Measurement of CO2 concentration based on supercontinuum laser absorption spectroscopy

被引:0
|
作者
李红莲 [1 ,2 ]
邸帅 [1 ,3 ]
李文铎 [1 ]
吕文静 [1 ]
王帆 [1 ]
方立德 [1 ]
机构
[1] College of Quality and Technical Supervision, Hebei University
[2] National and Local Joint Engineering Center of Measuring Instruments and Metrology Systems
[3] College of Chemistry & Environmental Science, Hebei University
关键词
D O I
暂无
中图分类号
TN249 [激光的应用];
学科分类号
0803 ; 080401 ; 080901 ;
摘要
The concentrations of CO2 were measured by the supercontinuum laser at normal temperature and variable temperature accurately in this paper. The absorption spectra of CO2 at different concentrations(1.2%—9.0%) were measured in the wavelength range of 1 425—1 445 nm under the optical path of 26.4 m at 293 K and 1 atm. The experimental results showed that the positions of the primary and secondary absorption peaks(1 432 nm, 1 437 nm) were consistent with the HITRAN database. A linear model of concentration and signal intensity at 1 432 nm was established. The maximum relative error of the concentration measurement was 3.3%. The line intensities of 99.9% CO2 in the 1 425—1 445 nm at different temperatures(298—373 K, interval of 15 K) were measured. The influence of temperature changes on the concentration measurement result was corrected and the relative error of the concentration measurement was reduced to 1.4%. Finally, the source of the uncertainty of the entire spectrum measurement system was analyzed and evaluated. This paper demonstrate that the supercontinuum laser can achieve the long-distance measurement of the CO2 under normal temperature or variable temperature environment accurately, which provide an important reference for the long-distance gas detection on site and simultaneous detection of multi-component gases.
引用
下载
收藏
页码:176 / 182
页数:7
相关论文
共 50 条
  • [1] Measurement of CO2 concentration based on supercontinuum laser absorption spectroscopy
    Li Hong-lian
    Di Shuai
    Li Wen-duo
    Lu Wen-jing
    Wang Fan
    Fang Li-de
    OPTOELECTRONICS LETTERS, 2021, 17 (03) : 176 - 182
  • [2] Measurement of CO2 concentration based on supercontinuum laser absorption spectroscopy
    Hong-lian Li
    Shuai Di
    Wen-duo Li
    Wen-jing Lü
    Fan Wang
    Li-de Fang
    Optoelectronics Letters, 2021, 17 : 176 - 182
  • [3] Measurement of CO2 concentration at high-temperature based on tunable diode laser absorption spectroscopy
    Chen, Jiuying
    Li, Chuanrong
    Zhou, Mei
    Liu, Jianguo
    Kan, Ruifeng
    Xu, Zhenyu
    INFRARED PHYSICS & TECHNOLOGY, 2017, 80 : 131 - 137
  • [4] Tunable diode laser absorption spectroscopy on 2.05 μm for the CO2 concentration measurement
    Pranovich, Alina
    Divoky, Martin
    Prochazka, Ivan
    Mocek, Tomas
    OPTICAL SENSORS 2015, 2015, 9506
  • [5] Research on CO2 concentration detection technology based on tunable diode laser absorption spectroscopy
    Zhang, Keke
    Zhang, Haidong
    Yang, Yingdong
    Wu, Yushang
    Yang, Xianglong
    Zheng, Xueyong
    SEMICONDUCTOR LASERS AND APPLICATIONS XIII, 2023, 12761
  • [6] On Line Simultaneous Measurement of CO/CO2/H2S Concentration Based on Laser Absorption Spectrum
    Zhang Xue-jun
    Chen Qin-gen
    Yang Zhan
    Deng Qin
    He Shuan-ling
    Peng Zhi-min
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44 (05) : 1412 - 1416
  • [7] 1 kHz MID-IR ABSORPTION SPECTROSCOPY FOR CO AND CO2 CONCENTRATION AND TEMPERATURE MEASUREMENT
    Fotia, Matthew L.
    Sell, Brian C.
    Hoke, John
    Wakefield, Stephen
    Schauer, Fred
    COMBUSTION SCIENCE AND TECHNOLOGY, 2015, 187 (12) : 1922 - 1936
  • [8] Measurements of CO2 Temperature and Concentration in High Temperature Environment Based on Mutable Diode Laser Absorption Spectroscopy
    Cui Haibin
    Wang Fei
    Li Meiyi
    LASER & OPTOELECTRONICS PROGRESS, 2018, 55 (05)
  • [9] CO2 Detection Based on Infrared Tunable Diode Laser Absorption Spectroscopy
    Wei Li
    Zhang Fan
    Liu Zhigang
    PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1-2, 2008, : 443 - 446
  • [10] Volcanic CO2 flux measurement at Campi Flegrei by tunable diode laser absorption spectroscopy
    Pedone, M.
    Aiuppa, A.
    Giudice, G.
    Grassa, F.
    Cardellini, C.
    Chiodini, G.
    Valenza, M.
    BULLETIN OF VOLCANOLOGY, 2014, 76 (04) : 1 - 13