Evaluation of CdZnTe spectrometer performance in measuring energy spectra during interventional radiology procedure

被引:0
|
作者
Lin Chai [1 ]
Lian Chen [2 ]
Cui-Ping Yang [1 ]
Dong-Dong Zhou [1 ]
Meng-Meng Yang [1 ]
Wei-Wei Qu [1 ,3 ]
Gao-Long Zhang [4 ,5 ]
Da-Qian Hei [6 ]
Shou-Ping Xu [7 ]
Xin-Jian Chen [1 ]
机构
[1] State Key Laboratory of Radiation Medicine and Protection,School of Radiation Medicine and Protection, Soochow University
[2] State Key Laboratory of Particle Detection and Electronics,University of Science and Technology of China
[3] Collaborative Innovation Center of Radiological Medicine of Jiangsu Higher Education Institutions
[4] School of Physics and Nuclear Energy Engineering, Beihang University
[5] Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, Beihang University
[6] College of Materials Science and Engineering, Nanjing University of Aeronautics and Astronautics
[7] Department of Radiation Oncology, PLA General Hospital
基金
中国博士后科学基金; 中国国家自然科学基金;
关键词
Interventional radiology procedures; Energyresponse curve; Energy spectrum; Radiation protection;
D O I
暂无
中图分类号
TL817 [谱仪];
学科分类号
082704 ;
摘要
Interventional radiology has been beneficial for patients for over 30 years of age with the combination of diagnostic and therapeutic methods. The radiation affecting occupationally exposed workers should be evaluated by means of the energy spectra and flux of X-rays in the treatment room. The present study aims to obtain the energy spectra of interventional procedures and study the capability of some detectors to evaluate the dose in interventional procedures. These measurements were taken by silicon-drift, Cd Te, and Cd Zn Te detectors. The energy spectra were corrected by the energy-response curve of each detector. The energy-response curves of silicon-drift and Cd Te detectors provided by the manufacturers specification were used. The energy response of the Cd Zn Te detector was measured by133Ba and152Eu c sources. The experimental data were compared with the simulation results, and their perfect agreement provides a way to correct the energy or dose response, which can be used for the personal dosimeter developed by our group.Moreover, the measured energy spectra can be used in individual radiation protection. The present study shows that the Cd Zn Te detector is a good candidate detector in interventional procedures.
引用
收藏
页码:91 / 96
页数:6
相关论文
共 50 条
  • [1] Evaluation of CdZnTe spectrometer performance in measuring energy spectra during interventional radiology procedure
    Lin Chai
    Lian Chen
    Cui-Ping Yang
    Dong-Dong Zhou
    Meng-Meng Yang
    Wei-Wei Qu
    Gao-Long Zhang
    Da-Qian Hei
    Shou-Ping Xu
    Xin-Jian Chen
    Nuclear Science and Techniques, 2019, 30
  • [2] Evaluation of CdZnTe spectrometer performance in measuring energy spectra during interventional radiology procedure
    Chai, Lin
    Chen, Lian
    Yang, Cui-Ping
    Zhou, Dong-Dong
    Yang, Meng-Meng
    Qu, Wei-Wei
    Zhang, Gao-Long
    Hei, Da-Qian
    Xu, Shou-Ping
    Chen, Xin-Jian
    NUCLEAR SCIENCE AND TECHNIQUES, 2019, 30 (09)
  • [3] Evaluation of the calibration procedure of active personal dosemeters for interventional radiology
    Bordy, J. -M.
    Daures, J.
    Clairand, I.
    Denoziere, M.
    Donadille, L.
    d'Errico, F.
    Gouriou, J.
    Itie, C.
    Struelens, L.
    RADIATION PROTECTION DOSIMETRY, 2008, 131 (01) : 87 - 92
  • [4] Measurement of internal constraints during.an interventional radiology procedure
    Ilic, D.
    Moix, I.
    Lambercy, O.
    Sache, L.
    Bleuler, H.
    Ohta, M.
    Augsburger, L.
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 2615 - 2618
  • [5] Intraprocedural Fall of an Obese Patient During an Interventional Radiology Procedure
    Liu, Brian
    Sadiq, Shayan
    Wang, Helen
    de Barros, Estele Odo
    Li, Zhuoxuan
    Nguyen, Kevin
    Jaipalli, Sujai
    Li, Molly
    Liddell, Robert P.
    JOURNAL OF PATIENT SAFETY, 2024, 20 (08) : e142 - e143
  • [6] Dose Evaluation in Medical Staff during Interventional Radiology Procedures
    Bacchim Neto, F. A.
    Alves, A. F. F.
    Alvarez, M.
    Rosa, M. E. D.
    Macedo de Freitas, C. C.
    Moura, R.
    Pina, D. R.
    Miranda, J. R. A.
    VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 : 927 - 930
  • [7] Evaluation of personal prescription treatment during an interventional procedure
    Richard, C.
    Berdougo, B.
    Cordonnier-Jourdin, C.
    Astier, A.
    Paul, M.
    Plaud, B.
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2014, 33 : A361 - A361
  • [8] Towards the estimation of the scattered energy spectra reaching the head of the medical staff during interventional radiology: A Monte Carlo simulation study
    Zagorska, A.
    Bliznakova, K.
    Buchakliev, Z.
    INTERNATIONAL CONFERENCE ON BIO-MEDICAL INSTRUMENTATION AND RELATED ENGINEERING AND PHYSICAL SCIENCES (BIOMEP 2015), 2015, 637
  • [9] An active pixel sensor to detect diffused X-ray during Interventional Radiology procedure
    Servoli, L.
    Battisti, D.
    Biasini, M.
    Checcucci, B.
    Conti, E.
    Di Lorenzo, R.
    Esposito, A.
    Fano, L.
    Paolucci, M.
    Passeri, D.
    Pentiricci, A.
    Placidi, P.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [10] Study on the Performance of an Alpha Energy Spectrum Radon Measuring Instrument Based on CdZnTe Detector
    Li, Songsong
    Shan, Jian
    Tan, Songbo
    HEALTH PHYSICS, 2021, 121 (02): : 87 - 91