Modelling and commissioning validation of eclipse conical cone collimator for stereotactic radiosurgery using Monte Carlo simulation

被引:0
|
作者
Shende, Ravindra [1 ,2 ]
Dhoble, S. J. [2 ]
Saroj, Dinesh [1 ]
Gupta, Gourav [1 ]
机构
[1] Balco Med Ctr, Dept Radiat Oncol, New Raipur, Chhattisgarh, India
[2] Rashtrasant Tukadoji Maharaj Nagpur Univ, Dept Phys, Nagpur, Maharashtra, India
关键词
commissioning validation; conical cone; Monte Carlo; PRIMO; SRS; VARIAN; DETECTORS; DEPTH; BEAMS;
D O I
10.1017/S1460396924000025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The miniaturized conical cones for stereotactic radiosurgery (SRS) make it challenging in measurement of dosimetric data needed for commissioning of treatment planning system. This study aims at validating dosimetric characteristics of conical cone collimator manufactured by Varian using Monte Carlo (MC) simulation technique.Methods & Material: Percentage depth dose (PDD), tissue maximum ratio (TMR), lateral dose profile (LDP) and output factor (OF) were measured for cones with diameters of 5mm, 7 center dot 5mm, 10mm, 12 center dot 5 mm, 15 mm and 17 center dot 5 mm using EDGE detector for 6MV flattening filter-free (FFF) beam from Truebeam linac. Similarly, MC modelling of linac for 6MVFFF beam and simulation of conical cones were performed in PRIMO. Subsequently, measured beam data were validated by comparing them with results obtained from MC simulation.Results: The measured and MC-simulated PDDs or TMRs showed close agreement within 3% except for cone of 5mm diameter. Deviations between measured and simulated PDDs or TMRs were substantially higher for 5mm cone. The maximum deviations at depth of 10cm, 20cm and at range of 50% dose were found 4 center dot 05%, 7 center dot 52%, 5 center dot 52% for PDD and 4 center dot 04%, 7 center dot 03%, 5 center dot 23% for TMR with 5mm cone, respectively. The measured LDPs acquired for all the cones showed close agreement with MC LDPs except in penumbra region around 80% and 20% dose profile. Measured and MC full-width half maxima of dose profiles agreed with nominal cone size within +/- 0 center dot 2 mm. Measured and MC OFs showed excellent agreement for cone sizes >= 10 mm. However, deviation consistently increases as the size of the cone gets smaller.Findings: MC model of conical cones for SRS has been presented and validated. Very good agreement was found between experimentally measured and MC-simulated data. The dosimetry dataset obtained in this study validated using MC model may be used to benchmark beam data measured for commissioning of SRS for cone planning.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Commissioning and Validation of a Monte Carlo Algorithm for Spine Stereotactic Radiosurgery
    Knill, C.
    Sandhu, R.
    Loughery, B.
    Lin, L.
    Halford, R.
    Drake, D.
    Snyder, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E456 - E456
  • [2] Commissioning and validation of a Monte Carlo algorithm for spine stereotactic radiosurgery
    Knill, Cory
    Sandhu, Raminder
    Loughery, Brian
    Lin, Lifeng
    Halford, Robert
    Drake, Doug
    Snyder, Michael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (11):
  • [3] Commissioning a multileaf collimator virtual cone for the stereotactic radiosurgery of trigeminal neuralgia
    Brown, Thomas A. D.
    Ayers, Rex G.
    Popple, Richard A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):
  • [4] Commissioning of the Eclipse TPS using a Monte Carlo software
    Calvo-Ortega, J.
    Hermida-Lopez, M.
    Casals-Farran, J.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1343 - S1343
  • [5] Validation of the PRIMO Monte Carlo software for stereotactic radiosurgery plans
    Hermida Lopez, M.
    Calvo-Ortega, J. F.
    Moragues-Femenia, S.
    Laosa Bello, C.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S732 - S733
  • [6] Monte Carlo Validation of Output Factors Measurements in Stereotactic Radiosurgery with Cones
    Garnier, Nicolas
    Amblard, Regis
    Villeneuve, Remy
    Haykal, Rodolphe
    Ortholan, Cecile
    Colin, Philippe
    Herault, Joel
    Belhomme, Sarah
    Benabdesselam, Mourad
    Serrano, Benjamin
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 469 - 475
  • [7] Inhomogeneity Correction using Monte Carlo Simulation for Leksell Gamma Knife Stereotactic Radiosurgery
    Moskvin, V. P.
    Henderson, M. A.
    Barriger, B.
    Can, M.
    DesRosiers, C.
    Johnstone, P. A. S.
    Das, I. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S680 - S680
  • [8] Monte Carlo simulation of conical collimators for stereotactic radiosurgery with a 6 MV flattening-filter-free photon beam
    Hermida-Lopez, Marcelino
    Sanchez-Artunedo, David
    Rodriguez, Miguel
    Brualla, Lorenzo
    MEDICAL PHYSICS, 2021, 48 (06) : 3160 - 3171
  • [9] Commissioning 6 MV photon beams of a stereotactic radiosurgery system for Monte Carlo treatment planning
    Deng, J
    Ma, CM
    Hai, J
    Nath, R
    MEDICAL PHYSICS, 2003, 30 (12) : 3124 - 3134
  • [10] Validation of a commercial Monte Carlo code used for stereotactic radiosurgery and stereotactic body radiation therapy
    Milroy, D.
    Dos Santos, G.
    Seuntjens, J.
    Patrocinio, H.
    MEDICAL PHYSICS, 2012, 39 (07) : 4620 - 4620