Electroanalytical Properties of Unmodified and Modified Solid-Contact Potentiometric β-Lactam Sensors in Aqueous and Biological Media

被引:0
|
作者
Kulapina, E. G. [1 ]
Mursalov, R. K. [1 ]
Kulapina, O. I. [2 ]
Ankina, V. D. [2 ]
机构
[1] Saratov NG Chernyshevskii State Univ, Saratov 410012, Russia
[2] Saratov State Med Univ, Saratov 410012, Russia
基金
俄罗斯科学基金会;
关键词
cefixime; cefuroxime; cefotaxime; amoxicillin; solid-contact potentiometric sensors; nanoparticles; polyaniline nanotubes; medicinal and biological media; GLASSY-CARBON ELECTRODE; VOLTAMMETRIC DETERMINATION; ELECTROCHEMICAL DETERMINATION; NANOPARTICLES; MEMBRANES;
D O I
10.1134/S002016852314008X
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A comparative study of the electroanalytical properties of solid-contact sensors (tubular, planar) in solutions of cefuroxime (Cefur), cefotaxime (Ceftx), cefixime (Cefix), and amoxicillin (Amox) is performed. Associates of tetraalkylammonium-tetradecylammonium (TDA) and dimethyldistearylammonium (DMDSA)-with complex compounds of silver(I) and beta-lactam antibiotics [Ag(beta-lac)(2)]TAA and modifiers such as ZnO, polyaniline, and polyaniline nanotubes are used as electrode-active components (EACs). The sensors under study based on [Ag(Cefur)(2)]TDA and [Ag(Amox)(2)]DMDSA are characterized by a short response time in solutions of cefotaxime, cefuroxime, cefixime, and amoxicillin: 20-25 and 12-17 s for tubular unmodified and modified, respectively; 20-25 and 10-15 s for planar unmodified and modified, respectively. The linear range of electrode functions in solutions of antibiotics is 1 x 10(-4)-1 x 10(-2) mol/L; the limit of detection is 2.5 x 10(-5)-8.9 x 10(-5) mol/L for unmodified, 5.6 x 10(-6)-7.5 x 10(-5) mol/L for modified, and 4.2 x 10(-5)-7.2 x 10(-5) mol/L for planar sensors. The potential drift is 8-12 mV/day for unmodified and 5-7 mV/day for modified planar sensors; the service life is 1.5-2 months. An advantage of planar sensors is their use in microassay detection, which is relevant in the analysis of biological media. Application of solid-contact sensors for determination of the antibiotics under study in model aqueous solutions, medicinal products, and oral fluid in the case of various infectious diseases is demonstrated.
引用
收藏
页码:1418 / 1426
页数:9
相关论文
共 25 条
  • [1] Electroanalytical Properties of Unmodified and Modified Solid-Contact Potentiometric β-Lactam Sensors in Aqueous and Biological Media
    E. G. Kulapina
    R. K. Mursalov
    O. I. Kulapina
    V. D. Ankina
    Inorganic Materials, 2023, 59 : 1418 - 1426
  • [2] Separate detection of homologous surfactants by means of solid-contact unmodified and modified with molecular sieves potentiometric sensors
    Kulapin, AI
    Chernova, RK
    Kulapina, EG
    Mikhaleva, NM
    TALANTA, 2005, 66 (03) : 619 - 626
  • [3] Solid-contact potentiometric sensors with plasticized PVC membranes (review)
    Kulapin, A.I.
    Materova, E.A.
    Kulapina, E.G.
    Zavodskaya Laboratoriya. Diagnostika Materialov, 2002, 68 (12): : 3 - 12
  • [4] Beer classification based on the array of solid-contact potentiometric sensors with thiacalixarene receptors
    Stoikova, E. E.
    Dolgova, N. N.
    Savel'ev, A. A.
    Galukhin, A. V.
    Stoikov, I. I.
    Antipin, I. S.
    Evtyugin, G. A.
    RUSSIAN CHEMICAL BULLETIN, 2014, 63 (01) : 223 - 231
  • [5] Beer classification based on the array of solid-contact potentiometric sensors with thiacalixarene receptors
    E. E. Stoikova
    N. N. Dolgova
    A. A. Savel’ev
    A. V. Galukhin
    I. I. Stoikov
    I. S. Antipin
    G. A. Evtyugin
    Russian Chemical Bulletin, 2014, 63 : 223 - 231
  • [6] Solid-Contact Potentiometric Sensors for the Determination of Some Cephalosporin Antibiotics in Pharmaceuticals and Oral Fluid
    E. G. Kulapina
    M. S. Tyutlikova
    O. I. Kulapina
    A. E. Dubasova
    Journal of Analytical Chemistry, 2019, 74 : 52 - 58
  • [7] Solid-Contact Potentiometric Sensors for the Determination of Some Cephalosporin Antibiotics in Pharmaceuticals and Oral Fluid
    Kulapina, E. G.
    Tyutlikova, M. S.
    Kulapina, O. I.
    Dubasova, A. E.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2019, 74 (Suppl 1) : S52 - S58
  • [8] Some regularities of interface formation in solid-contact potentiometric surfactant-selective sensors
    A. I. Kulapin
    R. K. Chernova
    E. G. Kulapina
    Journal of Analytical Chemistry, 2005, 60 : 282 - 288
  • [9] Some regularities of interface formation in solid-contact potentiometric surfactant-selective sensors
    Kulapin, AI
    Chernova, RK
    Kulapina, EG
    JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 60 (03) : 282 - 288
  • [10] Selectivity of solid-contact Ag potentiometric sensors based on thiacalix[4]arene derivatives
    Evtugyn, G. A.
    Stoikov, I. I.
    Belyakova, S. V.
    Stoikova, E. E.
    Shamagsumova, R. V.
    Zhukov, A. Yu.
    Antipin, I. S.
    Budnikov, H. C.
    TALANTA, 2008, 76 (02) : 441 - 447