Attenuation correction for dedicated breast PET using only emission data based on consistency conditions

被引:0
|
作者
王璐 [1 ,2 ]
柴培 [1 ]
武丽伟 [1 ,2 ]
贠明凯 [1 ]
周小林 [1 ,2 ]
刘双全 [3 ]
张玉包 [3 ]
单保慈 [1 ]
魏龙 [1 ,3 ]
机构
[1] Key Laboratory of Nuclear Analysis Techniques,Institute of High Energy Physics,Chinese Academy of Sciences
[2] Graduate University of Chinese Academy of Sciences
[3] Beijing Engineering Research Center of Radiographic Techniques and Equipment
基金
中国国家自然科学基金;
关键词
attenuation correction; breast PET; consistency conditions;
D O I
暂无
中图分类号
S865 [各种野生动物驯养];
学科分类号
090705 ;
摘要
Accurate attenuation correction is required in dedicated breast PET imaging systems for image artifact removal and quantitative studies. In this study, a method using only emission data based on consistency conditions is proposed for attenuation correction in breast PET imaging systems. The consistency conditions are exploited to evaluate the accuracy of the estimated attenuation distribution and find the appropriate parameters that yield the most consistent attenuation distribution with the measured emission data. We have proved the validity of the method with experimental investigations and single-patient studies using a dedicated breast PET. The results show that the method is capable of accurately estimating the attenuation distribution of a uniform attenuator from the experimental data with various relatively low activities. The results also show that in single-patient studies, the method is robust for the irregular boundary of breast tissue and provides a distinct improvement in image quality.
引用
收藏
页码:138 / 143
页数:6
相关论文
共 50 条
  • [1] Attenuation correction for dedicated breast PET using only emission data based on consistency conditions
    王璐
    柴培
    武丽伟
    贠明凯
    周小林
    刘双全
    张玉包
    单保慈
    魏龙
    [J]. Chinese Physics C., 2013, 37 (01) - 143
  • [2] Attenuation correction for dedicated breast PET using only emission data based on consistency conditions
    Wang Lu
    Chai Pei
    Wu Li-Wei
    Yun Ming-Kai
    Zhou Xiao-Lin
    Liu Shuang-Quan
    Zhang Yu-Bao
    Shan Bao-Ci
    Wei Long
    [J]. CHINESE PHYSICS C, 2013, 37 (01)
  • [3] Nonuniform attenuation correction using only emission data with data consistency conditions and a "knowledge set."
    Panin, VY
    Zeng, GL
    Gullberg, GT
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (05) : 177P - 177P
  • [4] Attenuation correction using SPECT emission data only
    Gourion, Daniel
    Noll, Dominikus
    Gantet, Pierre
    Celler, Anna
    Esquerré, Jean-Paul
    [J]. IEEE Transactions on Nuclear Science, 2002, 49 I (05) : 2172 - 2179
  • [5] Attenuation correction using SPECT emission data only
    Gourion, D
    Noll, D
    Gantet, P
    Celler, A
    Esquerré, JP
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) : 2172 - 2179
  • [6] Attenuation correction using SPECT emission data only
    Gourion, D
    Noll, D
    Gantet, P
    Celler, A
    Esquerré, JP
    [J]. 2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1522 - 1526
  • [7] Registration problem in attenuation map estimation using only emission data with data consistency conditions.
    Panin, VY
    Zeng, GL
    Gullberg, GT
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 294P - 294P
  • [8] Utility of the Helgason-Ludwig consistency conditions for emission based attenuation correction of PET studies.
    Madsen, MT
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (05) : 132P - 132P
  • [9] Attenuation correction in PET using consistency information
    Welch, A
    Campbell, C
    Clackdoyle, R
    Natterer, F
    Hudson, M
    Bromiley, A
    Mikecz, P
    Chillcot, F
    Dodd, M
    Hopwood, P
    Craib, S
    Gullberg, GT
    Sharp, P
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (06) : 3134 - 3141
  • [10] Attenuation correction in PET using consistency conditions and a three-dimensional template
    Bromiley, A
    Welch, A
    Chilcott, F
    Waikar, S
    McCallum, S
    Dodd, M
    Craib, S
    Schweiger, L
    Sharp, P
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (04) : 1371 - 1377