Evaluation of analysis volume in total reflection X-ray fluorescence analysis

被引:5
|
作者
Tsuji, Kouichi [1 ,2 ]
Taniguchi, Naoya [1 ]
Yamaguchi, Hiroki [1 ]
Matsuyama, Tsugufumi [1 ]
机构
[1] Osaka Metropolitan Univ OMU, Grad Sch Engn, Dept Chem & Bioengn, Osaka, Japan
[2] Osaka Metropolitan Univ OMU, Grad Sch Engn, Dept Chemistryand Bioengn, 3-3-138 Sugimoto,Sumiyoshi, Osaka 5588585, Japan
基金
日本学术振兴会;
关键词
analysis height; analysis volume; Au layer; total reflection X-ray fluorescence; trace analysis; ELEMENTAL ANALYSIS; SAMPLE PREPARATION; TRACE-ELEMENTS; TXRF; SPECTROMETRY; WATER;
D O I
10.1002/xrs.3335
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Total reflection X-ray fluorescence (TXRF) is widely used for trace elemental analysis, wherein a thin monochromatic X-ray beam is typically applied at glancing angles. The actual volume of the sample measured by the instrument (analysis volume) depends on the dimensions of the X-ray beam and glancing angle, and it is a key feature in sample preparation for TXRF analysis. Herein, the analysis volume, including the analysis region and analysis height, was experimentally evaluated. A thin Au layer approximately 5 nm thick and 1 mm in diameter was prepared at a specific position on a flat glass substrate using a vacuum evaporation device. The position of this Au layer was controlled by manipulating a mask plate with a small hole. Nine samples were prepared at different Au layer positions and analyzed at 8 azimuth angles using a tabletop TXRF instrument. Finally, the Au XRF intensity distribution was visualized on a glass substrate. It was found that the XRF emitted over a diameter of approximately 5 mm was effectively detected using the tabletop TXRF instrument. The analysis height was experimentally evaluated using a thin Au layer deposited on a polyimide film 5 mm in diameter and 5 nm thick. The Au layer height was adjusted by inserting polyimide thin films between the Au layer and the glass substrate. The Au XRF intensity drastically decreased with increasing Au layer height, suggesting that the XRF emitted from the sample is effectively detected at heights of <10 mu m by using the described setup.
引用
收藏
页码:357 / 363
页数:7
相关论文
共 50 条
  • [1] Total reflection x-ray fluorescence analysis - a review
    Wobrauschek, P.
    X-RAY SPECTROMETRY, 2007, 36 (05) : 289 - 300
  • [2] Analysis of Beverages by Total Reflection X-ray Fluorescence
    Schmeling, Martina
    SPECTROSCOPY, 2017, 32 (07) : 28 - 29
  • [3] Total reflection X-ray fluorescence analysis with chemical microchip
    Tsuji, Kouichi
    Hanaoka, Yousuke
    Hibara, Akihide
    Tokeshi, Manabu
    Kitamori, Takehiko
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (04) : 389 - 392
  • [4] Micro total reflection x-ray fluorescence (μ-TXRF) analysis
    Tsuji, Kouichi
    Kawamata, Masaya
    Nishida, Yousuke
    Nakano, Kazuhiko
    Sasaki, Ken-ichi
    X-RAY SPECTROMETRY, 2006, 35 (06) : 375 - 378
  • [5] Total reflection X-ray fluorescence analysis of Mongolian coals
    Ts. Amartaivan
    E. D. Greaves
    G. Bernasconi
    P. Wobrauschek
    Journal of Radioanalytical and Nuclear Chemistry, 1997, 220 : 249 - 252
  • [6] Mineral analysis of pollen by Total Reflection X-Ray Fluorescence
    Basso, Ines M.
    Lorenzo, Daniel S.
    Mouteira, Maria C.
    Custo, Graciela S.
    APPLIED RADIATION AND ISOTOPES, 2019, 152 : 168 - 171
  • [7] Total reflection X-ray fluorescence analysis of light elements
    Streli, C
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (03) : 281 - 293
  • [8] Total reflection X-ray fluorescence analysis of Mongolian coals
    Amartaivan, T
    Greaves, ED
    Bernasconi, G
    Wobrauschek, P
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 1997, 220 (02) : 249 - 252
  • [9] Total reflection X-ray fluorescence analysis and related methods
    DeBoer, DKG
    Klockenkaemper, R
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (09) : R7 - R7
  • [10] Surface analysis by total-reflection X-ray fluorescence
    Sanchez, HJ
    Perez, CA
    Perez, RD
    Rubio, M
    RADIATION PHYSICS AND CHEMISTRY, 1996, 48 (03): : 325 - 331