054, P = 0.065). Chord alpha and pupil size varied significantly between the two instruments (P < 0.001). OPD-Scan III results revealed a significant association between photopic chord kappa and mesopic chord kappa (r = 0.823, P < 0.001). There was a significant positive correlation between photopic and mesopic chord kappa measured by OPD-Scan III and chord kappa measured by Pentacam (r = 0.840, r = 0.757, P < 0.001). The chord alpha evaluated by the two tools had a moderate correlation (r = 0.442, P < 0.001) between them. Bland-Altman analysis showed that there was excellent agreement between the two instruments in measuring chord kappa and corneal spherical aberration, while the chord alpha and pupil size were not consistent. Conclusions: The chord kappa and corneal spherical aberration measured by OPD-Scan III and Pentacam were consistent in the preoperative decision making of multifocal intraocular lens implantation in cataract patients, both of which have clinical guiding significance. In addition, OPD-Scan III provides more comprehensive day and night chord kappa and pupil diameters, which improves clinical advice for patients with complicated ocular diseases and high demands for nocturnal vision.
引用
收藏
页数:5
相关论文
共 23 条
  • [1] Analysis of OPD-Scan and Pentacam Parameters for Early Keratoconus Detection
    Donoso, Rodrigo
    Rodriguez, Alvaro
    Esteffan, Karim
    Lagos, Claudio
    Aranguiz, Diego
    Hernandez, Nicolas
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2021, 226 : 235 - 242
  • [2] AcuTarget Measurements: Repeatability and Comparison to OPD-Scan III
    Guilbert, Emmanuel
    Saad, Alain
    Gatinel, Damien
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (03) : 180 - 185
  • [3] Analysis of biometric parameters of cataract eyes measured with optical biometer Lenstar LS900, IOL Master 700, and OPD-SCAN III
    Bao, Tong
    Wang, Lifeng
    Liu, Caihui
    Yang, Yajun
    Pang, Yunlei
    PHOTODIAGNOSIS AND PHOTODYNAMIC THERAPY, 2023, 43
  • [4] Precision of NIDEK OPD-scan measurements
    Holzer, Mike P.
    Goebels, Susanne
    Auffarth, Gerd U.
    JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (09) : S1021 - S1023
  • [5] Repeatability of pupil size measurements with NIDEK OPD-Scan III in myopic children
    Tan, Qi
    Ng, Alex L. K.
    Cheng, George P. M.
    Woo, Victor C. P.
    Cho, Pauline
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2021, 41 (02) : 431 - 436
  • [6] Repeatability of Pentacam-derived intraocular lens decentration measurements and the level of agreement with OPD-Scan III: A prospective observational case series
    Liu, Xiaobao
    Wu, Wenjie
    Huang, Yulong
    Fu, Yabo
    Huang, Yue
    Li, Qiong
    PLOS ONE, 2024, 19 (03):
  • [7] Wavefront aberrometry repeatability and agreement-A comparison between Pentacam AXL Wave, iTrace and OPD-Scan III
    Wan, Kelvin H.
    Liao, Xu Lin
    Yu, Marco
    Tsui, Rachel W. Y.
    Chow, Vanissa W. S.
    Chong, Kelvin K. L.
    Chan, Tommy C. Y.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2022, 42 (06) : 1326 - 1337
  • [8] Comparison of the VISX wavescan and OPD-scan III with the subjective refraction
    Zhu, R.
    Long, K. -L.
    Wu, X. -M.
    Li, Q. -D.
    EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 2016, 20 (14) : 2988 - 2992
  • [9] Evaluation of the auto-refraction function of the Nidek OPD-Scan III
    McGinnigle, Samantha
    Naroo, Shehzad A.
    Eperjesi, Frank
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2014, 97 (02) : 160 - 163
  • [10] Outcomes of Corneal Spherical Aberration-guided Cataract Surgery Measured by the OPD-Scan
    Solomon, Jonathan D.
    JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (11) : 863 - 869