Biomechanical Comparison of Cannulated Screw Osteosynthesis With Tension-Band Wiring for Proximal Fractures of the Fifth Metatarsal (Jones Fracture)

被引:1
|
作者
Unthan, Mark [1 ]
Graul, Isabel [1 ]
Hallbauer, Jakob [1 ]
Lindner, Robert [1 ,2 ]
Hofmann, Gunther O. [1 ]
Kohler, Felix C. [1 ,3 ]
机构
[1] Friedrich Schiller Univ Jena, Jena Univ Hosp, Dept Trauma Hand & Reconstruct Surg, D-07747 Jena, Germany
[2] Chirurg Orthopad Gemeinschaftspraxis Ingolstadt, Ostl Ringstr 4, D-85049 Ingolstadt, Germany
[3] Univ Klinikum Jena, Jena Univ Hosp, Am Klinikum 1, D-07747 Jena, Germany
来源
JOURNAL OF FOOT & ANKLE SURGERY | 2023年 / 62卷 / 02期
关键词
foot and ankle surgery; lag screw; proximal fifth metatarsal fracture; FIXATION; MANAGEMENT; METATARSAL;
D O I
10.1053/j.jfas.2022.08.004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Jones fractures, which lie at the junction of the diaphysis to the metaphysis of the fifth metatarsal, are a well -described clinical issue. There are various surgical approaches, including the commonly performed cannulated screw osteosyntheses, and the less frequently used tension-band approach. The aim is to compare the biomechan-ical stability of these osteosyntheses. We performed an osteotomy on 16 fresh frozen fifth metatarsal bones from body donors representing a Jones fracture. The fractures were treated pairwise with screw osteosynthesis or ten-sion-band wiring. This was followed by cyclic axial bending until osteosynthesis failure. Stability under axial bend-ing force was higher in the screw osteosynthesis (mean: 70.0 +/- 66.5 N) compared to the tension-band wiring (mean: 35.7 +/- 23.3 N) group although not reaching statistical significance (p = .116). The study shows no statisti-cally significant difference in biomechanical stability under axial loading between screw osteosynthesis and ten-sion band wiring. Based on the data obtained, no differences can be observed from a biomechanical point of view. The study supports the established method of treating Jones fractures primarily with screw osteosynthesis. In addition, the data suggest that tension band wiring may be a good alternative osteosynthesis, for example, after failed casting treatment or failure of primary osteosynthesis.(c) 2022 The Author(s). This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/)
引用
收藏
页码:300 / 303
页数:4
相关论文
共 50 条
  • [1] Tension-band wiring for fractures of the fifth metatarsal located in the junction of the proximal metaphysis and diaphysis
    Sarimo, J
    Rantanen, J
    Orava, S
    Alanen, J
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2006, 34 (03): : 476 - 480
  • [2] Treatment of fractures of the fifth metatarsal with the XS-nail retrospective study and comparison with tension-band wiring
    Renner, Christian
    Whyte, J.
    Singh, S.
    Friedl, W.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2010, 130 (09) : 1149 - 1156
  • [3] Treatment of fractures of the fifth metatarsal with the XS-nail retrospective study and comparison with tension-band wiring
    Christian Renner
    J. Whyte
    S. Singh
    W. Friedl
    Archives of Orthopaedic and Trauma Surgery, 2010, 130 : 1149 - 1156
  • [4] BIOMECHANICAL ASSESSMENT OF TENSION-BAND WIRING FOR OLECRANON FRACTURES
    Kuo, Chien-Chung
    Hsu, Horng-Chaung
    Hong, Shih-Wun
    Lu, Tung-Wu
    BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS, 2011, 23 (01): : 83 - 87
  • [5] Cannulated Screw Fixation of Jones Fifth Metatarsal Fractures: A Comparison of Titanium and Stainless Steel Screw Fixation
    DeVries, J. George
    Cuttica, Daniel J.
    Hyer, Christopher F.
    JOURNAL OF FOOT & ANKLE SURGERY, 2011, 50 (02): : 207 - 212
  • [6] Intramedullary screw fixation of proximal fifth metatarsal fractures: A biomechanical study
    Shah, SN
    Knoblich, GO
    Lindsey, DP
    Kreshak, J
    Yerby, SA
    Chou, LB
    FOOT & ANKLE INTERNATIONAL, 2001, 22 (07) : 581 - 584
  • [7] A Biomechanical Comparison of Fifth Metatarsal Jones Fracture Fixation Methods
    Duplantier, Neil L.
    Mitchell, Ronald J.
    Zambrano, Steve
    Stone, Aaron C.
    Delgado, Domenica A.
    Lambert, Bradley S.
    Moreno, Michael R.
    Harris, Joshua D.
    McCulloch, Patrick C.
    Lintner, David M.
    Varner, Kevin E.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2018, 46 (05): : 1220 - 1227
  • [8] Tension-band wiring through a single cannulated screw combined with suture anchors to treat inferior pole fracture of the patella
    Liu, Chen-dong
    Hu, Sun-jun
    Chang, Shi-Min
    Du, Shou-chao
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2023, 54 (04): : 1203 - 1209
  • [9] Comparison of Bicortical Headless Cannulated Screw Fixations, Tension Band Fixations, and Kirschner Wire Fixations of a Jones Fracture A Biomechanical Study
    Yilmaz, Onur
    Mutlu, Ibrahim
    Kuru, Tolgahan
    Bilge, Ali
    Gungor, Berna
    Kizilay, Hasan
    Olcar, Haci
    Ozkilic, Recai
    Erken, H. Yener
    JOURNAL OF THE AMERICAN PODIATRIC MEDICAL ASSOCIATION, 2024, 114 (06)
  • [10] Comparison of Cannulated Screw With Tension Band Wiring Versus Compressive Cannulated Locking Bolt and Nut Device (CompresSURE) in Patella Fractures-A Cadaveric Biomechanical Study
    Domby, Brian
    Henderson, Eric
    Nayak, Aniruddh
    Erdogan, Murat
    Gutierrez, Sergio
    Santoni, Brandon G.
    Sagi, H. Claude
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (12) : 678 - 683