A very thin MRPC developed for TOF-PET

被引:0
|
作者
Liu, J. [1 ]
Wang, Y. [1 ]
Guo, B. [1 ]
Han, D. [1 ]
Li, Y. [1 ]
机构
[1] Tsinghua Univ, Dept Engn Phys, Key Lab Particle & Radiat Imaging, Beijing 100084, Peoples R China
来源
JOURNAL OF INSTRUMENTATION | 2024年 / 19卷 / 02期
基金
中国国家自然科学基金;
关键词
Gaseous detectors; Resistive-plate chambers;
D O I
10.1088/1748-0221/19/02/C02068
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
PET (Positron Emission Tomography) is a medical molecular imaging modality employed in cancer diagnosis and monitoring. Multi -gap Resistive Plate Chambers (MRPCs) are utilized in Time of Flight (TOF)-PET due to their superior time resolution. Our research group has successfully developed a 4 -chamber 32 -gap MRPC prototype with 0.128 mm gap thickness. The measured time resolution for cosmic rays is 20 ps, while the time resolution for the time difference between 0.511 MeV photon pairs is 193 ps Full Width at Half Maximum (FWHM). The 23.6 mm thickness of the MRPC prototype will affect the time resolution. To investigate the influence of the number and total thickness of gas gaps on the time resolution, Geant4 software is used to simulate the interaction between particles (such as cosmic rays and photons) and MRPC detectors. Most of the gamma photons interact with the resistive plates in MRPCs. It is ejected mostly a Compton electron into one gas gap, causing the gas ionization. In this paper, two 8 -gap MRPC prototypes with 0.128 mm gap thickness and 4.2 mm total thickness are fabricated and tested. The time resolution of 39 ps for cosmic rays and 166 ps FWHM for 0.511 MeV photons is obtained using the fast front-end amplifier and waveform acquisition system. So, a real time PET image can be obtained with a pair of MRPCs, and the detection efficiency for gamma can be improved by increasing the number of thin MRPCs.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Clinical evaluation of PET image quality as a function of acquisition time in a new TOF-PET/MR compared to TOF-PET/CT - initial results
    Konstantinos Zeimpekis
    Martin Huellner
    Felipe De Galiza Barbosa
    Edwin Ter Voert
    Helen Davison
    Gaspar Delso
    Patrick Veit-Haibach
    EJNMMI Physics, 2 (Suppl 1)
  • [22] Simultaneous Dual Isotope ToF-PET Imaging
    Liu, Zhengzhi
    Lee, Min Sun
    Chinn, Garry
    Levin, Craig
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [23] Measurement on properties of MPPC for TOF-PET system
    Yamazaki, M.
    Takeshita, T.
    Hasegawa, Y.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [24] An educational tool for demonstrating the TOF-PET technique
    Bäck, T
    Cederkäll, J
    Cederwall, B
    Johnson, A
    Kerek, A
    Klamra, W
    van der Marel, J
    Molnár, J
    Novák, D
    Sohler, D
    Steèn, M
    Uhlèn, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 471 (1-2): : 200 - 204
  • [25] Data driven time alignment for TOF-PET
    Rezaei, Ahmadreza
    Schramm, Georg
    Van Laere, Koenraad
    Nuyts, Johan
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [26] Development of TOF-PET using cherenkov radiation
    Miyata, Manabu
    Tomita, Hideki
    Watanabe, Kenichi
    Kawarabayashi, Jun
    Iguchi, Tetsuo
    J Nucl Sci Technol, 4 (339-343):
  • [27] Evaluation of the performance limit of MPPC for ToF-PET
    Tsujikawa, Takayuki
    Kataoka, Jun
    Ambe, Takahiro
    Fujita, Takuya
    Kurei, Yota
    Sato, Goro
    Ikeda, Hirolazu
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [28] A real-size MRPC developed for CBM-TOF
    ZHU WeiPing
    WANG Yi
    FENG ShengQin
    WANG JingBo
    HUANG XinJie
    SHI Li
    BABKIN V.
    GOLOVATYUK V.
    RUMIANTCEV M.
    Science China(Technological Sciences), 2013, (11) : 2821 - 2826
  • [29] A real-size MRPC developed for CBM-TOF
    ZHU WeiPing
    WANG Yi
    FENG ShengQin
    WANG JingBo
    HUANG XinJie
    SHI Li
    BABKIN V
    GOLOVATYUK V
    RUMIANTCEV M
    Science China(Technological Sciences), 2013, 56 (11) : 2821 - 2826
  • [30] Development of the Thin TOF-PET scanner based on fast monolithic silicon pixel sensors
    Hayakawa, D.
    Iacobucci, G.
    Paolozzi, L.
    Valerio, P.
    Ripiccini, E.
    Benoit, M.
    Ferrere, D.
    Ratib, O.
    Weber, M.
    Forshaw, D.
    Miucchi, A.
    Bandi, Y.
    Cardarelli, R.
    Ruecker, H.
    Kaynak, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 958 (958):