Clinical paper Automated external defibrillator electrode size and termination of ventricular fibrillation in out-of-hospital cardiac arrest

被引:1
|
作者
Yin, Rose T. [1 ,3 ]
Taylor, Tyson G. [1 ]
de Graaf, Corina [2 ]
Ekkel, Mette M. [2 ]
Chapman, Fred W. [1 ]
Koster, Rudolph W. [2 ]
机构
[1] Stryker Emergency Care, Redmond, WA 98052 USA
[2] Amsterdam UMC, Acad Med Ctr, Amsterdam, Netherlands
[3] 11811 Willows Rd NE, Redmond, WA 98052 USA
关键词
Electrodes; Defibrillation; Out-of-hospital cardiac arrest; AED; Biphasic waveform; Impedance compensation; PADDLE PLACEMENT; IMPEDANCE;
D O I
10.1016/j.resuscitation.2023.109754
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Smaller electrodes allow more options for design of automated external defibrillator (AED) user interfaces. However, previous studies employing monophasic-waveform defibrillators found that smaller electrode sizes have lower defibrillation shock success rates. We hypothesize that, for impedance-compensated, biphasic truncated exponential (BTE) shocks, smaller electrodes increase transthoracic impedance (TTI) but do not Methods and Results: In this prospective before-and-after clinical study, Amsterdam police and firefighters used AEDs with BTE waveforms: an AED with larger electrodes in 2016-2017 (113 cm2), and an AED with smaller electrodes in 2017-2020 (65 cm2). We analyzed 157 and 178 patient cases with an initial shockable rhythm where the larger and smaller electrodes were used, respectively. A single 200-J shock terminated ventricular fibrillation (VF) in 86% of patients treated with large electrodes and 89% of patients treated with smaller electrodes. Small electrodes had a noninferior first shock defibrillation success rate compared to large electrodes, with a difference of 3% (95% CI: -3% -9%) with the lower confidence limit remaining above the defined non-inferiority threshold. TTI was significantly higher for the smaller electrodes (median: 100 X) compared to the Conclusions: For AEDs with impedance-compensating BTE waveforms, TTI was higher for smaller electrodes than the large electrode electrodes. Overall defibrillation shock success for AEDs with smaller electrodes was non-inferior to the AEDs with larger electrodes.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Automated external defibrillator and operator performance in out-of-hospital cardiac arrest
    Zijlstra, Jolande A.
    Bekkers, Loes E.
    Hulleman, Michiel
    Beesems, Stefanie G.
    Koster, Rudolph W.
    RESUSCITATION, 2017, 118 : 140 - 146
  • [2] The Outcomes of Traumatic or Injured Out-of-Hospital Cardiac Arrest and Ventricular Fibrillation: Implications for Use Automated External Defibrillator Use
    Chen, Yen-Pin
    Chen, Jiun-Wei
    Chong, Kah-Meng
    Cheng, Ming-Tai
    Lu, Tsung-Chien
    Chiang, Wen-Chu
    Ma, Matthew Huei-Ming
    Ko, Patrick Chow-In
    CIRCULATION, 2014, 130
  • [3] Failure of cardiac rhythm analysis by automated external defibrillator in an out-of-hospital cardiac arrest
    Cmorej, P. Ch.
    Smrzova, E.
    Mach, J.
    ANESTEZIOLOGIE A INTENZIVNI MEDICINA, 2018, 29 (03): : 135 - 138
  • [4] Refining Automated External Defibrillator Signage to Improve Out-of-Hospital Cardiac Arrest Survival
    Kovoor, Joshua G.
    Bacchi, Stephen
    Gupta, Aashray K.
    Stretton, Brandon
    Page, Gregory J.
    Kovoor, Pramesh
    HEART LUNG AND CIRCULATION, 2023, 32 (08): : E51 - E52
  • [5] Automated External Defibrillator Use After Out-of-Hospital Cardiac Arrest at Recreational Facilities
    Kolkailah, Ahmed A.
    Chan, Paul S.
    Li, Qiang
    Uzendu, Anezi
    Khan, Mirza S.
    Girotra, Saket
    JAMA INTERNAL MEDICINE, 2024, 184 (02) : 218 - 220
  • [6] Interruption of cardiopulmonary resuscitation with the use of the automated external defibrillator in out-of-hospital cardiac arrest
    Van Alem, AP
    Sanou, BT
    Koster, RW
    EUROPEAN HEART JOURNAL, 2003, 24 : 729 - 729
  • [7] Automated external defibrillator use in out-of-hospital cardiac arrest: Current limitations and solutions
    Delhomme, Clemence
    Njeim, Mario
    Varlet, Emilie
    Pechmajou, Louis
    Benameur, Nordine
    Cassan, Pascal
    Derkenne, Clement
    Jost, Daniel
    Lamhaut, Lionel
    Marijon, Eloi
    Jouven, Xavier
    Karam, Nicole
    ARCHIVES OF CARDIOVASCULAR DISEASES, 2019, 112 (03) : 217 - 222
  • [8] Interruption of cardiopulmonary resuscitation with the use of the automated external defibrillator in out-of-hospital cardiac arrest
    van Alem, AP
    Sanou, BT
    Koster, RW
    ANNALS OF EMERGENCY MEDICINE, 2003, 42 (04) : 449 - 457
  • [9] Automated external defibrillator: Rhythm analysis and defibrillation on paediatric out-of-hospital cardiac arrest
    Menant, Emma
    Lavignasse, Delphine
    Menetre, Sarah
    Didon, Jean-Philippe
    Jouven, Xavier
    RESUSCITATION PLUS, 2025, 22
  • [10] Use of a Drone-Delivered Automated External Defibrillator in an Out-of-Hospital Cardiac Arrest
    Schierbeck, Sofia
    Svensson, Leif
    Claesson, Andreas
    NEW ENGLAND JOURNAL OF MEDICINE, 2022, 386 (20): : 1953 - 1954