177Lu SPECT imaging in the presence of 90Y: does 90Y degrade image quantification? a simulation study

被引:0
|
作者
Miller, Cassandra [1 ,2 ]
Uribe, Carlos [1 ,3 ,4 ]
Hou, Xinchi [1 ,4 ]
Rahmim, Arman [1 ,2 ,4 ]
Celler, Anna [4 ]
机构
[1] BC Canc Res Inst, Dept Integrat Oncol, Vancouver, BC, Canada
[2] Univ British Columbia, Dept Phys, Vancouver, BC, Canada
[3] BC Canc, Funct Imaging, Vancouver, BC, Canada
[4] Univ British Columbia, Dept Radiol, Vancouver, BC, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
SPECT; Monte Carlo simulations; dual-isotope imaging; activity quantification; crosstalk; dosimetry; RADIONUCLIDE THERAPY; NEUROENDOCRINE TUMORS; SCATTER CORRECTION; RENAL-FUNCTION; DOSIMETRY; RADIOEMBOLIZATION; TOXICITY;
D O I
10.1088/2057-1976/accde4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work aims to investigate the accuracy of quantitative SPECT imaging of Lu-177 in the presence of Y-90, which occurs in dual-isotope radiopharmaceutical therapy (RPT) involving both isotopes. We used the GATE Monte Carlo simulation toolkit to conduct a phantom study, simulating spheres filled with Lu-177 and Y-90 placed in a cylindrical water phantom that was also filled with activity of both radionuclides. We simulated multiple phantom configurations and activity combinations by varying the location of the spheres, the concentrations of Lu-177 and Y-90 in the spheres, and the amount of background activity. We investigated two different scatter window widths to be used with triple energy window (TEW) scatter correction. We also created multiple realizations of each configuration to improve our assessment, leading to a total of 540 simulations. Each configuration was imaged using a simulated Siemens SPECT camera. The projections were reconstructed using the standard 3D OSEM algorithm, and errors associated with Lu-177 activity quantification and contrast-to-noise ratios (CNRs) were determined. In all configurations, the quantification error was within +/- 6% of the no-Y-90 case, and we found that quantitative accuracy may slightly improve when Y-90 is present because of reduction of errors associated with TEW scatter correction. The CNRs were not significantly impacted by the presence of Y-90, but they were increased when a wider scatter window width was used for TEW scatter correction. The width of the scatter windows made a small but statistically significant difference of 1%-2% on the recovered Lu-177 activity. Based on these results, we can conclude that activity quantification of Lu-177 and lesion detectability is not degraded by the presence of Y-90.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Quantitative Imaging of 177Lu in the Presence of 90Y for Peptide Receptor Radionuclide Therapy: A Simulation Study
    Miller, C.
    Rahmim, A.
    Uribe-munoz, C.
    Celler, A.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E550 - E550
  • [2] Quantitative SPECT/CT reconstruction for 177Lu and 177Lu/90Y targeted radionuclide therapies
    Shcherbinin, S.
    Piwowarska-Bilska, H.
    Celler, A.
    Birkenfeld, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (18): : 5733 - 5747
  • [3] 90Y microsphere therapy: does 90Y PET/CT imaging obviate the need for 90Y Bremsstrahlung SPECT/CT imaging?
    Zade, Anand A.
    Rangarajan, Venkatesh
    Purandare, Nilendu C.
    Shah, Sneha A.
    Agrawal, Archi R.
    Kulkarni, Suyash S.
    Shetty, Nitin
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2013, 34 (11) : 1090 - 1096
  • [4] Comparative evaluation of radionuclide therapy using 90Y and 177Lu
    Hanaoka, Hirofumi
    Hashimoto, Kazuyuki
    Watanabe, Satoshi
    Matsumoto, Shojiro
    Sakashita, Tetsuya
    Watanabe, Shigeki
    Ishioka, Noriko S.
    Endo, Keigo
    [J]. ANNALS OF NUCLEAR MEDICINE, 2023, 37 (01) : 52 - 59
  • [5] Managing 177Lu and 90Y Dotatate Patients on an Outpatient Basis
    Archer, J. K. J.
    Jones, G. O.
    Carroll, M. J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S334 - S334
  • [6] Uncertainties in measurement of 90Y and 177Lu activities in an ionisation chamber
    Ballinger, J. R.
    Fernandez, R.
    Allen, S. J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S140 - S140
  • [7] Microdosimetry of Simulated Heterogeneous 177Lu and 90Y Activity Distributions
    Grimes, J.
    Celler, A.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [8] Comparative evaluation of radionuclide therapy using 90Y and 177Lu
    Hirofumi Hanaoka
    Kazuyuki Hashimoto
    Satoshi Watanabe
    Shojiro Matsumoto
    Tetsuya Sakashita
    Shigeki Watanabe
    Noriko S. Ishioka
    Keigo Endo
    [J]. Annals of Nuclear Medicine, 2023, 37 : 52 - 59
  • [9] Production of glass microspheres comprising 90Y and 177Lu for treating of hepatic tumors with SPECT imaging capabilities
    Poorbaygi, Hosein
    Aghamiri, Seyed Mahmoud Reza
    Sheibani, Shahab
    Kamali-asl, Alireza
    Mohagheghpoor, Elham
    [J]. APPLIED RADIATION AND ISOTOPES, 2011, 69 (10) : 1407 - 1414
  • [10] Comparative analysis of 90Y Bremsstrahlung SPECT/CT and 90Y PET/CT imaging of 90Y microsphere therapy
    Rangarajan, V.
    Zade, A.
    Jha, A.
    Singh, A.
    Shah, S.
    Agrawal, A.
    Purandare, N.
    Shetty, N.
    Kulkarni, S.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S312 - S312