Analysis of refraction and scattering image artefacts in x-ray analyzer-based imaging

被引:2
|
作者
Zhao, Li-Ming [1 ]
Wang, Tian-Xiang [1 ]
Ma, Run-Kang [1 ]
Gu, Yao [1 ]
Luo, Meng-Si [1 ]
Chen, Heng [1 ]
Wang, Zhi-Li [1 ]
Ge, Xin [2 ]
机构
[1] Hefei Univ Technol, Sch Phys, Dept Opt Engn, Hefei 230009, Anhui, Peoples R China
[2] Shenzhen Bay Lab, Inst Biomed Engn, Shenzhen 518067, Peoples R China
基金
中国国家自然科学基金;
关键词
x-ray imaging; analyzer-based imaging; image artefacts; PHASE-CONTRAST; RETRIEVAL; ABSORPTION; NOISE;
D O I
10.1088/1674-1056/ac685d
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
X-ray analyzer-based imaging (ABI) is a powerful phase-sensitive technique that can provide a wide dynamic range of density and extract useful physical properties of the sample. It derives contrast from x-ray absorption, refraction, and scattering properties of the investigated sample. However, x-ray ABI setups can be susceptible to external vibrations, and mechanical imprecisions of system components, e.g., the precision of motor, which are unavoidable in practical experiments. Those factors will provoke deviations of analyzer angular positions and hence errors in the acquired image data. Consequently, those errors will introduce artefacts in the retrieved refraction and scattering images. These artefacts are disadvantageous for further image interpretation and tomographic reconstruction. For this purpose, this work aims to analyze image artefacts resulting from deviations of analyzer angular positions. Analytical expressions of the refraction and scattering image artefacts are derived theoretically and validated by synchrotron radiation experiments. The results show that for the refraction image, the artefact is independent of the sample's absorption and scattering signals. By contrast, artefact of the scattering image is dependent on both the sample's refraction and scattering signals, but not on absorption signal. Furthermore, the effect of deviations of analyzer angular positions on the accuracy of the retrieved images is investigated, which can be of use for optimization of data acquisition. This work offers the possibility to develop advanced multi-contrast image retrieval algorithms that suppress artefacts in the retrieved refraction and scattering images in x-ray analyzer-based imaging.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Analysis of refraction and scattering image artefacts in x-ray analyzer-based imaging
    赵立明
    王天祥
    马润康
    顾瑶
    罗梦丝
    陈恒
    王志立
    葛昕
    Chinese Physics B, 2023, 32 (02) : 625 - 630
  • [2] Absorption, refraction and scattering retrieval in X-ray analyzer-based imaging
    Wang, Zhili
    Liu, Dalin
    Zhang, Jin
    Huang, Wanxia
    Yuan, Qingxi
    Gao, Kun
    Wu, Zhao
    JOURNAL OF SYNCHROTRON RADIATION, 2018, 25 : 1206 - 1213
  • [3] Refraction and scattering of X-rays in analyzer-based imaging
    Suhonen, Heikki
    Fernandez, Manuel
    Bravin, Alberto
    Keyrilaeinen, Jani
    Suortti, Pekka
    JOURNAL OF SYNCHROTRON RADIATION, 2007, 14 (512-521) : 512 - 521
  • [4] Microbubbles as x-ray scattering contrast agents using analyzer-based imaging
    Arfelli, F.
    Rigon, L.
    Menk, R. H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (06): : 1643 - 1658
  • [5] Retrieval of multiple scattering contrast from x-ray analyzer-based imaging
    陈恒
    刘波
    赵立明
    任坤
    王志立
    Chinese Physics B, 2021, 30 (01) : 695 - 699
  • [6] Retrieval of multiple scattering contrast from x-ray analyzer-based imaging
    Chen, Heng
    Liu, Bo
    Zhao, Li-Ming
    Ren, Kun
    Wang, Zhi-Li
    CHINESE PHYSICS B, 2021, 30 (01)
  • [7] Biases of estimated signals in x-ray analyzer-based imaging
    夏健霖
    徐文
    周瑞成
    石晓敏
    任坤
    陈恒
    王志立
    Chinese Physics B, 2020, (06) : 666 - 671
  • [8] Biases of estimated signals in x-ray analyzer-based imaging
    Xia, Jianlin
    Xu, Wen
    Zhou, Ruicheng
    Shi, Xiaomin
    Ren, Kun
    Chen, Heng
    Wang, Zhili
    CHINESE PHYSICS B, 2020, 29 (06)
  • [9] Absorption, refraction and scattering in analyzer-based imaging: comparison of different algorithms
    Diemoz, Paul Claude
    Coan, Paola
    Glaser, Christian
    Bravin, Alberto
    OPTICS EXPRESS, 2010, 18 (04): : 3494 - 3509
  • [10] Analyzer-Based Phase Contrast X-ray Imaging for mouse tissues
    Li, H.
    Wang, M.
    Wang, Z.
    Luo, S. Q.
    BIOINFORMATICS AND BIOMEDICAL ENGINEERING: NEW ADVANCES, 2016, : 355 - 358