Efficacy and safety after posterior chamber Implantable Collamer Lens implantation according to preoperative anterior chamber depth: Short-term comparative study

被引:0
|
作者
Silva, Rui [1 ]
Franqueira, Nuno [1 ]
Faria-Correia, Fernando [1 ,2 ]
Mendes, Jose [1 ,2 ]
Oliveira, Mariana [1 ]
Monteiro, Tiago [1 ,2 ]
机构
[1] Hosp Braga, Dept Ophthalmol, P-4710243 Braga, Portugal
[2] Univ Minho, Life & Hlth Sci Res Inst ICVS, Escola Med, Braga, Portugal
关键词
Refractive Phakic IOLs < REFRACTIVE SURGERY; complications of refractive surgery < REFRACTIVE SURGERY; REFRACTIVE SURGERY; corneal topography/imaging systems < REFRACTIVE SURGERY; examination techniques:specular microscopy < CORNEA/EXTERNAL DISEASE; PHAKIC INTRAOCULAR-LENS; MYOPIA;
D O I
10.1177/11206721221131889
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare visual, refractive, and safety outcomes of posterior chamber Implantable Collamer Lens (ICL) implantation for the correction of myopia according to the preoperative anterior chamber depth (ACD). Methods: Retrospective, comparative study, patients submitted to implantation of myopic posterior-chamber phakic Implantable Collamer Lens (ICL), model V4C/V5, minimum follow-up of 12 months; two groups were created: Group 1 (ACD 2.80 to 2.99 mm) and Group 2 (ACD equal to or greater than 3.00 mm). The parameters evaluated were uncorrected and corrected visual acuity, subjective refraction, efficacy and safety index, predictability, endothelial cell density, central vault, anterior chamber angle and postoperative complications. A total of 558 eyes from 298 patients were evaluated: 111 eyes (19.9%) in group 1 and 447 eyes (80.1%) in group 2. Results: At 12 months, the efficacy index was similar in both groups ( p= 0.264); the safety index was higher in group 1 (p= 0.031); the mean central Vault was significantly lower in group 1 (212.8 vs 410.6 mu m; p < 0.001). Respectively, 93 (83.8%) and 366 (84.1%) eyes were within +/- 0.50 D of targeted refraction. Anterior chamber angle significantly decreased during follow-up in both groups ( p < 0.001; p < 0.001). Intraocular pressure did not change significantly ( p= 0.310 and p= 0.446, respectively). There were no significant differences in endothelial cell density loss ( p= 0.278) or in the rate of complications observed (p= 0.733). Conclusions: ICL implantation is an effective and safe procedure in eyes with shallow anterior chambers, with visual and refractive results and complication rates identical to those obtained in deeper anterior chambers.
引用
收藏
页码:822 / 830
页数:9
相关论文
共 50 条
  • [1] Short-term changes in and preoperative factors affecting vaulting after posterior chamber phakic Implantable Collamer Lens implantation
    Zhu, Qiu-Jian
    Chen, Wen-Jing
    Zhu, Wei-Jian
    Xiao, Hai-Xiang
    Zhu, Man-Hui
    Ma, Lie
    Yuan, You
    Song, E.
    BMC OPHTHALMOLOGY, 2021, 21 (01)
  • [2] Short‐term changes in and preoperative factors affecting vaulting after posterior chamber phakic Implantable Collamer Lens implantation
    Qiu-Jian Zhu
    Wen-Jing Chen
    Wei-Jian Zhu
    Hai-Xiang Xiao
    Man-Hui Zhu
    Lie Ma
    You Yuan
    E. Song
    BMC Ophthalmology, 21
  • [3] Anterior Segment Biometry During Accommodation After Posterior Chamber Phakic Implantable Collamer Lens Implantation
    Zhang, Guang-Ming
    Zhao, Li -Jun
    Xie, Hua-Tao
    Wu, Ming-Feng
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (03) : e164 - e172
  • [4] ANTERIOR-CHAMBER DEPTH AFTER POSTERIOR CHAMBER INTRAOCULAR-LENS IMPLANTATION
    ARAI, M
    OHZUNO, I
    ZAKO, M
    ACTA OPHTHALMOLOGICA, 1994, 72 (06): : 694 - 697
  • [5] Safety and anterior chamber structure of evolution implantable Collamer lens implantation with short white-to-white corneal diameters
    Chen, Xun
    Chen, Fang
    Wang, Xuanqi
    Xu, Yilin
    Cheng, Mingrui
    Han, Tian
    Wang, Xiaoying
    Zhou, Xingtao
    FRONTIERS IN MEDICINE, 2022, 9
  • [6] Safety and clinical results of posterior chamber phakic intraocular lens implantation in low anterior chamber depth
    Lee, Min Gyu
    Lim, Dong Hui
    Chung, Eui Sang
    Chung, Tae-Young
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [7] Influence of anterior chamber depth and vault on anterior chamber angle morphology after phakic posterior chamber intraocular lens implantation
    Sanchez-Trancon, Angel
    Manito, Santiago Cerpa
    Sierra, Oscar Torrado
    Baptista, Antonio Manuel
    Serra, Pedro Miguel
    INTERNATIONAL OPHTHALMOLOGY, 2024, 44 (01)
  • [8] Anterior Chamber Changes After Implantable Collamer Lens Implantation in High Myopia Using Pentacam: A Prospective Study
    Elmohamady M.N.
    Abdelghaffar W.
    Ophthalmology and Therapy, 2017, 6 (2) : 343 - 349
  • [9] Changes in anterior chamber depth and intraocular pressure after phacoemulsification and posterior chamber intraocular lens implantation
    Cekic, O
    Batman, C
    Totan, Y
    Emre, I
    Zilelioglu, O
    OPHTHALMIC SURGERY AND LASERS, 1998, 29 (08): : 639 - 642
  • [10] Analysis of the correlations between changes in posterior segment and anterior chamber segment after implantable collamer lens implantation in highly myopic patients
    Yifei Peng
    Lu Xiong
    Haokun Qu
    Yang Liu
    Zheng Wang
    BMC Ophthalmology, 25 (1)