Nonlinear Finite Element Analysis of Bone-Implant Contact in Three Short Dental Implant Models with Varying Osseointegration Percentages

被引:0
|
作者
Alemayehu, Dawit Bogale [1 ]
Todoh, Masahiro [2 ]
Huang, Song-Jeng [3 ]
机构
[1] Hokkaido Univ, Grad Sch Engn, Div Human Mech Syst & Design, Sapporo 0608628, Japan
[2] Hokkaido Univ, Fac Engn, Div Mech & Aerosp Engn, Sapporo 0608628, Japan
[3] Natl Taiwan Univ Sci & Technol, Dept Mech Engn, Taipei 10607, Taiwan
来源
ORAL | 2024年 / 4卷 / 04期
关键词
biomechanical; osseointegration; low density; high density; cancellous bone; short dental implant; finite element analysis; bone-implant contact; STRESS-DISTRIBUTION;
D O I
10.3390/oral4040040
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: Dental implants have become a cornerstone of restorative dentistry, providing a long-lasting method for tooth replacement. The degree of osseointegration has a significant effect on biomechanical stability at the bone-implant contact (BIC), determining the continued efficacy of these implants. However, the exact consequences of changing osseointegration levels on different implant designs, especially in bones with variable densities, are not well known. Methods: This study used 3D finite element analysis (FEA) to look at the biomechanical performance of three short dental implants: BioMet 3iT3, Straumann (R) Standard Plus Short-Regular Neck (SPS-RN), and Straumann (R) Standard Plus Short-Wide Neck (SPS-WN). This paper tested the implants at four stages of osseointegration: 25%, 50%, 75%, and 100% in both high-density (bone type III) and low-density (bone type IV) cancellous bone. It also created and examined realistic CAD models under static occlusal loading conditions to assess stress distribution and major strains at the bone-implant contact. Results: The study discovered that as osseointegration increases, von Mises stress and principal strains go down significantly for all implant types. The SPS-WN implant had the lowest strain values, especially for bone with low density. These reductions demonstrate increased mechanical stability as the bone-implant interface becomes more capable of dispersing mechanical stresses, minimizing the potential for localized deformation and bone resorption. Conclusions: The results highlight the importance of achieving optimum osseointegration to reduce mechanical stress and increase the lifespan of dental implants. The SPS-WN type implant performed better in biomechanical tests than the others, especially when bone conditions were not ideal. This makes it a great choice for clinical applications that need long-term implant success.
引用
收藏
页码:505 / 524
页数:20
相关论文
共 50 条
  • [1] The use of finite element analysis to model bone-implant contact with basal implants
    Ihde, Stefan
    Goldmann, Tomas
    Himmlova, Lucie
    Aleksic, Zoran
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2008, 106 (01): : 39 - 48
  • [2] A nonlinear homogenized finite element analysis of the primary stability of the bone-implant interface
    Ovesy, Marzieh
    Voumard, Benjamin
    Zysset, Philippe
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2018, 17 (05) : 1471 - 1480
  • [3] Biomechanical Analysis of Mandibular Bone-Implant Construct With Three Implant Screw Design: A Finite Element Study
    Banerjee, Anik
    Choudhury, Sandeep
    Rana, Masud
    Chakraborty, Arindam
    Gupta, Abhisek
    Chowdhury, Amit Roy
    Journal of Engineering and Science in Medical Diagnostics and Therapy, 2023, 6 (03)
  • [4] Effect of bone-implant contact percentage on bone remodeling surrounding a dental implant
    Lian, Zhiqiang
    Guan, Hong
    2010 International Conference on Biomedical Engineering and Computer Science, ICBECS 2010, 2010,
  • [5] Optimum Degree of Bone-Implant Contact in Bone Remodelling Induced by Dental Implant
    Lian, Z. Q.
    Guan, H.
    Loo, Y. C.
    PROCEEDINGS OF THE TWELFTH EAST ASIA-PACIFIC CONFERENCE ON STRUCTURAL ENGINEERING AND CONSTRUCTION (EASEC12), 2011, 14
  • [6] Dental implant stability and its measurements to improve osseointegration at the bone-implant interface: A review
    Kittur, Nidhi
    Oak, Radhika
    Dekate, Deepali
    Jadhav, Saakshi
    Dhatrak, Pankaj
    MATERIALS TODAY-PROCEEDINGS, 2021, 43 : 1064 - 1070
  • [7] Biomechanical investigations and finite element analysis of a bone-implant interface
    Widjaja, W
    Hartung, C
    BIOMEDIZINISCHE TECHNIK, 2001, 46 (12): : 351 - 354
  • [8] Osseointegration for transfemoral amputees: Influence of femur length and implant sizing on bone-implant contact and alignment
    Cao, Matthew
    Doherty, David B.
    He, Hongjia
    Vanodia, Vinay P.
    Reeves, Kristin
    Ismaily, Sabir K.
    Sauer, Jonathan J.
    Han, Shuyang
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2025, 56 (03):
  • [9] Bone-Implant Interface with Simulated Insertion Stress Around an Immediately Loaded Dental Implant in the Anterior Maxilla: A Three-Dimensional Finite Element Analysis
    Lee, Joon-Seok
    Cho, In-Ho
    Kim, Yung-Soo
    Heo, Seong-Joo
    Kwon, Ho-Beom
    Lim, Young-Jun
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (02) : 295 - 302
  • [10] High Bone-Implant Contact Achieved by Photofunctionalization to Reduce Periimplant Stress: A Three-Dimensional Finite Element Analysis
    Ohyama, Tetsuo
    Uchida, Tendo
    Shibuya, Norio
    Nakabayashi, Shinya
    Ishigami, Tomohiko
    Ogawa, Takahiro
    IMPLANT DENTISTRY, 2013, 22 (01) : 102 - 108