The possibility of automatic capillary blood testing in routine blood tests: an evaluation of the automatic mode of the Mindray BC-7500 CRP Auto Hematology Analyzer for capillary blood testing

被引:1
|
作者
He, Jiajing [1 ,2 ]
Zhang, Ge [1 ,2 ]
Wang, Yuefang [1 ,2 ]
Yang, Hao [1 ,2 ]
Dai, Qingkai [1 ,2 ]
Guo, Siqi [1 ,2 ]
Mai, Jia [1 ,2 ]
机构
[1] Sichuan Univ, West China Univ Hosp 2, Dept Lab Med, Chengdu, Peoples R China
[2] Sichuan Univ, Key Lab Birth Defects & Related Dis Women & Childr, Minist Educ, Chengdu, Peoples R China
关键词
Automated hematology analyzer; complete blood count (CBC); capillary blood; WHOLE-BLOOD; COUNTS; RHEOLOGY;
D O I
10.21037/cdt-23-84
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Capillary blood is a common specimen type used for infant blood routine tests. Until now, this specimen type could only be tested with the manual mode in hematology analyzers. Manual sample mixing and loading increases the amount labor force and can be more easily affected by human factors. This study was designed to investigate the proficiency of the automatic mode of the Mindray BC-7500 CRP Auto Hematology Analyzer for capillary blood testing. Methods: The complete blood count (CBC) results for capillary blood were compared between the automatic and manual modes. Special types of samples, including samples with high or low volume, thalassemia red cells, high fibrinogen, high hematocrit (HCT), or high triglyceride levels, were compared and evaluated. The intraclass correlation coefficient (ICC) was used to define the agreement between the 2 modes. The industry standard Analytical Quality Specifications for Routine Tests in Clinical Hematology (WS/T 406-2012), published by the National Health Commission of China, was used to evaluate the correlation between the results from the 2 modes.Results: There was good correlation between the automatic and manual modes for every type of sample, and the ICCs were all higher than 0.9. Except for high HCT or high triglyceride samples, there were no differences found between the 2 modes based on the WS/T 406-2012 standard.Conclusions: This new automatic mode utilized in the Mindray BC-7500 CRP Auto Hematology Analyzer for capillary blood yielded the same results as the manual mode except in the case of samples with high HCT or triglycerides. Capillary blood might be routinely tested automatically with hematology analyzers in the near future, which might reduce the labor required and improve standardization.
引用
收藏
页码:465 / 473
页数:9
相关论文
共 13 条
  • [1] Performance evaluation of routine blood and C-reactive protein analysis using Mindray BC-7500 CRP auto hematology analyzer
    Lin, Zhen
    Lin, Qiu
    Yu, Pingli
    Chen, Zhixin
    Lin, Haifeng
    Zhu, Bin
    Wang, Meihua
    Cao, Yingping
    ANNALS OF TRANSLATIONAL MEDICINE, 2022, 10 (10)
  • [2] DETERMINATION OF CAPILLARY BLOOD-GLUCOSE USING AUTOMATIC CLINICAL ANALYZER (ACA) DUPONT
    HEINEMANN, G
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1978, 16 (04): : 235 - 238
  • [3] A COMPARISON OF AUTOMATIC AND MANUAL BLOOD-GROUP ANTIBODY SCREENING IN BLOOD-DONORS FOR ROUTINE TESTING
    MCSHINE, RL
    DAS, PC
    SIBINGA, CTS
    JOURNAL OF AUTOMATIC CHEMISTRY, 1982, 4 (02): : 102 - 102
  • [4] Evaluation of automated capillary complete blood counts for routine clinical decision making in a large cohort of hematological patients, using Mindray BC-3000 Plus Auto and Sysmex XE-5000 hematology analyzers
    Oudatzis, Georgios
    Tsagarakis, Nikolaos J.
    Paterakis, Georgios
    Vasileiou, Paraskevi
    Xenou, Efthalia
    Maraki, Paraskevi
    Pangrati, Toula
    Angelopoulou, Maria K.
    Vassilakopoulos, Theodoros P.
    Konstantopoulos, Kostas
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (05) : 565 - 572
  • [5] Evaluation of blood cell count using an automatic hematology analyzer to optimize collection of peripheral blood progenitor cells by leukapheresis
    Pederzoli, Paula Renata Machado Passos
    Prata, Karen de Lima
    Cruz, Nathalia Gomide
    Marzano, Pedro Victorio de Almeida
    Martins, Mauricio Colombini
    Costa, Luciana de Almeida
    de Andrade, Roberta Kelly
    Libanio, Marcia Regina Issa Salomao
    Custer, Brian
    Belisario, Andre Rolim
    HEMATOLOGY TRANSFUSION AND CELL THERAPY, 2024, 46 : S150 - S156
  • [6] Detection of nucleated red blood cells using the Mindray BC-6800Plus hematology analyzer: a clinical performance evaluation
    Wang, Nengyong
    Jiang, Junyao
    Wang, Ji'an
    He, Yao
    Wen, Ning
    Liu, Yulin
    Wang, Yashu
    Li, Haijun
    Shi, Ping
    ANNALS OF PALLIATIVE MEDICINE, 2021, 10 (08) : 8808 - 8817
  • [7] Evaluation of a Blood Glucose Testing System with Automatic Ketone Body Measurement
    MacRury, S.
    Sachsenheimer, D.
    Mills, L.
    Deakin, S.
    Moore, K.
    Saini, S.
    Pfuetzner, A.
    DIABETES STOFFWECHSEL UND HERZ, 2016, 25 (06): : 343 - 347
  • [8] Evaluation of the point-of-care-testing DCA 2000 analyzer for HbA1c measurements in capillary blood and albumin/creatinine in urine.
    Hanslik, A
    Bürgstein, B
    Häring, HU
    Schmülling, RM
    Wahl, HG
    CLINICAL CHEMISTRY, 1998, 44 : A67 - A67
  • [9] Evaluation of remote triazole capillary blood testing to facilitate remote therapeutic drug monitoring (TDM): A validation study
    Nwankwo, Lisa
    Donovan, Jackie
    Ni, Melody
    Loebinger, Michael
    Armstrong-James, Darius
    Shah, Anand
    MEDICAL MYCOLOGY, 2022, 60 (SUPP 1) : 257 - 259
  • [10] Evaluation of three new methods of point-of-care testing in the measurement of blood cell counts as compared to the Siemens Advia 2120 hematology analyzer
    De laveleye, M.
    Detry, G.
    ACTA CLINICA BELGICA, 2017, 72 : 11 - 12