Implant-friendly MRI of deep brain stimulation electrodes at 7 T

被引:0
|
作者
Sadeghi-Tarakameh, Alireza [1 ,2 ]
DelaBarre, Lance [1 ]
Zulkarnain, Nur Izzati Huda [1 ]
Harel, Noam [1 ]
Eryaman, Yigitcan [1 ]
机构
[1] Univ Minnesota, Ctr Magnet Resonance Res, Minneapolis, MN USA
[2] Univ Minnesota, Ctr Magnet Resonance Res, 2021 6th St Southeast, Minneapolis, MN 55455 USA
关键词
deep brain stimulation safety; implant-friendly; MRI; radiofrequency heating; ultrahigh field; TRANSMISSION-LINE ARRAY; SUBTHALAMIC NUCLEUS; PARALLEL TRANSMIT; GLOBUS-PALLIDUS; SAFETY;
D O I
10.1002/mrm.29825
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study is to present a strategy to calculate the implant-friendly (IF) excitation modes-which mitigate the RF heating at the contacts of deep brain stimulation (DBS) electrodes-of multichannel RF coils at 7 T.Methods: An induced RF current on an implantable electrode generates a scattered magnetic field whose left-handed circularly polarizing component (B1(+)) is approximated using a B1(+)-mapping technique and subsequently used as a gauge for the electrode's induced current. Using this approach, the relative induced currents resulting from each channel of a multichannel RF coil on the DBS electrode were calculated. The IF modes of the corresponding multichannel coil were determined by calculating the null space of the relative induced currents. The proposed strategy was tested and validated for unilateral and bilateral commercial DBS electrodes (directional lead; Infinity DBS system, Abbott Laboratories) placed inside a uniform phantom by performing heating and imaging studies on a 7T MRI scanner using a 16-channel transceive RF coil.Results: Neither individual IF modes nor shim solutions obtained from IF modes induced significant temperature increase when used for a high-power turbo spin-echo sequence. In contrast, shimming with the scanner's toolbox (i.e., based on per-channel B1(+) fields) resulted in a more than 2(?)C temperature increase for the same amount of input power. Conclusion: A strategy for calculating the IF modes of a multichannel RF coil is presented. This strategy was validated using a 16-channel RF coil at 7 T for unilateral and bilateral commercial DBS electrodes inside a uniform phantom.
引用
下载
收藏
页码:2627 / 2642
页数:16
相关论文
共 50 条
  • [1] MRI in Patients with Deep Brain Stimulation Electrodes: Balancing Risks and Benefits
    Martin, Alastair J.
    RADIOLOGY, 2019, 293 (01) : 184 - 185
  • [2] Accuracy and distortion of deep brain stimulation electrodes on postoperative MRI and CT
    Pinsker, M. O.
    Herzog, J.
    Falk, D.
    Volkmann, J.
    Deuschl, G.
    Mehdorn, M.
    ZENTRALBLATT FUR NEUROCHIRURGIE, 2008, 69 (03): : 144 - 147
  • [3] Quantification of deep medullary veins at 7 T brain MRI
    Hugo J. Kuijf
    Willem H. Bouvy
    Jaco J. M. Zwanenburg
    Tom B. Razoux Schultz
    Max A. Viergever
    Koen L. Vincken
    Geert Jan Biessels
    European Radiology, 2016, 26 : 3412 - 3418
  • [4] Quantification of deep medullary veins at 7 T brain MRI
    Kuijf, Hugo J.
    Bouvy, Willem H.
    Zwanenburg, Jaco J. M.
    Schultz, Tom B. Razoux
    Viergever, Max A.
    Vincken, Koen L.
    Biessels, Geert Jan
    EUROPEAN RADIOLOGY, 2016, 26 (10) : 3412 - 3418
  • [5] CONCENTRATION OF THE SPECIFIC ABSORPTION RATE AROUND DEEP BRAIN STIMULATION ELECTRODES DURING MRI
    Mohsin, S. A.
    PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER, 2011, 121 : 469 - 484
  • [6] Multimodal Localization of Electrodes in Deep Brain Stimulation: Comparison of Stereotactic CT and MRI with Teleradiography
    Sauner, Dieter
    Runge, Matthias
    Poggenborg, Joerg
    Maarouf, Mohammad
    Sturm, Volker
    Treuer, Harald
    Hunsche, Stefan
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2010, 88 (04) : 253 - 258
  • [7] Deep brain stimulation and intraoperative MRI
    Elias, W. Jeff
    JOURNAL OF NEUROSURGERY, 2016, 124 (01) : 59 - 60
  • [8] Deep whole brain segmentation of 7T structural MRI
    Ramadass, Karthik
    Yu, Xin
    Cai, Leon Y.
    Tang, Yucheng
    Bao, Shunxing
    Kerley, Cailey
    D'Archangel, Micah
    Barquero, Laura A.
    Newton, Allen T.
    Gauthier, Isabel
    McGugin, Rankin Williams
    Dawant, Benoit M.
    Cutting, Laurie E.
    Huo, Yuankai
    Landman, Bennett A.
    MEDICAL IMAGING 2023, 2023, 12464
  • [9] High Efficiency Electrodes for Deep Brain Stimulation
    Grill, Warren M., Sr.
    Wei, Xuefeng F.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 3298 - 3301
  • [10] Utility of 7T Imaging for Deep Brain Stimulation Surgery
    Abosch, Aviva
    Harel, Noam
    Sapiro, Guillermo
    Duchin, Yuval
    Yacoub, Essa
    NEUROSURGERY, 2012, 71 (02) : E569 - E570