A Conjoined Rectilinear Collision Cell and Pulsed Extraction Ion Trap with Auxiliary DC Electrodes

被引:14
|
作者
Stewart, Hamish [1 ]
Grinfeld, Dmitry [1 ]
Wagner, Alexander [1 ]
Kholomeev, Alexander [1 ]
Biel, Matthias [1 ]
Giannakopulos, Anastassios [1 ]
Makarov, Alexander [1 ]
Hock, Christian [1 ]
机构
[1] Thermo Fisher Sci, D-28199 Bremen, Germany
基金
欧盟地平线“2020”;
关键词
PROTEOME COVERAGE; MASS-SPECTROMETRY; PERFORMANCE;
D O I
10.1021/jasms.3c00311
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Ion traps are routinely directly coupled to mass analyzers, where they serve to suitably cool and shape an ion population prior to pulsed extraction into the analyzer proper. Such devices benefit from high duty cycle and transmission but suffer slow ion processing times caused by a compromise in the buffer gas pressure range that suitably dampens the ion kinetic energy without causing excessive scatter during extraction or within the analyzer. A rectilinear RF quadrupole ion trap has been characterized, conjoining a pressurized collision region with a pumped extraction region, and an unbroken RF interface for seamless ion transfer between them. Auxiliary electrodes mounted between the RF electrodes provide DC voltage gradients that serve to both guide ions through the device and position them at the extraction slot. The influence of the auxiliary DC upon the trapping RF field was measured, and suitable parameters were defined. A mode of operation was developed that allowed parallel processing of ions in both regions, enabling a repetition rate of 200 Hz when the device was coupled to a high-resolution accurate-mass analyzer.
引用
收藏
页码:74 / 81
页数:8
相关论文
共 17 条
  • [1] Comparison between leaky and pulsed modes of ion extraction from an electron beam ion trap
    Hu, Zhimin
    Ishiguro, Yusuke
    Watanabe, Hirofumi
    Ohtani, Shunsuke
    Nakamura, Nobuyuki
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 298 : 96 - 99
  • [2] Pulsed ion extraction diagnostics in a quadrupole ion trap linear time-of-flight mass spectrometer
    Dangi, Beni B.
    Sassin, Nicholas A.
    Ervin, Kent M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (06):
  • [3] Lithium-Ion Cell Balancing Using Auxiliary Battery and DC/DC Unidirectional Converter
    Hashim, Muhammad Hazwan
    Jamaluddin, Mohd Saiful
    Hun, Mohd Hafidzuddin Sam
    Aziz, Mohd Junaidi Abdul
    Nordin, Norjulia Mohamad
    5TH IEEE CONFERENCE ON ENERGY CONVERSION 2021 (CENCON 2021), 2021, : 52 - 57
  • [4] Characterization of a high-pressure laser ion source with dc and pulsed extraction
    Yeates, P.
    Costello, J. T.
    Kennedy, E. T.
    PLASMA SOURCES SCIENCE & TECHNOLOGY, 2010, 19 (06):
  • [5] Dipolar Direct Current Driven Collision-Induced Dissociation in a Digital Ceramic-Based Rectilinear Ion Trap Mass Spectrometer
    Wang, Liang
    Xu, Fuxing
    Ding, Chuan-Fan
    ANALYTICAL CHEMISTRY, 2013, 85 (03) : 1271 - 1275
  • [6] Improving the sensitivity of miniature linear ion trap mass spectrometer by a DC voltage applied on the eject electrodes
    Li, Gang
    Li, Detian
    Cheng, Yongjun
    Sun, Wenjun
    Pei, Xiaoqiang
    Han, Xiaodong
    Zhang, Huzhong
    Wang, Yongjun
    Sun, Jian
    Dong, Meng
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2018, 24 (04) : 322 - 329
  • [7] Axial focusing following low kinetic energy pulsed extraction from a miniature linear ion trap
    Dobson, Gareth S.
    Enke, Christie G.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (17) : 2905 - 2912
  • [8] Hybridization of Pulsed-Q Dissociation and Collision-Activated Dissociation in Linear Ion Trap Mass Spectrometer for iTRAQ Quantitation
    Guo, Tiannan
    Gan, Chee Sian
    Zhang, Huoming
    Zhu, Yi
    Kon, Oi Lian
    Sze, Siu Kwan
    JOURNAL OF PROTEOME RESEARCH, 2008, 7 (11) : 4831 - 4840
  • [9] Enhanced Performance of Pulsed Q Collision Induced Dissociation-Based Peptide Identification on a Dual-Pressure Linear Ion Trap
    Loessner, Christopher
    Blackstock, Walter
    Gunaratne, Jayantha
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2012, 23 (01) : 186 - 189
  • [10] Particle-in-cell simulation of ion beam extraction from a pulsed plasma through a grid
    Nam, Sang Ki
    Economou, Demetre J.
    Donnelly, Vincent M.
    PLASMA SOURCES SCIENCE & TECHNOLOGY, 2007, 16 (01): : 90 - 96