Comparison of 3 different bone-borne type expansion appliances used in surgically-assisted rapid palatal expansion: A finite element analysis

被引:2
|
作者
Koc, Osman [1 ]
Pamukcu, Hande [2 ]
Kocabalkan, Azize Atakan [2 ]
机构
[1] Yildiz Tech Univ, Fac Mech Engn, Istanbul, Turkiye
[2] Baskent Univ, Sch Dent, Dept Orthodont, Ankara, Turkiye
关键词
MAXILLARY EXPANSION; MINI-IMPLANTS; TOOTH-BORNE; EXPANDERS; STABILITY; SKELETAL; ANCHORAGE;
D O I
10.1016/j.ajodo.2022.12.001
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: This study aimed to compare the effects of 3 different bone-borne type expansion appliances used in the surgically-assisted rapid palatal expansion (SARPE) by finite element analysis. Methods: Three different miniscrew-supported palatal expansion appliances were modeled. Median and lateral osteotomies were performed without pterygomaxillary suture separation. Model I consisted of a palatal expander with 2 miniscrews placed 4 mm far from the midpalatal suture. In model II, 2 miniscrews were located at the alveolar ridge between the first molar and the second premolar. In model III, 4 miniscrews were placed as a combination of the first and second models. Stress distributions and amount of displacements were evaluated with Ansys software (version 19.2; Ansys, Canonsburg, Pa) for 5-mm expansion in a symmetrical finite element analysis model to reflect the clinical situation. Results: SARPE simulation using miniscrew-assisted maxillary expanders for all models showed a rotation and tipping of the maxilla. The largest displacement was found for the anterior part of the palate in model II and the posterior part in model III. Although a wedge-shaped expansion pattern was observed in all models, this form was more prominent in model II. The highest stress value (0.91 MPa) was measured in model I, and the lowest value (0.004 MPa) was measured in model II for the anterior nasal spine region. The highest stress value (0.51 MPa) was measured in model III, and the lowest value (0.12 MPa) was measured in model II for the posterior nasal spine region. The lowest stress values were measured in model II for all the craniofacial and maxillofacial structures. Conclusions: Among the models, the lowest stress distribution conditions for craniofacial and maxil-lofacial structures were found in model II. The largest displacement was found at the incisors and anterior part of the maxilla for model II. The greatest displacement was found at the posterior region for model III.
引用
收藏
页码:e23 / e33
页数:11
相关论文
共 50 条
  • [1] Following Surgically Assisted Rapid Palatal Expansion, Do Tooth-Borne or Bone-Borne Appliances Provide More Skeletal Expansion and Dental Expansion?
    Hamedi-Sangsari, Adrien
    Chinipardaz, Zahra
    Carrasco, Lee
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2017, 75 (10) : 2211 - 2222
  • [2] Maxillary changes with bone-borne surgically assisted rapid palatal expansion: A prospective study
    Asscherickx, Karlien
    Govaerts, Elke
    Aerts, Johan
    Vande Vannet, Bart
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2016, 149 (03) : 374 - 383
  • [3] Simulation of three surgical techniques combined with two different bone-borne forces for surgically assisted rapid palatal expansion of the maxillofacial complex: a finite element analysis
    Moehlhenrich, S. C.
    Modabber, A.
    Kniha, K.
    Peters, F.
    Steiner, T.
    Hoelzle, F.
    Fritz, U.
    Raith, S.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2017, 46 (10) : 1306 - 1314
  • [4] Surgically assisted palatal expansion with a bone-borne self-retaining palatal expander
    Aziz, Shahid R.
    Tanchyk, Amy
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2008, 66 (09) : 1788 - 1793
  • [5] Comparison of Tooth- and Bone-Borne Appliances on the Stress Distributions and Displacement Patterns in the Facial Skeleton in Surgically Assisted Rapid Maxillary Expansion-A Finite Element Analysis (FEA) Study
    Nowak, Rafal
    Olejnik, Anna
    Gerber, Hanna
    Fratczak, Roman
    Zawislak, Ewa
    MATERIALS, 2021, 14 (05) : 1 - 17
  • [6] Surgically assisted rapid maxillary expansion with bone-borne versus tooth-borne distraction appliances-a systematic review
    Blaehr, T. L.
    Mommaerts, M. Y.
    Kjellerup, A. D.
    Starch-Jensen, T.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2019, 48 (04) : 492 - 501
  • [7] Finite Element Analysis Model for Assessing Expansion Patterns from Surgically Assisted Rapid Palatal Expansion
    Lin, Jia-Hong
    Wu, Guan-Lin
    Chiu, Chun-Kai
    Wang, Steven
    Chung, Chun-Hsi
    Li, Chenshuang
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2023, (200):
  • [8] Is There a Significant Difference in Relapse and Complication Rate of Surgically Assisted Rapid Palatal Expansion Using Tooth-Borne, Bone-Borne, and Orthodontic Mini-Implant-Borne Appliances?
    Ploder, Oliver
    Winsauer, Heinz
    Juengling, Katharina
    Grill, Florian
    Bissinger, Oliver
    Wolff, Klaus-Dietrich
    Kolk, Andreas
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2021, 79 (01) : 213 - 224
  • [9] Surgically-assisted rapid palatal expansion for management transverse maxillary deficiency
    Gallagher, V
    Gallagher, C
    O'Brien, C
    Sleeman, D
    JOURNAL OF DENTAL RESEARCH, 2001, 80 (04) : 1157 - 1157
  • [10] Surgically-assisted rapid palatal expansion for management of transverse maxillary deficiency
    Silverstein, K
    Quinn, PD
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1997, 55 (07) : 725 - 727