Signal-correction errors in the EasyOne Pro LAB multiple-breath washout device significantly impact outcomes in children and adults

被引:1
|
作者
Oestreich, Marc-Alexander [1 ]
Wyler, Florian [1 ]
Buess, Christian [2 ]
Etter, Blaise [1 ]
Ramsey, Kathryn A. [1 ]
Latzin, Philipp [1 ]
机构
[1] Bern Univ Hosp, Univ Bern, Dept Paediat, Div Paediat Resp Med & Allergol,Inselspital, Bern, Switzerland
[2] ndd Medizintechn AG, Zurich, Switzerland
基金
瑞士国家科学基金会;
关键词
cystic fibrosis; lung clearance index; lung disease; pulmonary function testing; ventilation inhomogeneity; CYSTIC-FIBROSIS;
D O I
10.1152/japplphysiol.00096.2022
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Multiple-breath washout (MBW) is an established technique to assess functional residual capacity (FRC) and ventilation inhomogeneity in the lung. Indirect calculation of nitrogen concentration requires accurate measurement of gas concentrations. To investigate the accuracy of the CO2 concentration and molar mass (MM) values used for the indirect calculation of nitrogen concentration in a commercial MBW device [EasyOne Pro LAB (EOPL), ndd Medizintechnik AG, Switzerland] and its impact on outcomes. We used high-precision gas mixtures to evaluate CO2 and MM sensor output in vivo and in vitro. We developed updated algorithms to correct observed errors and assessed the impact on MBW outcomes and FRC measurement accuracy compared with body plethysmography. The respiratory exchange ratio (RER)-based adjustment of the measured CO2 signal used in the EOPL led to an overestimated CO2 signal (range -0.1% to 1.0%). In addition, an uncorrected dependence on humidity was identified. These combined effects resulted in an overestimation of expired nitrogen concentrations (range -0.7% to 2.6%), and consequently MBW outcomes. Corrected algorithms reduced the mean (SD) cumulative expired volume by 15.8% (9.7%), FRC by 6.6% (3.0%), and lung clearance index by 9.9% (7.6%). Differences in FRC between the EOPL and body plethysmography further increased. Inadequate signal correction causes RER- and humidity-dependent expired nitrogen concentration errors and overestimation of test outcomes. Updated algorithms reduce average signal error, however, RER values far from the population average still cause measurement errors. Despite improved signal accuracy, the updated algorithm increased the difference in FRC between the EOPL and body plethysmography.
引用
收藏
页码:460 / 471
页数:12
相关论文
共 20 条
  • [1] A non-systematic signal-correction error in the commercial EasyOne Pro multiple-breath washout device significantly impacts outcomes in children and adults
    Oestreich, M-A
    Wyler, F.
    Etter, B.
    Ramsey, K. A.
    Latzin, P.
    KLINISCHE PADIATRIE, 2022, 234 (05): : 329 - 329
  • [2] Characterization of a signal-correction error in a commercial multiple-breath washout device
    Oestreich, M.
    Wyler, F.
    Etter, B.
    Ramsey, K.
    Latzin, P.
    EUROPEAN RESPIRATORY JOURNAL, 2022, 60
  • [3] VALIDATION OF THE NDD EASYONE PRO LAB MULTIPLE BREATH NITROGEN WASHOUT SYSTEM IN CHILDREN
    Isaac, S. M.
    Jensen, R.
    Davies, J. C.
    Fuchs, S.
    Gappa, M.
    Latzin, P.
    Saunders, C.
    Short, C.
    Singer, F.
    Stanojevic, S.
    Zwitserloot, A.
    Ratjen, F.
    PEDIATRIC PULMONOLOGY, 2018, 53 : 231 - 231
  • [4] Correction of sensor crosstalk error in Exhalyzer D multiple-breath washout device significantly impacts outcomes in children with cystic fibrosis
    Wyler, Florian
    Oestreich, Marc-Alexander
    Frauchiger, Bettina S.
    Ramsey, Kathryn A.
    Latzin, Philipp
    JOURNAL OF APPLIED PHYSIOLOGY, 2021, 131 (03) : 1148 - 1156
  • [5] Impact of different breathing protocols on multiple-breath washout outcomes in children
    Yammine, Sophie
    Singer, Florian
    Gustafsson, Per
    Latzin, Philipp
    JOURNAL OF CYSTIC FIBROSIS, 2014, 13 (02) : 190 - 197
  • [6] Multiple-breath washout measurements can be significantly shortened in children
    Yammine, Sophie
    Singer, Florian
    Abbas, Chiara
    Roos, Markus
    Latzin, Philipp
    THORAX, 2013, 68 (06) : 586 - 587
  • [7] Correction of measurement error in a commercial multiple-breath washout device
    Wyler, Florian
    Oestreich, Marc-Alexander
    Frauchiger, Bettina
    Ramsey, Kathryn
    Latzin, Philipp
    EUROPEAN RESPIRATORY JOURNAL, 2021, 58
  • [8] Impact of Software Settings on Multiple-Breath Washout Outcomes
    Summermatter, Selina
    Singer, Florian
    Latzin, Philipp
    Yammine, Sophie
    PLOS ONE, 2015, 10 (07):
  • [9] Agreement between multiple-breath nitrogen washout systems in children and adults
    Poncin, William
    Singer, Florian
    Aubriot, Anne-Sophie
    Lebecque, Patrick
    JOURNAL OF CYSTIC FIBROSIS, 2017, 16 (02) : 258 - 266
  • [10] Impact of different breathing protocols on nitrogen multiple-breath washout in children
    Yammine, Sophie
    Singer, Florian
    Latzin, Philipp
    SWISS MEDICAL WEEKLY, 2013, 143 : 37S - 37S