A finite element modeling of posterior atlantoaxial fixation and biomechanical analysis of C2 intralaminar screw fixation

被引:0
|
作者
Ma Xuexiao [1 ]
Peng Xianbo [1 ]
Xiang Hongfei [1 ]
Zhang Yan [1 ]
Zhang Guoqing [1 ]
Chen Bohua [1 ]
机构
[1] Department of Orthopaedic Surgery, the Affiliated Hospital of Qingdao University
关键词
D O I
暂无
中图分类号
R687.3 [骨骼手术];
学科分类号
1002 ; 100210 ;
摘要
Background The objective of this study was to use three-dimensional finite element(FE)models to analyze the stability and the biomechanics of two upper cervical fixation methods:the C2 intralaminar screw method and the C2 pedicle screw method.Methods From computed tomography images,a nonlinear three-dimensional FE model from C0(occiput)to C3 was developed with anatomic detail.The C2 intralaminar screw and the C2 pedicle screw systems were added to the model,in parallel to establish the interlaminar model and the pedicle model.The two models were operated with all possible states of motion and physiological loads to simulate normal movement.Results Both the C2 intralaminar screw method and the C2 pedicle screw method significantly reduced motion compared with the intact model.There were no statistically significant differences between the two methods.The Von Mises stresses of the internal and external laminar walls were similar between the two methods.Stability was also similar.Conclusions The C2 intralaminar screw method can complement but cannot completely replace the C2 pedicle screw method.Clinicians would need to assess and decide which approach to adopt for the best therapeutic effect.
引用
收藏
页码:1266 / 1271
相关论文
共 50 条
  • [1] A finite element modeling of posterior atlantoaxial fixation and biomechanical analysis of C2 intralaminar screw fixation
    Ma Xuexiao
    Peng Xianbo
    Xiang Hongfei
    Zhang Yan
    Zhang Guoqing
    Chen Bohua
    [J]. CHINESE MEDICAL JOURNAL, 2014, 127 (07) : 1266 - 1271
  • [2] Finite element modelling of posterior occiput-axis fixation and biomechanical analysis of C2 intralaminar screw fixation with offset connectors
    Gu, Jiyong
    Zhao, Minchan
    Wu, Yi
    Li, Zhibin
    Feng, Zize
    Jiang, Yalan
    Hu, Huixian
    Zhong, Minggu
    Luan, Zhongqin
    Xie, Zhengyuan
    Qi, Songtao
    Lu, Yuntao
    [J]. BRITISH JOURNAL OF NEUROSURGERY, 2024, 38 (01) : 104 - 110
  • [3] Biomechanical properties of different anterior and posterior techniques for atlantoaxial fixation: a finite element analysis
    Li, Jie
    Cao, Shuai
    Guo, Dong
    Lu, Teng
    Zang, Quanjin
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2023, 18 (01)
  • [4] Biomechanical properties of different anterior and posterior techniques for atlantoaxial fixation: a finite element analysis
    Jie Li
    Shuai Cao
    Dong Guo
    Teng Lu
    Quanjin Zang
    [J]. Journal of Orthopaedic Surgery and Research, 18
  • [5] Biomechanical analysis of the C2 intralaminar fixation technique using a cross-link and offset connector for an unstable atlantoaxial joint
    Lehman, Ronald A., Jr.
    Dmitriev, Anton E.
    Wilson, Kevin W.
    [J]. SPINE JOURNAL, 2012, 12 (02): : 151 - 156
  • [6] Biomechanical analysis of atlantoaxial dorsal fixation using finite element models
    Bae, Beomju
    Kim, Dongwook
    Oh, Hyejong
    Kim, Gonhyung
    [J]. THAI JOURNAL OF VETERINARY MEDICINE, 2022, 52 (01): : 185 - 192
  • [7] Biomechanical Effect of the C2 Laminar Decortication on the Stability of C2 Intralaminar Screw Construct and Biomechanical Comparison of C2 Intralaminar Screw and C2 Pars Screw
    Hong, Jae Taek
    Takigawa, Tomoyuki
    Udayakunmar, Ranjith
    Shin, Hun Kyu
    Simon, Peter
    Orias, Alejandro A. Espinoza
    Inoue, Nozomu
    An, Howard S.
    [J]. NEUROSURGERY, 2011, 69 : 1 - 6
  • [8] POSTERIOR ATLANTOAXIAL FACET SCREW FIXATION
    MARCOTTE, P
    DICKMAN, CA
    SONNTAG, VKH
    KARAHALIOS, DG
    DRABIER, J
    [J]. JOURNAL OF NEUROSURGERY, 1993, 79 (02) : 234 - 237
  • [9] Posterior atlantoaxial transarticular screw fixation
    Ko, Byung-Su
    Lee, Jung-Kil
    Kim, Yeon-Seong
    Moon, Sung-Jun
    Kim, Jae-Hyoo
    Kim, Soo-Han
    [J]. JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2007, 42 (03) : 179 - 183
  • [10] Biomechanical Effect of the C2 Laminar Decortication on the Stability of C2 Intralaminar Screw Construct and Biomechanical Comparison of C2 Intralaminar Screw and C2 Pars Screw COMMENTS
    Benzel, Edward C.
    Crawford, Neil R.
    Sonntag, Volker K. H.
    [J]. NEUROSURGERY, 2011, 69 : 6 - 7