Measurement of sensitivity of transducer in the field of NDT by the reciprocity technique

被引:0
|
作者
ZHAO Qichang and CHEN Zhaohua(Institute of Acoustics
机构
关键词
Measurement of sensitivity of transducer in the field of NDT by the reciprocity technique; NDT;
D O I
10.15949/j.cnki.0217-9776.1990.01.010
中图分类号
学科分类号
摘要
The sensitivity of a transducer should be an important characteristic inthe field of NDT.But so far,no method for this measurement of the sensitivity of atransducer applied to the ultrasonic flaw detection is available.In this paper thetransducer is regarded as a plane source and the reciprocity parameter under thecondition of the far field is derived.We have calibrated the sensitivity of ultrasonictransducers with the reciprocity technique in the solid sample.The experimentsshow and confirm that the transmitting and receiving sensitivities of the transducerapplied to NDT can be exactly measured.The precision of the measurement is with-in ± 1dB under a certain coupling condition.
引用
收藏
页码:82 / 88
页数:7
相关论文
共 50 条
  • [1] RECIPROCITY TECHNIQUE FOR ESTIMATING THE DIFFUSE-FIELD SENSITIVITY OF PIEZOELECTRIC TRANSDUCERS
    HILL, EVK
    EGLE, DM
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1980, 67 (02): : 666 - 672
  • [2] A note on determination of the diffuse-field sensitivity of microphones using the reciprocity technique
    Barrera-Figueroa, Salvador
    Rasmussen, Knud
    Jacobsen, Finn
    Journal of the Acoustical Society of America, 2008, 124 (03): : 1505 - 1512
  • [3] A note on determination of the diffuse-field sensitivity of microphones using the reciprocity technique
    Barrera-Figueroa, Salvador
    Rasmussen, Knud
    Jacobsen, Finn
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2008, 124 (03): : 1505 - 1512
  • [4] ESTIMATION OF EFFICIENCY OF A TRANSDUCER USING SELF-RECIPROCITY TECHNIQUE
    BINDAL, VN
    SAKSENA, TK
    SOM, JN
    CHANDRA, S
    ULTRASONICS, 1987, 25 (06) : 375 - 376
  • [5] MEASUREMENT OF TRANSDUCER EFFICIENCY USING SELF-RECIPROCITY TECHNIQUES
    WIDENER, MW
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1980, 67 (03): : 1058 - 1060
  • [6] Measurement methods of ultrasonic transducer sensitivity
    Xiao, Dingguo
    Fan, Qiong
    Xu, Chunguang
    Zhang, Xiuhua
    ULTRASONICS, 2016, 68 : 150 - 154
  • [7] NDT of on and offshore oil and gas installations using the alternating current field measurement (ACFM) technique
    Raine, GA
    Smith, N
    MATERIALS EVALUATION, 1996, 54 (04) : 461 - &
  • [8] NDT of on and offshore oil and gas installations using the alternating current field measurement (ACFM) technique
    Tochnical Software Consultants Ltd., 6 Mill Square, Featherstone Road, Milton Keynes MK12 5RB, United Kingdom
    不详
    Mater Eval, 4 (461-465):
  • [10] Calibration of a focusing transducer and miniature hydrophone as well as acoustic power measurement based on free-field reciprocity in a spherically focused wave field
    Shou, WD
    Duan, SM
    He, PZ
    Xia, RM
    Qian, DC
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (03) : 564 - 570