A survey and biomechanical analysis of the feasibility of the thumb test for determining the cancellous bone quality for stemless shoulder prosthesis

被引:1
|
作者
Hatta, Taku [1 ,7 ]
Shinagawa, Kiyotsugu [2 ]
Kawakami, Jun [3 ]
Kanazawa, Kenji [4 ]
Hayakawa, Takashi [5 ]
Yamamoto, Nobuyuki [3 ]
Yamakado, Kotaro [6 ]
机构
[1] Sports Clin Ishinomaki, Dept Orthoped Surg, Joint Surg, Ishinomaki, Japan
[2] Tohoku Rosai Hosp, Dept Orthoped Surg, Sendai, Japan
[3] Tohoku Univ, Sch Med, Dept Orthoped Surg, Sendai, Japan
[4] South Miyagi Med Ctr, Dept Orthoped Surg, Ogawara, Japan
[5] Niigata Cent Hosp, Dept Orthoped Surg, Niigata, Japan
[6] Fukui Gen Hosp, Dept Orthoped Surg, Fukui, Japan
[7] Sports Clin Ishinomaki, Dept Orthoped Surg, Joint Surg, 5-2-1 Ayumino, Ishinomaki 9860850, Japan
关键词
biomechanics; bone quality; stemless shoulder prosthesis; survey; thumb test; ARTHROPLASTY; FIXATION; DENSITY;
D O I
10.1177/10225536231218869
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundThe effect of the thumb test for assessing the cancellous bone quality at the resection plane of the proximal humerus on determining the application of a stemless shoulder prosthesis remains unclear. This study was conducted to survey the current utilization of the thumb test among surgeons and to investigate biomechanical features of the thumb test.MethodA survey among shoulder surgeons who had experience with stemless prostheses was conducted to investigate the current utilization of preoperative assessments and intraoperative thumb test when applying stemless prosthesis. Biomechanical experiments for the thumb test using artificial bone models were performed to assess the compression force, contact pressure and area. According to the preliminary survey, three compression techniques were assessed: compression perpendicular to the surface with thumb pad (P-pad technique) or tip of the thumb (P-tip technique), or compression in the vertical direction simulating compression along the longitudinal axis of the humeral shaft with tip-pad of the thumb (H-axis technique). The contact area was separated into three subregions (proximal, middle and distal) to assess the distribution of contact pressure.ResultsAmong 38 surgeons, 66% utilized the thumb test intraoperatively. The P-pad technique was more frequently applied than the P-tip or H-axis techniques (80%, 4% and 16%, respectively). Although with wide variation among the examiners, biomechanical assessments revealed the P-pad technique showed larger contact area and less compression force than the P-tip technique. The P-pad technique provided no significant localized differences in the mean contact pressure on the compressed plane, whereas the P-tip and H-axis techniques showed significant differences among subregions.ConclusionThis survey demonstrated relatively frequent application of the thumb test on applying the stemless shoulder prosthesis. Biomechanical assessment revealed the thumb test can hinder objective reproducibility among examiners; therefore, further investigations to identify feasible assessments of the bone quality is required.
引用
收藏
页数:9
相关论文
共 1 条
  • [1] The biomechanical assessment of two stemless shoulder arthroplasty prostheses in uniformly poor-quality bone mineral density cadaveric specimens
    Razavi, Ahmad Hedayatzadeh
    Nafisi, Nazanin
    Stewart, Isabella
    Abbasian, Mohammadreza
    Kheir, Nadim
    Shariyate, Mohammad Javad
    Khak, Mohammad
    Momenzadeh, Kaveh
    Asciutto, Dominic
    Ramappa, Arun J.
    Ross, Glen
    Shah, Sarav
    Nazarian, Ara
    [J]. CLINICAL BIOMECHANICS, 2024, 120