Investigating the impact of ghost images on the accuracy of modulation transfer function measurement using knife-edge method

被引:0
|
作者
Ryu, Dongok [1 ]
Kim, Seonghui [2 ]
Seong, Sehyun [2 ]
Seo, Hyunju [3 ]
Youk, Youngchun [1 ]
Kim, Goeun [1 ]
Kim, Shinwook [1 ]
Yoon, Jeeyeon [1 ]
Lee, Deog Gyu [1 ]
Heo, Haeng Pal [1 ]
机构
[1] Korea Aerosp Res Inst, 169 84 Gwahak Ro, Daejeon 34133, South Korea
[2] Telepix Co Ltd, 17 Techno 4 Ro, Daejeon 34013, South Korea
[3] Hanwha Syst Co Ltd, 491-23 Gyeonggidong Ro,Namsa eup, Yongin 17121, Gyeonggi Do, South Korea
关键词
modulation transfer function; ghost; ray tracing; knife-edge method; edge spread function; optical performance; performance validation; space telescope;
D O I
10.1117/12.2676634
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
In this study, we investigated the impact of ghost images on the modulation transfer function (MTF) of a Korsch-type telescope using nonsequential ray-tracing simulations and the experimental measurements of the knife-edge method with a collimator and light source targets. Our findings showed that ghost images introduce a directional bias into the edge spread function depending on the field position, which affects the line spread function and MTF. Furthermore, our measurement results demonstrated that ghost images can significantly affect the MTF on the edge field of the green channel. The ghost-to-signal ratio in the multispectral (MS) green channel was approximately 2.5%, which is approximately 0.25% higher than that in the panchromatic channel. To estimate the impact of ghost images in the MS green channel, we performed a parametric analysis using a nonsequential ray-tracing simulation, exploring potential strategies, such as adjusting the window thickness, the distance between the detector and the window, the transmittance of the window surface, and the reflectance of the detector surface. By comparing the positions and intensities of the ghost images obtained from the simulations with those measured experimentally, we identified the simulation input parameters that best reproduced the measured results. Our study provides valuable insight into the importance of managing ghost images when designing and operating Korsch-type telescopes to achieve the optimal image quality.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Non-uniform sampling knife-edge method for camera modulation transfer function measurement
    Duan, Yaxuan
    Xue, Xun
    Chen, Yongquan
    Tian, Liude
    Zhao, Jianke
    Gao, Limin
    [J]. OPTICAL METROLOGY AND INSPECTION FOR INDUSTRIAL APPLICATIONS IV, 2016, 10023
  • [2] A study of multi-angles knife-edge method applied to digital modulation transfer function measurement
    Yi-Chin Fang
    Ho-Lin Tsay
    [J]. Microsystem Technologies, 2021, 27 : 1429 - 1437
  • [3] A study of multi-angles knife-edge method applied to digital modulation transfer function measurement
    Fang, Yi-Chin
    Tsay, Ho-Lin
    [J]. MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2021, 27 (04): : 1429 - 1437
  • [4] An experimental study of the accuracy in measurement of modulation transfer function using an edge method
    Lee, Dong-Hoon
    Kim, Ye-seul
    Park, Hye-Suk
    Lee, Young-Jin
    Kim, Hee-Joung
    [J]. MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [5] A Knife-edge Input Point Spread Function Estimation Method for Document Images
    Zhong, Jiangiang
    [J]. PROCEEDINGS OF THE 2016 2ND INTERNATIONAL CONFERENCE ON SOCIAL SCIENCE AND TECHNOLOGY EDUCATION (ICSSTE 2016), 2016, 55 : 618 - 622
  • [6] EFFECT OF SPATIAL COHERENCE ON KNIFE-EDGE MEASUREMENTS OF DETECTOR MODULATION TRANSFER-FUNCTION
    WERNICK, MN
    MORRIS, GM
    [J]. APPLIED OPTICS, 1994, 33 (25): : 5906 - 5913
  • [7] Analysis of Edge Method Accuracy and Practical Multidirectional Modulation Transfer Function Measurement
    Wu, Yongjie
    Xu, Wei
    Piao, Yongjie
    Yue, Wei
    [J]. APPLIED SCIENCES-BASEL, 2022, 12 (24):
  • [8] Lunar limb knife-edge optical transfer function measurements
    Shen, JJ
    [J]. JOURNAL OF ELECTRONIC IMAGING, 1999, 8 (02) : 196 - 208
  • [10] MODELING THE IMAGES OF KNIFE-EDGE EXPOSURES AND MACRO CONTRAST OF DIFFUSION TRANSFER MATERIALS
    FYSON, JR
    [J]. JOURNAL OF PHOTOGRAPHIC SCIENCE, 1990, 38 (4-5): : 130 - 134