Acetone Detection and Classification as Biomarker of Diabetes Mellitus Using a Quartz Crystal Microbalance Gas Sensor Array

被引:5
|
作者
Rodriguez-Torres, Marcos [1 ]
Altuzar, Victor [1 ]
Mendoza-Barrera, Claudia [1 ]
Beltran-Perez, Georgina [1 ]
Castillo-Mixcoatl, Juan [1 ]
Munoz-Aguirre, Severino [1 ]
机构
[1] Benemerita Univ Autonoma Puebla, Fac Ciencias Fis Matemat, Ave San Claudio & 18 Sur,Edificio FM1-101B,Ciudad, Puebla 72570, Mexico
关键词
quartz crystal microbalance; ultrasonic atomization; acetone biomarker; principal component analysis; discriminant analysis; diabetes mellitus; EXHALED BREATH ANALYSIS; ELECTRONIC-NOSE; TECHNOLOGIES; FREQUENCY; DISEASE;
D O I
10.3390/s23249823
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A gas sensor array was developed and evaluated using four high-frequency quartz crystal microbalance devices (with a 30 MHz resonant frequency in fundamental mode). The QCM devices were coated with ethyl cellulose (EC), polymethylmethacrylate (PMMA), Apiezon L (ApL), and Apiezon T (ApT) sensing films, and deposited by the ultrasonic atomization method. The objective of this research was to propose a non-invasive technique for acetone biomarker detection, which is associated with diabetes mellitus disease. The gas sensor array was exposed to methanol, ethanol, isopropanol, and acetone biomarkers in four different concentrations, corresponding to 1, 5, 10, and 15 mu L, at temperature of 22 degrees C and relative humidity of 20%. These samples were used because human breath contains them and they are used for disease detection. Moreover, the gas sensor responses were analyzed using principal component analysis and discriminant analysis, achieving the classification of the acetone biomarker with a 100% membership percentage when its concentration varies from 327 to 4908 ppm, and its identification from methanol, ethanol, and isopropanol.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Chitosan Coating on Quartz Crystal Microbalance Gas Sensor for Isopropyl Alcohol and Acetone Detection
    Hekiem, Nurul Liyana Lukman
    Ralib, Aliza Aini Md
    Akmal, M. H. Maziati
    Ahmad, Farah B.
    Ab Rahim, Rosminazuin
    Za'bah, Nor Farahidah
    Sugandi, Gandi
    INTERNATIONAL JOURNAL OF INTEGRATED ENGINEERING, 2022, 14 (03): : 241 - 249
  • [2] Multichannel Monolithic Quartz Crystal Microbalance Gas Sensor Array
    Jin, Xiaoxia
    Huang, Yue
    Mason, Andrew
    Zeng, Xiangqun
    ANALYTICAL CHEMISTRY, 2009, 81 (02) : 595 - 603
  • [3] Design and Optimization of a Multichannel Quartz Crystal Microbalance Sensor Array for Multiple Target Gas Detection
    Joseph, Aashish
    Emadi, Arezoo
    2019 IEEE SENSORS, 2019,
  • [4] Quartz crystal microbalance sensor array for the detection of volatile organic compounds
    Xu, Xiuming
    Cang, Huaiwen
    Li, Changzhi
    Zhao, Zongbao K.
    Li, Haiyang
    TALANTA, 2009, 78 (03) : 711 - 716
  • [5] Multichannel quartz crystal microbalance array: Fabrication, evaluation, application in biomarker detection
    Tao, Wenyan
    Lin, Peng
    Ai, Yanqing
    Wang, Hairui
    Ke, Shanming
    Zeng, Xierong
    ANALYTICAL BIOCHEMISTRY, 2016, 494 : 85 - 92
  • [6] Detection of β-caryophyllene in mango using a quartz crystal microbalance sensor
    Ali, Sk Babar
    Ghatak, Barnali
    Debabhuti, Nilava
    Sharma, Prolay
    Ghosh, Arunangshu
    Tudu, Bipan
    Bhattacharya, Nabarun
    Bandyopadhyay, Rajib
    SENSORS AND ACTUATORS B-CHEMICAL, 2018, 255 : 3064 - 3073
  • [7] Detection of volatile organic compounds using Quartz Crystal Microbalance sensor array and artificial neural network
    Noorsal, E
    Sidek, O
    Mohamad-Saleh, J
    Ahmad, MN
    2004 IEEE CONFERENCE ON CYBERNETICS AND INTELLIGENT SYSTEMS, VOLS 1 AND 2, 2004, : 931 - 936
  • [8] ITO thin films coated quartz crystal microbalance as gas sensor for NO detection
    Zhang, J
    Hu, JQ
    Zhu, FR
    Gong, H
    O'Shea, SJ
    SENSORS AND ACTUATORS B-CHEMICAL, 2002, 87 (01) : 159 - 167
  • [9] Detection of linalool in black tea using a quartz crystal microbalance sensor
    Sharma, Prolay
    Ghosh, Arunangshu
    Tudu, Bipan
    Bhuyan, Lakshi Prasad
    Tamuly, Pradip
    Bhattacharyya, Nabarun
    Bandyopadhyay, Rajib
    Chatterjee, Anutosh
    SENSORS AND ACTUATORS B-CHEMICAL, 2014, 190 : 318 - 325
  • [10] Quartz Crystal Microbalance Sensor Using Ionophore for Ammonium Ion Detection
    Kosaki, Yasuhiro
    Takano, Kosuke
    Citterio, Daniel
    Suzuki, Koji
    Shiratori, Seimei
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2012, 12 (01) : 563 - 567